Broviac 0600060Recalls & reported problems
Bard Access Systems, Inc. · Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Trav with SureCuff Tissue Ingrowth Cuff · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Broviac single-lumen central venous catheter is a long-term implanted line for delivering fluids and medicines. The 33 reports mostly describe the catheter cracking or breaking after placement, sometimes while it is being removed, along with leaking and flow problems; 1 report mentions a death and 2 mention injuries. One open recall, started in December 2025, covers damage to the outer tray that can compromise the sterile barrier.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00801741051692), received between Nov 17, 2015 and Apr 17, 2026 — classified as 1 death, 2 injuries, 30 malfunctions.
The most frequently reported specific device problem is “Fracture” (21 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2025Open, Classified
Due to damage to outer tray that can potentially compromise the sterile barrier
Z-1367-2026 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Fracture21
- Fluid/Blood Leak6
- Break4
- Material Separation3
- Adverse Event Without Identified Device or Use Problem2
- Device Dislodged or Dislocated2
- Difficult to Flush2
- Failure to Infuse2
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Known Impact Or Consequence To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Consequences Or Impact To Patient2
- Swelling2
- Bradycardia1
- Death1
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Fluid/Blood Leak; Fracture
THE PATIENT UNDERWENT A CATHETER PLACEMENT PROCEDURE USING A HICKMAN/LEONARD/BROVIAC CENTRAL VENOUS CATHETER. AFTER SOME TIME, THE BROVIAC CATHETER WAS FOUND TO BE BROKEN AND LEAKING, WHICH NECESSITATED THE USE OF A REPAIR KIT. REPORTEDLY, THE CATHETER WAS REPLACED COMPLETELY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24906947 ↗MalfunctionMar 11, 2025Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT SOMETIMES POST CATHETER PLACEMENT PROCEDURE, THE CATHETER WAS ALLEGEDLY BROKEN. IT WAS FURTHER REPORTED THAT FLUID WAS LEAKING OUT OF THE INNER CYLINDER SIDE OF THE DUPLEX DURING THE PROCESS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 21580438 ↗MalfunctionSep 1, 2024Degraded; Fracture; Degraded; Fracture
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE CATHETER ALLEGEDLY BROKE DOWN OVER TIME. REPORTEDLY, THE CATHETER WAS REPLACED. THERE WAS NO REPORTER PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20115474 ↗MalfunctionJun 17, 2024Fracture; Material Separation; Fracture; Material Separation
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, WHILE DISSECTING THE CUFF, THE CATHETER ALLEGEDLY GOT DAMAGED AND GOT BROKEN DURING REMOVAL. REPORTEDLY, PATIENT REQUIRED SURGERY TO REMOVE THE FRAGMENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19546008 ↗MalfunctionDec 14, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY BROKE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18325258 ↗MalfunctionNov 11, 2023Fracture; Fracture
IT WAS REPORTED THAT SIX DAYS POST CATHETER PLACEMENT PROCEDURE, THE CATHETER WAS ALLEGEDLY FRACTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18117013 ↗MalfunctionJul 17, 2023Difficult to Flush; Fracture; Obstruction of Flow; Difficult to Flush; Fracture; Obstruction of Flow
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY HAD A BREAK AT THIN/THICK JUNCTURE. IT WAS FURTHER REPORTED THAT RESISTANCE WAS ALLEGEDLY FELT/OCCLUSION ALARMS FROM PUMP PRIOR TO BREAKS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17332163 ↗MalfunctionJul 17, 2023Difficult to Flush; Fracture; Obstruction of Flow; Difficult to Flush; Fracture; Obstruction of Flow
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE CATHETER ALLEGEDLY HAD A BREAK AT THIN/THICK JUNCTURE. IT WAS FURTHER REPORTED THAT THE RESISTANCE WAS ALLEGEDLY FELT AND OCCLUSION ALARMS FROM PUMP PRIOR TO BREAKS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17331914 ↗MalfunctionJul 17, 2023Break; Fracture; Stretched; Material Protrusion/Extrusion; Break; Fracture; Stretched; Material Protrusion/Extrusion
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY HAD A BREAK. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17331855 ↗MalfunctionJul 6, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING CATHETER PLACEMENT PROCEDURE, THE CATHETER WAS ALLEGEDLY BROKEN AT THE THIN/THICK JUNCTURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17266378 ↗MalfunctionJul 6, 2023Break; Fracture; Break; Fracture
HOLD FOR JF 7.6. IT WAS REPORTED THAT DURING CATHETER PLACEMENT PROCEDURE, THE CATHETER WAS ALLEGEDLY BROKEN AT THE THIN/THICK JUNCTURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17266377 ↗MalfunctionJul 6, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER WAS ALLEGEDLY BROKEN AT THE THIN/THICK JUNCTURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17264549 ↗MalfunctionJun 30, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT PRIOR TO A CHRONIC CATHETER PLACEMENT PROCEDURE, THE CATHETER ALLEGEDLY HAD HOLE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17235500 ↗MalfunctionMar 31, 2023Loss of or Failure to Bond; Disconnection; Fracture; Material Separation; Loss of or Failure to Bond; Disconnection; Fracture; Material Separation
IT WAS REPORTED THAT ONE DAY POST CATHETER PLACEMENT, THE OUTER TUBE ALLEGEDLY BECAME DISCONNECTED FROM THE INNER TUBE AT THE JUNCTURE OF THE TWO-TUBE SECTION. IT WAS FURTHER REPORTED THAT THE DEVICE WAS REPAIRED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16658698 ↗MalfunctionJul 15, 2022Fracture; Material Separation; Detachment of Device or Device Component; Fracture; Material Separation; Detachment of Device or Device Component
IT WAS REPORTED THAT APPROXIMATELY ONE MONTH AND TEN DAYS POST CHRONIC CATHETER PLACEMENT, THE PATIENT ALLEGEDLY EXPERIENCED PAIN. IT WAS FURTHER REPORTED THAT AN EXAMINATION DEMONSTRATED THAT THE CATHETER WAS ALLEGEDLY BROKEN AT THE EXTERNAL PART. REPORTEDLY CATHETER WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15025673 ↗MalfunctionJan 21, 2022Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT TWO DAYS POST CATHETER PLACEMENT PROCEDURE, THE LEAK WAS ALLEGEDLY FOUND. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13315237 ↗MalfunctionDec 10, 2021Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT ONE MONTH POST CATHETER PLACEMENT, THE DEVICE ALLEGEDLY HAD HOLE WITH FLUID LEAKING OUT AT JUNCTION OF THE HUB. IT WAS FURTHER REPORTED THAT DURING CATHETER EXCHANGE IN OPERATING ROOM THE CATHETER ALLEGEDLY FRACTURED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12974234 ↗MalfunctionOct 19, 2021Fracture; Fracture
IT WAS REPORTED THAT EIGHT DAYS POST PORT PLACEMENT, THE HUB WAS ALLEGED FOUND TO BE DAMAGED. IT WAS FURTHER REPORTED THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12653323 ↗MalfunctionOct 14, 2021Fracture; Fracture
IT WAS REPORTED THAT THIRTY DAYS POST PORT PLACEMENT, SILICON NEAR HUB ALLEGEDLY HAD A SPLIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12629563 ↗MalfunctionAug 31, 2021Fracture; Material Integrity Problem; Fracture; Material Integrity Problem
IT WAS REPORTED THAT TWELVE DAYS POST DEVICE IMPLANT, THE CATHETER OF THE KIT WAS ALLEGEDLY DAMAGED. THE DAMAGED SECTION OF THE CATHETER WAS REPAIRED USING A REPAIR KIT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12391636 ↗InjuryAug 17, 2021Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ONE DAY POST PORT DEVICE PLACEMENT, REDNESS WAS ALLEGEDLY FOUND ALONG THE CATHETER IMPLANTED AREA. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12329700 ↗MalfunctionNov 23, 2020Fracture; Fracture
IT WAS REPORTED THAT APPROXIMATELY THREE MONTHS POST PORT PLACEMENT, THE CATHETER ALLEGEDLY FRACTURED NEAR CUFF. THE PROCEDURE WAS COMPLETED WITH REINFORCEMENT OF THE CATHETER WITH A CLAMP. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10882635 ↗DeathApr 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT APPROXIMATELY FIVE HOURS AFTER PORT INSTALLATION AND ANTIBIOTIC TREATMENT IN A FIVE WEEK INFANT WITH A HISTORY OF BRONCHIOLITIS AND ENDOCARDITIS, THE PATIENT WAS TRANSFERRED TO THE INTENSIVE CARE UNIT FOR COMPRESSION HEMOPERICARDIUM, AND TAMPONADE. IT WAS FURTHER REPORTED THAT THE PATIENT ALLEGEDLY EXPIRED THE NEXT DAY FROM UNFAVORABLE DEVELOPMENT.
Patient problems: Bradycardia; Death
MAUDE report 10012417 ↗MalfunctionApr 19, 2019Failure to Infuse; Failure to Infuse
THIS REPORT SUMMARIZES ONE(1) MALFUNCTION EVENT. A REVIEW OF THE EVENTS INDICATED THAT MODEL 0600060 CHRONIC CATHETER SOME TIME POST INSERTION, THERE WAS ALLEGED RESISTANCE WHEN INFUSING THE CHRONIC CATHETER. THIS REPORT WAS RECEIVED FROM ONE SOURCE. THIS EVENT INVOLVED ONE PATIENT WHOSE AGE, WEIGHT AND GENDER WAS NOT PROVIDED. PATIENT HAD NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8533414 ↗MalfunctionApr 18, 2019Break; Defective Component; Break; Defective Component
THIS REPORT SUMMARIZES ONE MALFUNCTION EVENT. A REVIEW OF THE EVENT INDICATED THAT MODEL 600060 CHRONIC CATHETER EXPERIENCED A BREAK. THIS REPORT WAS RECEIVED FROM ONE SOURCE. OF THIS EVENT, THE DEVICE WAS USED ON THE PATIENT WITH NO REPORTED INJURY. THE PATIENT¿S AGE, SEX, AND WEIGHT WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8528896 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page