C. R. Bard, Inc. · Bard® PreVent® Anti-Reflux Filter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741052286), received between Mar 1, 2019 and May 15, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Markings/Labelling Problem3
  • Device Contamination with Chemical or Other Material2
  • Degraded1

Patient problems

  • No Consequences Or Impact To Patient4
  • No Known Impact Or Consequence To Patient4
  • Gastrointestinal Regurgitation1
  • Vomiting1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2024

    A SALEM SUMP NASOGASTRIC TUBE WAS PLACED FOR DECOMPRESSION. THE BARD PREVENT ANTI-REFLUX FILTER WAS PLACED UPSIDE DOWN IN THE AIR VENT. THE TUBE WAS NOT DECOMPRESSING PATIENT'S STOMACH, AND PATIENT WAS ABLE TO VOMIT AROUND TUBE. DURING NURSING PEER REVIEW, IT WAS IDENTIFIED THE PRODUCT'S INSTRUCTIONS ARE NOT CLEAR ON WAY TO INSERT. THE INSTRUCTIONS STATE "FIRMLY SEAT TAPERED END OF ANTI-REFLUX FILTER IN BLUE AIR LUMEN VENT OF NG TUBE." THE REGISTERED NURSES (RNS) INVOLVED AND OTHER RNS ASKED ABOUT THE PLACEMENT OF FILTER FELT THE PACKAGE INSTRUCTIONS LEFT TOO ROOM FOR INTERPRETATION AND IT WAS NOT CLEAR. THE RNS WOULD LIKE MORE DESCRIPTION OF INSERTION BLUE END INTO THE BLUE AIR VENT OR AN ARROW TO SHOW DIRECTION FOR INSERTION. CURRENTLY THE DEVICE HAS A LABEL STATING INFUSE AIR ONLY.

    Patient problems: Gastrointestinal Regurgitation; Vomiting

    MAUDE report 19319890 ↗
  • MalfunctionMar 2, 2019Degraded; Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THE MARKINGS ON THE NG TUBE HAD RUBBED OFF. WITHOUT THE MARKINGS, THE NURSING STAFF COULD NOT DETERMINE WHERE THE NG TUBE WAS INSERTED, OR WHERE IT SHOULD BE LOCATED. ADDITIONALLY, IT WAS ALLEGED THAT RESIDUE REMAINED IN THE FILTERS AFTER USE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8384423 ↗
  • MalfunctionMar 2, 2019Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE MARKINGS ON THE NG TUBE HAD RUBBED OFF. WITHOUT THE MARKINGS, THE NURSING STAFF COULD NOT DETERMINE WHERE THE NG TUBE WAS INSERTED, OR WHERE IT SHOULD BE LOCATED. ADDITIONALLY, IT WAS ALLEGED THAT RESIDUE REMAINED IN THE FILTERS AFTER USE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8384361 ↗
  • MalfunctionMar 2, 2019Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE MARKINGS ON THE NG TUBE HAD RUBBED OFF. WITHOUT THE MARKINGS, THE NURSING STAFF COULD NOT DETERMINE WHERE THE NG TUBE WAS INSERTED, OR WHERE IT SHOULD BE LOCATED. ADDITIONALLY, IT WAS ALLEGED THAT RESIDUE REMAINED IN THE FILTERS AFTER USE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8384358 ↗
  • MalfunctionMar 1, 2019Device Markings/Labelling Problem; Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THE MARKINGS ON THE NGT RUBBED OFF AND WITHOUT THE MARKINGS THE NURSING STAFF COULD NOT DETERMINE WHERE THE NGT WAS INSERTED OR ITS LOCATION. ADDITIONALLY, IT WAS ALLEGED THAT RESIDUE REMAINED IN THE FILTERS AFTER USE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8384097 ↗

Other TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices

All TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices →