Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with… recallZ-1367-2026 · Bard Peripheral Vascular Inc · initiated Dec 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to damage to outer tray that can potentially compromise the sterile barrier
FDA root cause category: Process control
What the company is doing
On 01/12/1026, the firm sent via FedEx and/or email an "URGENT: Medical Device Product Removal" Letter informing customers that BD has identified via internal inspections that the packaging of some Hickman Catheter and Broviac Catheter kits may have damage present on the outer tray, potentially compromising the sterile barrier. Affected products were distributed between 06/12/2025 to 11/28/2025. Customers are instructed to: 1. Discontinue use of product listed in the letter and immediately dispose of all affected product remaining in their possession in accordance with their local facility s process. 2. Share this notification with all users within their facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, forward this notification to ensure that these entities are contacted and informed of the recall. 3. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not they have any of the product so that BD may acknowledge their receipt of this notification and subsequently process any required product credit. 4. Indicate on the response form the quantity of product identified at their facility and confirm that this product inventory was disposed of. For questions or assistance, contact: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: [email protected]
Recall details
- Recall number
- Z-1367-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2025
- Posted by the FDA
- Feb 13, 2026
- Recalling firm
- Bard Peripheral Vascular Inc
- Quantity
- 14,935 units
- Product code
- LJS
- FDA submissions
- K830233, K830256
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Argentina, Australia, Brazil, Canada, Chile, Hong Kong, India, Japan, Korea, Mexico, Malaysia, Singapore, Thailand, Uruguay, Vietnam.
Product description
Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Cat 60060
Code information (models, lots, serial numbers)
Show the full code information (14k characters)
Product Name: Broviac 4.2 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600060 UDI code: (01)00801741051692(17)300728(10)HUKV0547 Lot Number: HUKV0547 Product Name: Broviac 6.6 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600100 UDI code: (01)00801741051708(17)300628(10)HUKU0307 (01)00801741051708(17)300828(10)HUKW1312 Lot Number: HUKU0307 HUKW1312 Product Name: Broviac 6.6 Fr Single-Lumen CV Catheter Cutdown Tray (Short External Length) with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600120 UDI code: (01)00801741051715(17)300628(10)HUKU0274 Lot Number: HUKU0274 Product Name: Hickman 7 Fr Dual-Lumen CV Catheter, Cutdown Tray with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600310 UDI code: (01)00801741051739(17)300428(10)HUKS0783 (01)00801741051739(17)300628(10)HUKU0250 (01)00801741051739(17)300628(10)HUKU0310 Lot Number: HUKS0783 HUKU0250 HUKU0310 Product Name: Hickman 12 Fr Dual-Lumen CV Catheter, Cutdown Tray with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600350 UDI code: (01)00801741051777(17)300428(10)HUKR1359 (01)00801741051777(17)300428(10)HUKS0811 (01)00801741051777(17)300628(10)HUKU0316 Lot Number: HUKR1359 HUKS0811 HUKU0316 Product Name:Broviac 4.2 Fr Single-Lumen CV Catheter Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600520 UDI code: (01)00801741051807(17)270528(10)HUKS1468 (01)00801741051807(17)270528(10)HUKU0246 (01)00801741051807(17)270528(10)HUKU0278 (01)00801741051807(17)270528(10)HUKU0289 (01)00801741051807(17)270528(10)HUKU0305 (01)00801741051807(17)270528(10)HUKU0327 (01)00801741051807(17)270528(10)HUKU0338 (01)00801741051807(17)270528(10)HUKU0353 (01)00801741051807(17)270528(10)HUKW0164 (01)00801741051807(17)280328(10)HUKW1278 (01)00801741051807(17)280328(10)HUKW1285 Lot Number: HUKS1468 HUKU0246 HUKU0278 HUKU0289 HUKU0305 HUKU0327 HUKU0338 HUKU0353 HUKW0164 HUKW1278 HUKW1285 Product Name:Broviac 6.6 Fr Single-Lumen CV Catheter Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600540 UDI code: (01)00801741051821(17)290428(10)HUKR0064 (01)00801741051821(17)290628(10)HUKR0065 (01)00801741051821(17)290628(10)HUKR0067 (01)00801741051821(17)300128(10)HUKU0252 (01)00801741051821(17)300128(10)HUKU0270 (01)00801741051821(17)300128(10)HUKU0279 (01)00801741051821(17)300128(10)HUKU0308 (01)00801741051821(17)300128(10)HUKU0334 (01)00801741051821(17)300328(10)HUKV0359 (01)00801741051821(17)300328(10)HUKV0550 (01)00801741051821(17)300328(10)HUKV0560 (01)00801741051821(17)300328(10)HUKW0156 (01)00801741051821(17)300328(10)HUKW0159 Lot Number: HUKR0064 HUKR0065 HUKR0067 HUKU0252 HUKU0270 HUKU0279 HUKU0308 HUKU0334 HUKV0359 HUKV0550 HUKV0560 HUKW0156 HUKW0159 Product Name:Hickman 9.6 Fr Single-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600560 UDI code: (01)00801741051845(17)290828(10)HUKS0776 (01)00801741051845(17)290828(10)HUKS0816 (01)00801741051845(17)290928(10)HUKS1220 (01)00801741051845(17)290928(10)HUKU0302 (01)00801741051845(17)290928(10)HUKU0311 (01)00801741051845(17)300128(10)HUKU0345 (01)00801741051845(17)300128(10)HUKU0350 (01)00801741051845(17)300128(10)HUKV0376 (01)00801741051845(17)300128(10)HUKV0555 Lot Number: HUKS0776 HUKS0816 HUKS1220 HUKU0302 HUKU0311 HUKU0345 HUKU0350 HUKV0376 HUKV0555 Product Name:Hickman 7 Fr Dual-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600570 UDI code: (01)00801741051869(17)290628(10)HUKR1342 (01)00801741051869(17)290628(10)HUKR1344 (01)00801741051869(17)300128(10)HUKR1346 (01)00801741051869(17)290628(10)HUKR1347 (01)00801741051869(17)300128(10)HUKR1348 (01)00801741051869(17)300128(10)HUKR1356 (01)00801741051869(17)290628(10)HUKR1363 (01)00801741051869(17)290628(10)HUKR1367 (01)00801741051869(17)300128(10)HUKR1368 (01)00801741051869(17)300128(10)HUKS0774 (01)00801741051869(17)300128(10)HUKU0238 (01)00801741051869(17)300128(10)HUKU0242 (01)00801741051869(17)300128(10)HUKU0247 (01)00801741051869(17)300128(10)HUKU0253 (01)00801741051869(17)300128(10)HUKU0254 (01)00801741051869(17)300128(10)HUKU0266 (01)00801741051869(17)300128(10)HUKU0280 (01)00801741051869(17)300128(10)HUKU0283 (01)00801741051869(17)300128(10)HUKU0284 (01)00801741051869(17)300128(10)HUKU0290 (01)00801741051869(17)300128(10)HUKU0293 (01)00801741051869(17)300128(10)HUKU0294 (01)00801741051869(17)300128(10)HUKU0298 (01)00801741051869(17)300128(10)HUKU0312 (01)00801741051869(17)300128(10)HUKU0339 (01)00801741051869(17)300328(10)HUKU0340 (01)00801741051869(17)300128(10)HUKU0344 (01)00801741051869(17)300128(10)HUKU0354 (01)00801741051869(17)300328(10)HUKV0549 (01)00801741051869(17)300328(10)HUKV0569 (01)00801741051869(17)300328(10)HUKV0576 (01)00801741051869(17)300328(10)HUKV0581 (01)00801741051869(17)300328(10)HUKW0148 Lot Number: HUKR1342 HUKR1344 HUKR1346 HUKR1347 HUKR1348 HUKR1356 HUKR1363 HUKR1367 HUKR1368 HUKS0774 HUKU0238 HUKU0242 HUKU0247 HUKU0253 HUKU0254 HUKU0266 HUKU0280 HUKU0283 HUKU0284 HUKU0290 HUKU0293 HUKU0294 HUKU0298 HUKU0312 HUKU0339 HUKU0340 HUKU0344 HUKU0354 HUKV0549 HUKV0569 HUKV0576 HUKV0581 HUKW0148 Product Name:Hickman 12 Fr Dual-Lumen CV Catheter, Peel-Apart Introducer kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600620 UDI code: (01)00801741051913(17)291028(10)HUJY0302 (01)00801741051913(17)291028(10)HUJY1095 (01)00801741051913(17)291128(10)HUJZ1332 (01)00801741051913(17)291128(10)HUJZ1336 (01)00801741051913(17)291128(10)HUJZ1339 (01)00801741051913(17)291128(10)HUJZ1340 (01)00801741051913(17)300328(10)HUKR1349 (01)00801741051913(17)300328(10)HUKR1364 (01)00801741051913(17)300428(10)HUKS0784 (01)00801741051913(17)300428(10)HUKS0806 (01)00801741051913(17)300428(10)HUKS1217 (01)00801741051913(17)230428(10)HUKS1218 (01)00801741051913(17)300628(10)HUKU0259 (01)00801741051913(17)300628(10)HUKU0351 (01)00801741051913(17)300728(10)HUKV0349 (01)00801741051913(17)300728(10)HUKV0567 (01)00801741051913(17)300828(10)HUKW0534 Lot Number: HUJY0302 HUJY1095 HUJZ1332 HUJZ1336 HUJZ1339 HUJZ1340 HUKR1349 HUKR1364 HUKS0784 HUKS0806 HUKS1217 HUKS1218 HUKU0259 HUKU0351 HUKV0349 HUKV0567 HUKW0534 Product Name:Hickman 12.5 Fr Triple-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 600650 UDI code: (01)00801741051975(17)300228(10)HUKQ0535 (01)00801741051975(17)300428(10)HUKS1474 (01)00801741051975(17)300428(10)HUKS1483 (01)00801741051975(17)300628(10)HUKU0313 (01)00801741051975(17)300628(10)HUKU0317 (01)00801741051975(17)300628(10)HUKU0328 (01)00801741051975(17)300728(10)HUKV0354 (01)00801741051975(17)300728(10)HUKV0371 (01)00801741051975(17)300828(10)HUKW0171 (01)00801741051975(17)300828(10)HUKW1283 Lot Number: HUKQ0535 HUKS1474 HUKS1483 HUKU0313 HUKU0317 HUKU0328 HUKV0354 HUKV0371 HUKW0171 HUKW1283 Product Name:Hickman 10 Fr Triple-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 606560 UDI code: (01)00801741052019(17)290828(10)HUKR1370 (01)00801741052019(17)290928(10)HUKS1219 (01)00801741052019(17)300128(10)HUKU0255 (01)00801741052019(17)290928(10)HUKU0324 (01)00801741052019(17)300128(10)HUKV0578 (01)00801741052019(17)300128(10)HUKW0172 (01)00801741052019(17)300128(10)HUKW1315 Lot Number: HUKR1370 HUKS1219 HUKU0255 HUKU0324 HUKV0578 HUKW0172 HUKW1315 Product Name:Broviac 6.6 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600100CE UDI code: (01)00801741051708(17)300628(10)HUKU0307 (01)00801741051708(17)300828(10)HUKW1312 Lot Number: HUKU0307 HUKW1312 Product Name:Hickman 7 Fr Dual-Lumen CV Catheter, Cutdown Tray with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600310CE UDI code: (01)00801741051739(17)300428(10)HUKS0783 (01)00801741051739(17)300628(10)HUKU0250 (01)00801741051739(17)300628(10)HUKU0310 Lot Number: HUKS0783 HUKU0250 HUKU0310 Product Name:Broviac 4.2 Fr Single-Lumen CV Catheter Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600520CE UDI code: (01)00801741051807(17)270528(10)HUKS1468 (01)00801741051807(17)270528(10)HUKU0246 (01)00801741051807(17)270528(10)HUKU0278 (01)00801741051807(17)270528(10)HUKU0289 (01)00801741051807(17)270528(10)HUKU0305 (01)00801741051807(17)270528(10)HUKU0338 (01)00801741051807(17)270528(10)HUKU0353 (01)00801741051807(17)270528(10)HUKW0164 (01)00801741051807(17)280328(10)HUKW1278 (01)00801741051807(17)280328(10)HUKW1285 Lot Number: HUKS1468 HUKU0246 HUKU0278 HUKU0289 HUKU0305 HUKU0338 HUKU0353 HUKW0164 HUKW1278 HUKW1285 Product Name:Broviac 6.6 Fr Single-Lumen CV Catheter Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600540CE UDI code: (01)00801741051821(17)290428(10)HUKR0064 (01)00801741051821(17)290628(10)HUKR0065 (01)00801741051821(17)290628(10)HUKR0067 (01)00801741051821(17)300128(10)HUKU0252 (01)00801741051821(17)300128(10)HUKU0270 (01)00801741051821(17)300128(10)HUKU0279 (01)00801741051821(17)300128(10)HUKU0308 (01)00801741051821(17)300128(10)HUKU0334 (01)00801741051821(17)300328(10)HUKV0359 (01)00801741051821(17)300328(10)HUKV0550 (01)00801741051821(17)300328(10)HUKW0156 (01)00801741051821(17)300328(10)HUKW0159 Lot Number: HUKR0064 HUKR0065 HUKR0067 HUKU0252 HUKU0270 HUKU0279 HUKU0308 HUKU0334 HUKV0359 HUKV0550 HUKW0156 HUKW0159 Product Name:Hickman 9.6 Fr Single-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600560CE UDI code: (01)00801741051845(17)290828(10)HUKS0776 (01)00801741051845(17)290828(10)HUKS0816 (01)00801741051845(17)290928(10)HUKS1220 (01)00801741051845(17)290928(10)HUKU0302 (01)00801741051845(17)290928(10)HUKU0311 (01)00801741051845(17)300128(10)HUKU0345 (01)00801741051845(17)300128(10)HUKU0350 (01)00801741051845(17)300128(10)HUKV0555 Lot Number: HUKS0776 HUKS0816 HUKS1220 HUKU0302 HUKU0311 HUKU0345 HUKU0350 HUKV0555 Product Name:Hickman 7 Fr Dual-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600570CE UDI code: (01)00801741051869(17)290628(10)HUKR1342 (01)00801741051869(17)290628(10)HUKR1344 (01)00801741051869(17)300128(10)HUKR1346 (01)00801741051869(17)290628(10)HUKR1347 (01)00801741051869(17)300128(10)HUKR1348 (01)00801741051869(17)300128(10)HUKR1356 (01)00801741051869(17)290628(10)HUKR1363 (01)00801741051869(17)290628(10)HUKR1367 (01)00801741051869(17)300128(10)HUKR1368 (01)00801741051869(17)300128(10)HUKS0774 (01)00801741051869(17)300128(10)HUKU0238 (01)00801741051869(17)300128(10)HUKU0242 (01)00801741051869(17)300128(10)HUKU0247 (01)00801741051869(17)300128(10)HUKU0253 (01)00801741051869(17)300128(10)HUKU0254 (01)00801741051869(17)300128(10)HUKU0266 (01)00801741051869(17)300128(10)HUKU0280 (01)00801741051869(17)300128(10)HUKU0283 (01)00801741051869(17)300128(10)HUKU0284 (01)00801741051869(17)300128(10)HUKU0290 (01)00801741051869(17)300128(10)HUKU0293 (01)00801741051869(17)300128(10)HUKU0294 (01)00801741051869(17)300128(10)HUKU0298 (01)00801741051869(17)300128(10)HUKU0312 (01)00801741051869(17)300128(10)HUKU0339 (01)00801741051869(17)300328(10)HUKU0340 (01)00801741051869(17)300128(10)HUKU0344 (01)00801741051869(17)300128(10)HUKU0354 (01)00801741051869(17)300328(10)HUKV0549 (01)00801741051869(17)300328(10)HUKV0576 (01)00801741051869(17)300328(10)HUKW0148 Lot Number: HUKR1342 HUKR1344 HUKR1346 HUKR1347 HUKR1348 HUKR1356 HUKR1363 HUKR1367 HUKR1368 HUKS0774 HUKU0238 HUKU0242 HUKU0247 HUKU0253 HUKU0254 HUKU0266 HUKU0280 HUKU0283 HUKU0284 HUKU0290 HUKU0293 HUKU0294 HUKU0298 HUKU0312 HUKU0339 HUKU0340 HUKU0344 HUKU0354 HUKV0549 HUKV0576 HUKW0148 Product Name:Hickman 12 Fr Dual-Lumen CV Catheter, Peel-Apart Introducer kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600620CE UDI code: (01)00801741051913(17)291028(10)HUJY0302 (01)00801741051913(17)291028(10)HUJY1095 (01)00801741051913(17)291128(10)HUJZ1332 (01)00801741051913(17)291128(10)HUJZ1336 (01)00801741051913(17)291128(10)HUJZ1339 (01)00801741051913(17)291128(10)HUJZ1340 (01)00801741051913(17)300328(10)HUKR1349 (01)00801741051913(17)300328(10)HUKR1364 (01)00801741051913(17)300428(10)HUKS0784 (01)00801741051913(17)300428(10)HUKS0806 (01)00801741051913(17)300428(10)HUKS1217 (01)00801741051913(17)230428(10)HUKS1218 (01)00801741051913(17)300628(10)HUKU0259 (01)00801741051913(17)300628(10)HUKU0351 (01)00801741051913(17)300728(10)HUKV0349 (01)00801741051913(17)300728(10)HUKV0567 (01)00801741051913(17)300828(10)HUKW0534 Lot Number: HUJY0302 HUJY1095 HUJZ1332 HUJZ1336 HUJZ1339 HUJZ1340 HUKR1349 HUKR1364 HUKS0784 HUKS0806 HUKS1217 HUKS1218 HUKU0259 HUKU0351 HUKV0349 HUKV0567 HUKW0534 Product Name:Hickman 12.5 Fr Triple-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0600650CE UDI code: (01)00801741051975(17)300228(10)HUKQ0535 (01)00801741051975(17)300428(10)HUKS1474 (01)00801741051975(17)300428(10)HUKS1483 (01)00801741051975(17)300628(10)HUKU0313 (01)00801741051975(17)300628(10)HUKU0317 (01)00801741051975(17)300628(10)HUKU0328 (01)00801741051975(17)300728(10)HUKV0354 (01)00801741051975(17)300728(10)HUKV0371 (01)00801741051975(17)300828(10)HUKW0171 (01)00801741051975(17)300828(10)HUKW1283 Lot Number: HUKQ0535 HUKS1474 HUKS1483 HUKU0313 HUKU0317 HUKU0328 HUKV0354 HUKV0371 HUKW0171 HUKW1283 Product Name:Hickman 10 Fr Triple-Lumen CV Catheter, Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff Product Catalog Number: 0606560CE UDI code: (01)00801741052019(17)290828(10)HUKR1370 (01)00801741052019(17)290928(10)HUKS1219 (01)00801741052019(17)300128(10)HUKU0255 (01)00801741052019(17)290928(10)HUKU0324 (01)00801741052019(17)300128(10)HUKV0578 (01)00801741052019(17)300128(10)HUKW0172 (01)00801741052019(17)300128(10)HUKW1315 Lot Number: HUKR1370 HUKS1219 HUKU0255 HUKU0324 HUKV0578 HUKW0172 HUKW1315
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Hickman 0600570Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall1 death182 reports
- Broviac 0600540Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall113 reports
- Hickman 0600650Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall82 reports
- Hickman 0606560Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall80 reports
- Hickman 0600560Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall53 reports
- Hickman 0600620Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall40 reports
- Broviac 0600060Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall1 death33 reports
- Hickman 0600310Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall8 reports
- Broviac 0600120Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall6 reports
- Broviac 0600100Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall4 reports
- Hickman 0600350Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall1 report
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.