Skip to content

Broviac 0600100Recalls & reported problems

Bard Access Systems, Inc. · Broviac 6.6 Fr Single-Lumen CV Catheter Cutdown Tray with SureCuff Tissue Ingrowth Cuff · Class 2 · Implant

All models of Broviac (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741051708), received between Jul 19, 2018 and Nov 3, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Failure to Infuse” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2025Open, Classified

    Due to damage to outer tray that can potentially compromise the sterile barrier

    Z-1367-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Infuse1
  • Fluid/Blood Leak1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Loss of or Failure to Bond1
  • Material Puncture/Hole1
  • Migration1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 3, 2025Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Migration; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Migration

    ON (B)(6) 2025, A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE HICKMAN CATHETER. ON (B)(6) 2025, POST A PROCEDURE AN ACCIDENTAL DISLODGEMENT OF THE BROVIAC CV CATHETER, SINGLE-LUMEN, 6.6F OCCURRED. UPON EXAMINING THE CUFF SECTION OF THE DISLODGED CATHETER, IT APPEARS THAT ALMOST NO INGROWTH INTO THE SUBCUTANEOUS TISSUE WAS OBSERVED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23447500 ↗
  • MalfunctionFeb 1, 2021Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    IT WAS REPORTED THAT APPROXIMATELY SIX MONTH POST CATHETER PLACEMENT, THE CATHETER ALLEGEDLY HAD A HOLE IN THE THICK PART AND FLUID LEAKED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11259040 ↗
  • MalfunctionFeb 1, 2021Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT DURING A CATHETER PLACEMENT, THE CATHETER ALLEGEDLY SHORTENED OVER THE OUTER SECTION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11258543 ↗
  • MalfunctionJul 19, 2018Failure to Infuse; Failure to Infuse

    IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO FLUSH THE CATHETER PRIOR TO INSERTION AND RESISTANCE WAS FELT IN THE LUMEN AND THE CATHETER COULD NOT BE FLUSHED. THE PROCEDURE WAS COMPLETED USING A NEW DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7702849 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page