Skip to content

Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ EN0046160Recalls & reported problems

C. R. Bard, Inc. · Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741113253), received between May 31, 2018 and Feb 14, 2024 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Nonstandard Device” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 20, 2023Class IIOpen, Classified

    BD (C.R. Bard) has received user complaints reporting that during use of BD Bard NG Sump Tubes with ENFit Connectors, inadequate or reduced suction, decompression, and drainage occurred.

    Z-0799-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Nonstandard Device2
  • Defective Component1
  • Fitting Problem1
  • Incorrect Measurement1
  • Insufficient Device Problem Information1
  • Partial Blockage1
  • Suction Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Exposure to Body Fluids1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 14, 2024Defective Component; Defective Component

    IT WAS REPORTED THAT THE NASOGASTRIC TUBE HAD UNKNOWN ISSUE. (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18702037 ↗
  • MalfunctionNov 27, 2023Incorrect Measurement; Fitting Problem; Incorrect Measurement; Fitting Problem

    IT WAS REPORTED THAT NASOGASTRIC SUMP TUBE AND IT HAS THE ENFIT DEVICE ON IT. WHAT THEY WERE TRYING TO FIGURE OUT WAS HOW TO CONNECT IT SUCTION BECAUSE THE LUER LOCK DECREASED THE DIAMETER OF THE TUBE (EN0046160 AND EN0066000).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18209856 ↗
  • InjuryJun 8, 2023Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT THERE WAS A RECENT INCIDENT WHERE ONE OF THEIR PROVIDERS WAS USING EN0046160 FOR DECOMPRESSION. UNFORTUNATELY, THEY HAD AN EVENT IN WHICH ONE OF THESE BARD TUBES WAS PLACED IN A PATIENT WHO WAS ACTIVELY CODING. THEY WERE UNABLE TO DECOMPRESS THE PATIENT¿S STOMACH AND ENDED UP USING SCISSORS TO CUT THE END OF IT OFF TO CONNECT SUCTION. IN DOING SO, THE CRISIS NURSE ALSO HAD GASTRIC CONTENTS SPRAYED ALL OVER THEM. THEY WERE AWARE THAT THERE WERE ADAPTER OPTIONS BUT STATED THEY COULD NOT DECOMPRESS THROUGH THE SMALL ENFIT HOLE. THEY WOULD ASSUME THIS TUBE WOULD ALSO HAVE PROBLEMS IN OTHER SITUATIONS, SUCH AS BLEEDING OR POISONING SITUATIONS. THIS WAS A MAJOR PATIENT SAFETY EVENT. IF THE PRODUCT WAS UNABLE TO DO WHAT IT WAS SUPPOSED TO DO, THEY HAD A PROBLEM. NO MEDICAL INTERVENTION WAS REPORTED. PER FOLLOW UP VIA EMAIL ON (B)(6) 2023, IT WAS REPORTED THAT THE SAMPLE WAS DISCARDED. THEY DID NOT THINK THE BARD SOLUTION WOULD MEET THE ANESTHESIOLOGISTS¿ NEEDED IN AN EMERGENCY. ALSO STATED THAT THIS TUBE WAS NEEDED DURING A CODE. ULTIMATELY BECAUSE THE STAFF DIDN¿T HAVE THE ADAPTERS, THEY NEEDED TO CUT OFF THE FIXED ENFIT ADAPTER TO UTILIZE THE SUCTION FEATURE. THERE

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Exposure to Body Fluids

    MAUDE report 17086018 ↗
  • MalfunctionMay 8, 2023Partial Blockage; Partial Blockage

    IT WAS REPORTED THAT PROVIDER ATTEMPTED TO PLACE A NASOGASTRIC TUBE (NG TUBE) (THE NEW ENFIT STYLE THAT WE HAVE JUST STARTED TO CARRY) 16 FR OR 18 FR AND HAD THE FIRST 3 NASOGASTRIC TUBES CLOG (BATCH# NGGR2656), WAS SUCCESSFUL ON THE 4TH TUBE. PER FOLLOW-UP VIA EMAIL ON 10-MAY-2023, IT WAS REPORTED THAT THE NEW BD ENFIT COMPATIBLE SUMP TUBES HAD AN ENFIT TIP ON THEM WHICH REQUIRED A SUCTION ADAPTER TO PERFORM SUCTION. THE PORT AND ADAPTER HAD A NARROW CHANNEL AND BRONSON CLINICIANS HAD REPORTED CLOGGING ISSUES (BATCH# NGFW1855). PER FOLLOW UP VIA PHONE ON 02-JUN-2023, CUSTOMER HAS ISSUES WITH TWO LOTS NNGY3160 AND NNGV4485 THE CUSTOMER HAS A SAMPLE TO SEND IN AND THE ONLY PATIENT IMPACT WAS THAT CATHETERS HAD TO BE RE-INSERTED, BUT NO HARM TO ANY PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16886173 ↗
  • MalfunctionSep 29, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE NASOGASTRIC TUBE DID NOT DO WHAT INTENDED IN MEDICAL SETTING. THE ISSUE WITH THE NASOGASTRIC TUBE WAS UNKNOWN. THE PRODUCT CATALOG NUMBERS PROVIDED BY CUSTOMER WERE EN0046100, EN0046120, EN0046140, EN0046160 AND EN0046180.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508528 ↗
  • MalfunctionFeb 11, 2020Nonstandard Device; Therapeutic or Diagnostic Output Failure; Nonstandard Device

    IT WAS REPORTED THAT THE OPENING WAS TOO NARROW TO EFFECTIVELY DRAIN THE STOMACH.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9695650 ↗
  • MalfunctionMay 31, 2018Suction Problem

    IT WAS REPORTED THAT THE NG TUBE DID NOT SUCTION WELL. THE COMPLAINANT REPORTED THAT THE NURSING STAFF HAD TO BREAK THE "INNER MOAT" SECTION OUT OF THE ENFIT CONNECTOR TO IMPROVE THE SUCTION.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7557792 ↗

Other TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices

All TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page