Channel DrainRecalls & reported problems across all models
C. R. Bard, Inc. · 27 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 410
- Deaths reported
- 0
- Injuries reported
- 60
- Recalls
- 0
We found 27 models of Channel Drain by C. R. Bard, Inc. in the FDA's device registry (GUDID). Together they are named in 410 adverse event reports in the FDA's MAUDE database, received between Mar 28, 2017 and Sep 17, 2026 — classified as 60 injuries, 350 malfunctions.
The most frequently reported specific device problem is “Dull, Blunt” (57 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Channel Drain problems
Device problems
- Dull, Blunt57
- Device Contamination with Chemical or Other Material54
- Material Fragmentation51
- Break27
- Disconnection23
- Wrong Label21
- Incorrect Measurement17
- Suction Failure17
Patient problems
- No Clinical Signs, Symptoms or Conditions240
- No Known Impact Or Consequence To Patient70
- No Consequences Or Impact To Patient66
- Foreign Body In Patient35
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available27
- No Patient Involvement21
- Pain11
- No Information6
Channel Drain models
27 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Channel Drain 072229CATHETER, IRRIGATION65 reports
- Channel Drain 072221CATHETER, IRRIGATION53 reports
- Channel Drain 072227CATHETER, IRRIGATION46 reports
- Channel Drain 072223CATHETER, IRRIGATION27 reports
- Channel Drain 072189CATHETER, IRRIGATION23 reports
- Channel Drain V072229CATHETER, IRRIGATION23 reports
- Channel Drain 072231CATHETER, IRRIGATION19 reports
- Channel Drain 072187CATHETER, IRRIGATION15 reports
- Channel Drain 072188CATHETER, IRRIGATION15 reports
- Channel Drain V072230CATHETER, IRRIGATION15 reports
- Channel Drain 072211CATHETER, IRRIGATION14 reports
- Channel Drain 072230CATHETER, IRRIGATION12 reports
- Channel Drain 072190CATHETER, IRRIGATION10 reports
- Channel Drain 072217CATHETER, IRRIGATION10 reports
- Channel Drain 072234CATHETER, IRRIGATION10 reports
- Channel Drain 072191CATHETER, IRRIGATION8 reports
- Channel Drain 072228CATHETER, IRRIGATION8 reports
- Channel Drain V072231CATHETER, IRRIGATION8 reports
- Channel Drain 072225CATHETER, IRRIGATION6 reports
- Channel Drain 072214CATHETER, IRRIGATION5 reports
- Channel Drain V072228CATHETER, IRRIGATION5 reports
- Channel Drain 072212CATHETER, IRRIGATION4 reports
- Channel Drain 072216CATHETER, IRRIGATION4 reports
- Channel Drain V072234CATHETER, IRRIGATION2 reports
- Channel Drain 072186CATHETER, IRRIGATION1 report
- Channel Drain 072215CATHETER, IRRIGATION1 report
- Channel Drain 072235CATHETER, IRRIGATION1 report
Recent Channel Drain adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 17, 2026Nonstandard Device
It was reported that Methodist Hospital OR has been using our wound drains as an alternative to their usual wound drains. The item numbers are the exact same, but the quality was poor. The hub where the channel meets the drain seems to be thicker than normal and has caused 3 hematomas. No medical intervention was reported.
Patient problems: Hematoma
MAUDE report 50802887 ↗MalfunctionSep 15, 2026Wrong Label
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50781989 ↗MalfunctionSep 15, 2026Wrong Label
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50781537 ↗MalfunctionAug 27, 2026Failure to Capture
It was reported that after a kidney transplant, a drain was placed in the donor patient after surgery. The drain doesn't show up on the X ray and wanted to check if there was something wrong with the product. As usual, took the images with the Fujifilm Carneo GO Plus. The patient was a 75 year old woman and not obese. The device was still in place. Typically, a postoperative drainage tube would be visible in the image. Health professional plan to retrieve the device later if possible; for now, need to investigate whether any similar incidents have occurred with this specific lot.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50482356 ↗MalfunctionJul 1, 2026Material Too Rigid or Stiff
IT WAS REPORTED THAT ACCORDING TO THE CLINICAL RESOURCE DIRECTOR, THEY ARE GETTING RECENT COMPLAINTS FROM MY SATILLA FACILITY THAT THE WOUND DRAINS WERE TOO STIFF. THE FACILITY AND DOCTORS THERE HAVE USED THESE SAME DRAINS FOR A WHILE NOW, SO THEY ARE NOT SURE WHAT HAD CHANGED. DRAINS ARE TOO STIFF FOR THE CURRENT STAFF, AND THEY WERE WONDERING IF THERE IS A LINE OF PERHAPS SOFTER MORE FLEXIBLE SILICONE DRAINS, THEY CAN HAVE THEM TRY UNTIL THIS IS RESOLVED. MATERIAL (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25677176 ↗MalfunctionMay 27, 2026Material Split, Cut or Torn
IT WAS REPORTED BY CUSTOMER THAT THE LOSED SYSTEM DRAINAGE BAG WITH ANTI-REFLUX HAS HOLES IN THE PORTION CONNECTED TO THE BULB EVACUATOR OUTSIDE OF THE PATIENT, WHICH WAS AFFECTING ITS ABILITY TO MAINTAIN PROPER SUCTION. THIS WAS HAPPENING ONCE PATIENTS GO HOME, AND WHEN THEY ARE INSTRUCTED TO STRIP THE DRAINS TO MOVE FLUID, THE HOLES APPEAR TO GET LARGER. THEY ALSO NOTED THAT A BRAND-NEW DRAIN HAD HOLES UPON OPENING, AND THAT THIS HAS HAPPENED BEFORE AND DOES NOT SEEM TO BE RELATED TO A SPECIFIC LOT NUMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25309629 ↗MalfunctionMay 16, 2026Material Too Rigid or Stiff
IT WAS REPORTED BY CUSTOMER THAT THEY INCLUDED A TROCAR THAT WAS PERCEIVED AS TOO DULL, RAISING FEAR OF POTENTIAL PATIENT HARM DURING BREAST AND PLASTIC RECONSTRUCTION CASES, AS WELL AS THE DRAIN BEING NOTICEABLY STIFFER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25196021 ↗MalfunctionMay 16, 2026Dull, Blunt
IT WAS REPORTED BY CUSTOMER THAT THEY INCLUDED A TROCAR THAT WAS PERCEIVED AS TOO DULL, RAISING FEAR OF POTENTIAL PATIENT HARM DURING BREAST AND PLASTIC RECONSTRUCTION CASES, AS WELL AS THE DRAIN BEING NOTICEABLY STIFFER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25196015 ↗MalfunctionMay 14, 2026Material Fragmentation
IT WAS REPORTED THAT A REGISTERED NURSE WAS ATTEMPTING TO STRIP THE JP (JACKSONPRATT) DRAIN, AND THE TUBING BROKE APPROXIMATELY 6 INCHES FROM THE INSERTION SITE ON THE WOUND DRAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25174129 ↗MalfunctionApr 17, 2026Complete Blockage
IT WAS REPORTED THAT THE CUSTOMER HAD OCCURRED THE PREVIOUS WEEK WITH A 15FR ROUND DRAIN. DR HAD BEEN THE PROVIDER, AND IT HAD BEEN A BREAST CASE. THE CHARGE NURSE FROM GENERAL SURGERY HAD MENTIONED THAT THE DOCTOR DID NOT PREFER THE BARD DRAINS. A 15FR DRAIN HAD BEEN TRIED AND HAD CLOTTED, SO IT WAS CHANGED TO A 19FR DRAIN, WHICH HAD ALSO CLOTTED. THE CHARGE NURSE HAD STATED THAT A 24FR DRAIN WOULD NOT BE USED FOR A BREAST CASE. THE TEAM HAD RETURNED TO USING THE ETHICON DRAIN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24918000 ↗MalfunctionMar 31, 2026Incorrect Measurement
IT WAS REPORTED THAT THE CUSTOMER RECENTLY CONVERTED OUR BLAKE DRAINS TO THE BD 072227 DRAINS. CUSTOMER WAS RECEIVING SOME COMPLAINTS ON THE BD DRAINS, PRIMARILY ON THE SIZE OF THE TROCAR COMPARED TO THE BLAKE DRAINS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24749089 ↗MalfunctionMar 30, 2026Incorrect Measurement; Insufficient Device Problem Information
IT WAS REPORTED THAT THE CUSTOMER RECENTLY CONVERTED OUR BLAKE DRAINS TO THE BD 072227 AND 072229 DRAINS. CUSTOMER WAS RECEIVING SOME COMPLAINTS ON THE BD DRAINS, PRIMARILY ON THE SIZE OF THE TROCAR COMPARED TO THE BLAKE DRAINS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24719024 ↗InjuryMar 25, 2026Material Fragmentation
IT WAS REPORTED THAT A FEW DAYS AGO, AT THE FACILITY, STAFF WAS ATTEMPTED TO REMOVE A WOUND DRAIN FROM A PATIENT. THE DRAIN SNAPPED OFF INTO THE PATIENT. THEY HAD TO TAKE THE PATIENT TO THE OPERATING ROOM FOR REMOVAL OF THE RETAINED DRAIN PIECE. THEY UNSURE OF WHAT THEY DID WITH THE DRAIN PIECE WHETHER THAT WAS SENT TO PATHOLOGY FOR EVALUATION OR JUST DISCARDED. THEY JUST WANTED TO REPORT THAT INCIDENT IN CASE OTHER HAVE REPORTED A SIMILAR ISSUE.
Patient problems: Foreign Body In Patient
MAUDE report 24690499 ↗MalfunctionMar 19, 2026Fluid/Blood Leak
IT WAS REPORTED THAT THE PATIENT HAD A FOLEY CATHETER INSERTED PRIOR TO CESAREAN SECTION. URINE DRAINED IN COLLECTION BAG. WHEN THE SURGERY WAS COMPLETED AND THE BLANKET WAS PULLED BACK, THE FOLEY TUBING AND BULB WERE FOUND TO BE LAYING ON THE BED BETWEEN THE PATIENTS LEG, COMPLETELY OUT OF THE URETHRA. WHEN THE NURSE TRIED TO BLOW UP THE BALLOON AGAIN WITH WATER THERE WAS A SMALL HOLE WHERE THE WATER WAS LEAKING. PER FOLLOW-UP INFORMATION RECEIVED VIA EMAIL ON 12MAR2026, BOTH PRODUCTS ARE AVAILABLE. FOLEY CATHETER WAS REPLACED. NO TROUBLESHOOTING WAS DONE AS BOTH HAD LEAKS THAT WERE NOT REPAIRABLE. THE LOT #NGKR2507 AND NGKW1965.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24644398 ↗InjuryMar 4, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT THERE WAS A COMPLAINT OF INCREASED PAIN, LIMITED MOBILITY, AND THE INCREASED NEED FOR OPIOID MEDICATION WITH THE USE OF WOUND DRAIN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain
MAUDE report 24506148 ↗
Other C. R. Bard, Inc. products
- Bardex I.C. Foley Catheter26 models · 3,607 reports
- Bardex Lubricath Foley Catheter38 models · 1,670 reports · 1 recall
- Bardex Lubri-Sil I.C. All-Silicone Foley Catheter26 models · 1,518 reports
- Magic3 Coude Intermittent Catheter with Sure-Grip Sleeve5 models · 1,466 reports
- Magic3 Intermittent Catheter with Sure-Grip Sleeve14 models · 1,310 reports
- Bardia Foley Catheter Silicone Coated19 models · 1,283 reports
- PureWick Female External Catheter5 models · 1,184 reports
- Bardex All-Silicone Foley Catheter12 models · 1,170 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page