Skip to content

CARDINAL HEALTH 8888266122ERecalls & reported problems

Cardinal Health 200, LLC · Salem Sump Dual Lumen Stomach Tube with ENFit Connection · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10192253012811), received between Jan 24, 2025 and Sep 15, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2025Class IIOpen, Classified

    Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

    Z-0206-2026 · Root cause: Use error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3
  • Difficult to Remove3
  • Separation Failure3

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50774379 ↗
  • MalfunctionNov 20, 2025Break; Break

    THE CUSTOMER REPORTED THE ARV BROKE WHEN TRYING TO REMOVE IT WHEN ORDERED FOR THE NG TO DRAIN AFTER BEING CLAMPED. PER ADDITIONAL INFORMATION PROVIDED ON 17NOV2025, THE NG WAS CLAMPED AND THE WHITE PART OF THE ARV WAS IN THE SUCTION LUMEN, UPON ATTEMPT TO REMOVE IT TO REATTACH TO SUCTION, IT BROKE AT THE WHITE PART OF THE ARV.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23609796 ↗
  • MalfunctionNov 20, 2025Break; Break

    THE CUSTOMER REPORTED THE ARV BROKE WHEN TRYING TO REMOVE IT WHEN ORDERED FOR THE NG TO DRAIN AFTER BEING CLAMPED. PER ADDITIONAL INFORMATION PROVIDED ON 17NOV2025, THE NG WAS CLAMPED AND THE WHITE PART OF THE ARV WAS IN THE SUCTION LUMEN, UPON ATTEMPT TO REMOVE IT TO REATTACH TO SUCTION IT BROKE AT THE BLUE PART OF THE ARV. THE DEVICE WAS REMOVED FROM THE PATIENT AND A NEW ONE HAS TO BE RE-INSERTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23609645 ↗
  • MalfunctionJun 9, 2025Difficult to Remove; Difficult to Remove

    THE CUSTOMER REPORTED THAT THE BLUE AND WHITE PIECE IS UNABLE TO BE REMOVED FROM TUBE. UNABLE TO USE ITEM DUE TO THIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22169283 ↗
  • MalfunctionJun 9, 2025Difficult to Remove; Difficult to Remove

    THE CUSTOMER REPORTED THAT THE CAP WILL NOT COME OFF THE ANTI-REFLUX VALVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168114 ↗
  • MalfunctionJun 9, 2025Difficult to Remove; Difficult to Remove

    THE CUSTOMER REPORTED THAT THE SALEM SUMP ANTI-REFLUX VALVE WAS UNABLE TO BE REMOVED FROM THE BLUE PORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22167498 ↗
  • MalfunctionMay 8, 2025Break; Break

    THE CUSTOMER REPORTED THAT THE BLUE/WHITE FILTERED PLUG END FOR NG TUBE PULLED APART AT THE WHITE END WITH WHITE PORTION STUCK IN TUBE. REMOVED WHITE PORTION WITH USE OF KELLY CLAMP. STILL ABLE TO UTILIZE NG. ADDITIONAL DETAILS ABOUT INCIDENT: WHEN THE FILTER/VALVE HAD BEEN ATTEMPTED TO BE REMOVED FROM THE TUBING IT HAD GOTTEN STUCK. MANUAL REMOVAL WAS UNSUCCESSFUL AND THE WHITE PORTION OF THE FILTER SNAPPED OFF. THE BLUE PORTION WAS ABLE TO BE REMOVED FROM THE TUBING WITH THE ASSIST OF A KELLY CLAMP. THE NG TUBING WAS UNAFFECTED AND DID NOT HAVE TO BE REPLACED, ONLY THE FILTER/VALVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21990404 ↗
  • MalfunctionMar 5, 2025Separation Failure; Separation Failure

    THE CUSTOMER REPORTED THEY CAN'T REMOVE THE ARV FROM THE SALEM SUMP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21526398 ↗
  • MalfunctionJan 24, 2025Separation Failure; Separation Failure

    THE CUSTOMER REPORTED THEY HAD NO LUCK GETTING THE ANTI-REFUX VALVE OFF BY TWISTING, PULLING, ETC. THIS ISSUE OCCURRED STRAIGHT OUT OF THE PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21233174 ↗

Other TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices

All TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page