Broviac 0600540Recalls & reported problems
Bard Access Systems, Inc. · Broviac 6.6 Fr Single-Lumen CV Catheter Peel-Apart Introducer Kit with SureCuff Tissue Ingrowth Cuff · Class 2 · Implant
All models of Broviac (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard Broviac catheter is a long-term central venous line. Reports describe the catheter clogging, cracking or fracturing, leaking fluid, and sometimes being hard to flush or aspirate, with some catheters removed. There are 111 reports, including 14 injuries and no deaths. There is 1 open recall, which began in December 2025 over outer trays damaged in a way that could compromise the sterile packaging.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 111
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 111 adverse event reports that list this exact device (DI 00801741051821), received between Nov 10, 2015 and Jan 9, 2026 — classified as 14 injuries, 97 malfunctions.
The most frequently reported specific device problem is “Fracture” (37 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2025Class IIOpen, Classified
Due to damage to outer tray that can potentially compromise the sterile barrier
Z-1367-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture37
- Fluid/Blood Leak24
- Hole In Material9
- Stretched8
- Device Dislodged or Dislocated6
- Illegible Information6
- Material Erosion6
- Migration6
Patient problems
- No Clinical Signs, Symptoms or Conditions61
- No Known Impact Or Consequence To Patient16
- No Consequences Or Impact To Patient15
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Unspecified Infection3
- Swelling2
- Bacterial Infection1
- Burning Sensation1
Recent adverse event reports
The 25 most recent of 111 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2026Difficult to Flush; Suction Problem; Obstruction of Flow
A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE HICKMAN/LEONARD/BROVIAC CENTRAL VENOUS CATHETER. ON (B)(6) 2025, POST THE PROCEDURE CLOGS WAS NOTED WITHIN 02 TO 03 WEEKS WITH EPO-PLASTINOL. ADDITIONALLY, THERE WAS AN ISSUE WITH FLUSHING AND ASPIRATION. CATHETER WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24018429 ↗MalfunctionDec 2, 2025Nonstandard Device; Failure to Advance; Deformation Due to Compressive Stress; Nonstandard Device; Failure to Advance; Deformation Due to Compressive Stress
ON (B)(6) 2025, A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE BROVIAC CV CATHETER, SINGLE-LUMEN, 6.6F. DURING THE PROCEDURE IT WAS NOTED THAT THE TIP OF THE DILATOR NOT GLIDING ON GUIDEWIRE AND DILATOR SHEATH TIP ALLEGEDLY BROADENED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23691070 ↗MalfunctionDec 2, 2025Nonstandard Device; Failure to Advance; Deformation Due to Compressive Stress; Nonstandard Device; Failure to Advance; Deformation Due to Compressive Stress
ON (B)(6) 2025, A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE BROVIAC CV CATHETER, SINGLE-LUMEN, 6.6F. DURING THE PROCEDURE IT WAS NOTED THAT THE TIP OF THE DILATOR NOT GLIDING ON GUIDEWIRE AND DILATOR SHEATH TIP ALLEGEDLY BROADENED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23691065 ↗MalfunctionOct 15, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
ON (B)(6) 2025, A PATIENT WAS PREPPED FOR A CATHETER PLACEMENT PROCEDURE USING THE BROVIAC CV CATHETER. PRIOR TO PROCEDURE THE PACKAGE WAS ALLEGEDLY FOUND WITH FOREIGN OBJECTS. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23276255 ↗MalfunctionAug 19, 2025Fracture; Fracture
ON (B)(6) 2025, A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE HICKMAN S/L CATHETERS. ON (B)(6) 2025, IT WAS REPORTED THAT POST PLACEMENT PROCEDURE, THERE WAS A CRACK IN THE DEVICE. THE BROKEN CATHETER WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22837403 ↗MalfunctionJul 24, 2025Loss of or Failure to Bond; Expulsion; Loss of or Failure to Bond; Expulsion
ON (B)(6) 2024, A PATIENT UNDERWENT THE MULTIPLE HICKMAN LINE WAS INSERTED. ON AUG 23,2024, IT WAS REPORTED THAT POST PLACEMENT PROCEDURE THE CUFF WAS ALLEGEDLY NOTICED TO BE AT EXIT SITE. THERE WAS NO PATIENT REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22584417 ↗MalfunctionJul 23, 2025Loss of or Failure to Bond; Expulsion; Migration; Loss of or Failure to Bond; Expulsion; Migration
ON (B)(6) 2024, A PATIENT UNDERWENT THE MULTIPLE HICKMAN LINE WAS INSERTED. ON (B)(6) 2024, IT WAS REPORTED THAT POST PLACEMENT PROCEDURE THE LINE WAS ALLEGEDLY RE-X RAYED. REPORTEDLY THE LINE FOUND TO HAVE MIGRATED OUT TIP LYING IN BRACHIOCEPHALIC VEIN DESPITE ANCHORING SUTURE AT EXIT SITE. THE LINE REPLACEMENT WAS REQUIRED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions
MAUDE report 22573961 ↗InjuryJul 23, 2025Loss of or Failure to Bond; Expulsion; Migration; Loss of or Failure to Bond; Expulsion; Migration
ON (B)(6) 2024, A PATIENT UNDERWENT THE MULTIPLE HICKMAN LINE WAS INSERTED. ON (B)(6) 2025, IT WAS REPORTED THAT POST PLACEMENT PROCEDURE THE CUFF WAS ALLEGEDLY NOTICED TO BE MIGRATED 2CM. IT WAS FURTHER REPORTED THAT THE ANCHOR SUTURE WAS REMOVED. THE LINE REPLACEMENT WAS REQUIRED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22573759 ↗MalfunctionJun 5, 2025Burst Container or Vessel; Fracture; Burst Container or Vessel; Fracture
A PATIENT UNDERWENT A CHRONIC CATHETER PLACEMENT PROCEDURE USING THE BROVIAC CV CATHETER APPROXIMATELY SIX MONTHS AGO. ON (B)(6) 2025, THE CATHETER ALLEGEDLY BROKE. REMOVAL, REPOSITIONING WAS DONE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22147390 ↗MalfunctionMar 4, 2025Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE HUB ALLEGEDLY LEAKED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21512124 ↗MalfunctionFeb 18, 2025Stretched; Material Protrusion/Extrusion; Stretched; Material Protrusion/Extrusion
IT WAS REPORTED THAT SOMETIMES POST A CHRONIC CATHETER PLACEMENT, THE DEVICE ALLEGEDLY HAD BALLOONING IN THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21399084 ↗MalfunctionJan 13, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE CATHETER'S CLAMP WAS ALLEGEDLY SLIPPED AND UNCLAMPED ON ITS OWN. IT WAS FURTHER REPORTED THAT A HEMOSTAT FORCEPS WAS USED TO CLAMP THE LINE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21137874 ↗MalfunctionJan 11, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE CATHETER'S CLAMP WAS ALLEGEDLY SLIPPED AND UNCLAMPED ON ITS OWN. IT WAS FURTHER REPORTED THAT A HEMOSTAT FORCEPS WAS USED TO CLAMP THE LINE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21133081 ↗MalfunctionOct 25, 2024Expulsion; Loosening of Implant Not Related to Bone-Ingrowth; Expulsion; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT SOMETIMES POST A CHRONIC CATHETER PLACEMENT, THE LINE ALLEGEDLY MIGRATED. IT WAS FURTHER REPORTED THAT AFTER REINSERTION, SUTURE WAS INTACT BUT THE LINE HAD MIGRATED THROUGH THE STITCH DUE TO LOOSENING. FURTHERMORE, THE CUFF WAS NOT DETACHED AND THE ENTIRE DEVICE HAD PARTIALLY MIGRATED OUT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20531141 ↗InjuryOct 25, 2024Migration; Migration
IT WAS REPORTED THAT TWENTY FOUR DAYS POST A CHRONIC CATHETER PLACEMENT, THE LINE ALLEGEDLY MIGRATED AND REQUIRED REINSERTION. REPORTEDLY, THE CATHETER WAS REPLACED TWICE. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20531039 ↗MalfunctionSep 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SOMETIMES POST CHRONIC CATHETER PLACEMENT, THE DEVICE ALLEGEDLY COULD HAVE GAVE THE PATIENT AN ALLERGIC REACTION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 20229080 ↗MalfunctionAug 22, 2024Fluid/Blood Leak; Fracture; Material Integrity Problem; Fluid/Blood Leak; Fracture; Material Integrity Problem
IT WAS REPORTED THAT SOMETIME POST A PORT PLACEMENT, DAMAGE WAS ALLEGEDLY OBSERVED ON THE EXTERNAL PART OF THE CATHETER. FLUID LEAKAGE WAS CONFIRMED BY FLUSHING WITH SALINE BEFORE ADMINISTRATION OF THE DRUG. REPORTEDLY, THE PORT WAS REMOVED AND REPLACED. THERE WAS NO REPORTED SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20030926 ↗MalfunctionAug 10, 2024Material Discolored; Stretched; Material Protrusion/Extrusion; Material Discolored; Stretched; Material Protrusion/Extrusion
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT VIA THE LEFT SAPHENOUS VEIN, THE BUBBLE WAS ALLEGEDLY FOUND TO BE PRESENT DISTALLY TO THE CLAMP. IT WAS FURTHER REPORTED THAT THE FLUID WAS ASPIRATED WHILE CLAMPED AND BUBBLE DE-INFLATED. FURTHERMORE, THE BLOOD RETURN WAS PRESENT BUT PINK DISCOLORATION WAS ALLEGED FOUND IN THE THIN AREA OF LINE. MOREOVER, THE LINE WAS FLUSHED AND WHEN EXTRA PRESSURE ADDED, NEW BUBBLE ALLEGEDLY FORMED IN THE SAME AREA. EVIDENTLY, ABLE TO REMOVE THE FLUID AGAIN AND FLUSHED SLOWLY WITHOUT FORCE. REPORTEDLY, THE CATHETER WAS REPAIRED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19957671 ↗MalfunctionMay 2, 2024Fracture; Difficult to Insert; Difficult to Remove; Material Separation; Stretched; Unraveled Material; Improper or Incorrect Procedure or Method; Difficult to Advance; Material Integrity Problem; Physical Resistance/Sticking; Fracture; Difficult to Insert; Difficult to Remove; Material Separation; Stretched; Unraveled Material; Improper or Incorrect Procedure or Method; Difficult to Advance; Material Integrity Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, WHEN THE GUIDEWIRE WAS INSERTED, RESISTANCE WAS ALLEGEDLY FELT AND THE GUIDEWIRE WAS ALLEGEDLY DIFFICULT TO BE ADVANCED. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE WAS ALLEGEDLY DIFFICULT TO BE REMOVED. FURTHERMORE, WHEN PULLED BACK, THE GUIDEWIRE WAS ALLEGEDLY FOUND TO BE DAMAGED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19228261 ↗MalfunctionDec 29, 2023Fluid/Blood Leak; Fracture; Material Puncture/Hole; Fluid/Blood Leak; Fracture; Material Puncture/Hole
IT WAS REPORTED THAT APPROXIMATELY A COUPLE OF YEARS POST A CHRONIC CATHETER PLACEMENT, THE CATHETER ALLEGEDLY HAD A PINHOLE FRACTURE. IT WAS FURTHER REPORTED THAT THE INFUSION WAS ALLEGEDLY OOZING OUT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18417159 ↗MalfunctionNov 27, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT SOMETIME POST A CHRONIC CATHETER PLACEMENT, THE CATHETER ALLEGEDLY HAD PINHOLE A FEW MILLIMETERS FROM THE HUB. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18207991 ↗MalfunctionNov 14, 2023Separation Failure; Separation Failure
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, THE INTRODUCER ALLEGEDLY DID NOT OPEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18129549 ↗MalfunctionSep 1, 2023Difficult to Flush; Suction Problem; Difficult to Flush; Suction Problem
IT WAS REPORTED THAT NINE DAYS POST CHRONIC CATHETER PLACEMENT, THE CATHETER ALLEGEDLY DOES NOT HAVE BACK FLOW. IT WAS FURTHER REPORTED THAT THE DEVICE ALLEGEDLY HAD DIFFICULTY IN FLUSHING. REPORTEDLY, THE CATHETER WAS REMOVED AND REPLACED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17668894 ↗MalfunctionAug 1, 2023Defective Component; Deformation Due to Compressive Stress; Defective Component; Deformation Due to Compressive Stress
IT WAS REPORTED THAT DURING A CHRONIC CATHETER PLACEMENT PROCEDURE, THE PEEL AWAY SHEATH WAS ALLEGEDLY TOO SMALL FOR THE CATHETER TO GO THROUGH. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17434078 ↗MalfunctionJul 13, 2023Fracture; Deformation Due to Compressive Stress; Material Split, Cut or Torn; Fracture; Deformation Due to Compressive Stress; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO A CHRONIC CATHETER PLACEMENT PROCEDURE, TIP OF THE SHEATH ALLEGEDLY SPLIT WHEN INSERTING THE INTRODUCER SHEATH OVER THE WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17316629 ↗
FDA clearance and classification
The FDA regulates product code LJS (catheter,intravascular,therapeutic,long-term greater than 30 days) under 21 CFR 880.5970 as a Class II device. 21 CFR 880.5970 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Broviac 0600540
Has the Broviac 0600540 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1367-2026 (Class II), began on Dec 12, 2025: Due to damage to outer tray that can potentially compromise the sterile barrier. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Broviac 0600540?
The FDA's MAUDE database lists 111 adverse event reports for this device, received between Nov 10, 2015 and Jan 9, 2026. In those reports, the most common specific device problems are “Fracture” (37), “Fluid/Blood Leak” (24) and “Hole In Material” (9); the most common patient problems are “Unspecified Infection” (3), “Swelling” (2) and “Bacterial Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Broviac 0600540?
None of the reports are classified as deaths, 14 as injuries and 97 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Broviac 0600540 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Broviac 0600540, and is it still sold?
The FDA's device database lists Bard Access Systems, Inc. as the company responsible for it (brand Broviac, model 0600540). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,692 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,507 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall398 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls244 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall263 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days276 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page