Skip to content

Channel Drain 072214Recalls & reported problems

C. R. Bard, Inc. · Channel Drain, 10mm Flat Full Fluted · Class 1

All models of Channel Drain (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741049798), received between Feb 13, 2019 and Sep 17, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material2
  • Break1
  • Defective Device1
  • Material Fragmentation1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Hematoma1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Nonstandard Device

    It was reported that Methodist Hospital OR has been using our wound drains as an alternative to their usual wound drains. The item numbers are the exact same, but the quality was poor. The hub where the channel meets the drain seems to be thicker than normal and has caused 3 hematomas. No medical intervention was reported.

    Patient problems: Hematoma

    MAUDE report 50802887 ↗
  • MalfunctionAug 15, 2025Defective Device; Device Contamination with Chemical or Other Material; Defective Device; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN ADDITIONAL COMPLAINTS HAVE BEEN REPORTED REGARDING THE SAME ITEM, AND IT HAS BEEN IDENTIFIED THAT THE ISSUE MAY AFFECT MULTIPLE ITEMS BEYOND THE ORIGINALLY REPORTED 072221. A MEETING IS REQUESTED TO DISCUSS THE ISSUE FURTHER. CONSIDERATION IS BEING GIVEN TO REMOVING THE AFFECTED PRODUCT AND RELATED ITEMS FROM INVENTORY UNTIL CONFIRMATION OF RESOLUTION IS OBTAINED. IMPACTED ITEMS INCLUDE: 072211. 072190. 072189. 072214. 072186. ADDITIONAL REPORTS DATED (B)(6) FROM THE CUSTOMER ARE INCLUDED BELOW. COMMUNICATION FROM (B)(6) WAS SUBMITTED AND DOCUMENTED UNDER PR (B)(4) AND (B)(4), WITH CUSTOMER OUTREACH PENDING. NEW REPORTS SHOULD BE REVIEWED TO CONFIRM ALIGNMENT WITH PREVIOUSLY OPENED RECORDS. ENSURE INCLUSION OF BD CUSTOMER QUALITY AND RELEVANT STAKEHOLDERS IN ALL CUSTOMER COMMUNICATIONS. PER ADDITIONAL INFORMATION RECEIVED VIA EMAIL ON 14AUG2025, IT WAS REPORTED THAT THEY SHARED CLINICIANS RAISED MULTIPLE CONCERNS RELATED TO ITEM 072221 AND THEN INSPECTED SIMILAR ITEMS AT THEIR DISTRIBUTION CENTER AND FOUND FUZZY HAIR FOREIGN MATTER IN THOSE ITEMS AS WELL. THERE WAS NO HARM TO A PATIENT AS THESE WERE FOUND BEFORE USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22816273 ↗
  • MalfunctionAug 12, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT PATIENT WAS THERE FOR A SURGERY. SURGEON ALWAYS USED TWO 10FR ROUND DRAINS (LOT# NGKP0821). SURGEON OPENED TWO DRAINS AT BEGINNING OF CASE, BOTH DRAINS HAD FUZZY MATERIAL ON THE DRAIN IN THE STERILE PACKAGING. THIS HAS HAPPENED BEFORE. SURGEON SWITCHED AND HAD TO USE 15FR DRAINS WHICH SHOWED NO CONTAMINATION. NO PATIENT HARM WAS REPORTED. NO PHYSICAL SAMPLE WAS AVAILABLE. IT WAS REPORTED THAT PATIENT WAS THERE FOR A SURGERY. SURGEON ALWAYS USED 10FR ROUND DRAINS WITH THIS SURGERY (LOT# NGKP0821). SURGEON OPENED 3 DIFFERENT PACKAGES OF STERILE DRAINS AND ALL WERE UNUSABLE. THEY HAD FUZZY STICKY STUFF ON DRAINS AND THE THIRD ONE HAD FUZZY SUBSTANCE AND A HAIR INSIDE THE STERILE PACKAGING. SURGEON PROCEEDED TO USE DIFFERENT SIZE DRAIN THAN NEEDED DUE TO THIS ONGOING PROBLEM. NO PATIENT HARM WAS REPORTED. PHYSICAL SAMPLE WAS AVAILABLE. IT WAS REPORTED THAT PHYSICIANS HAD SEVERAL ISSUES WITH 10F DRAINS FOR A WHILE (LOT# NGJW3604). WOUND DRAIN HAD DEBRIS, POWDER- LIKE SUBSTANCE ON THEM. VERY OFTEN THE DRAINS HAD SMALL WHITE PIECES OF PLASTIC THAT ARE STUCK TO THE CLEAR PORTION OF THE DRAIN. IT WAS SO FREQUENT THAT PHYSICIAN HAD THE STAFF CHECK THE TUBING BEFORE INSER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22784168 ↗
  • InjuryJan 18, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE WOUND DRAIN SUDDENLY SNAPPED WITHOUT PENROSE DRAIN PORTION EXITING THE SKIN AND THUS LEAVING REMAINDER OF DRAIN WITH THE PATIENT. STATED THAT THE DOCTOR ASSESSED THE SITE UNABLE TO LOCATE REMAINDER OF DRAIN, SO THE PATIENT WAS BROUGHT TO THE OPERATING ROOM TO HAVE THE EMBEDDED PORTION OF THE DRAIN REMOVED. AS PER SAMPLE FORM RECEIVED ON 13JAN2023, THE CUSTOMER STATED THAT WHEN WOUND DRAIN REMOVED ON (B)(6) 2022 IT SUDDENLY SNAPPED WITHOUT PENROSE DRAIN PORTION EXITING THE SKIN AND THUS LEAVING REMAINDER OF DRAIN WITH THE PATIENT. STATED THAT THE DOCTOR ASSESSED THE SITE UNABLE TO LOCATE REMAINDER OF DRAIN, SO THE PATIENT WAS BROUGHT TO THE OPERATING ROOM FOR REMOVAL. PATIENT HAD A SURGERY TO REMOVE BROKEN DEVICE OUT OF BODY.

    Patient problems: Foreign Body In Patient

    MAUDE report 16195633 ↗
  • MalfunctionFeb 13, 2019Break; Break

    IT WAS REPORTED THAT THE WOUND DRAIN BROKE APART WHERE THE 2 PIECES CONNECT TOGETHER, DURING AN ATTEMPT TO INSERT INTO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8337325 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page