CARDINAL HEALTH 8888266122ERecalls & reported problems
Cardinal Health 200, LLC · Salem Sump Dual Lumen Stomach Tube with ENFit Connection · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10192253012811), received between Jan 24, 2025 and Sep 15, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 18, 2025Class IIOpen, Classified
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Z-0206-2026 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Break3
- Difficult to Remove3
- Separation Failure3
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Separation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50774379 ↗MalfunctionNov 20, 2025Break; Break
THE CUSTOMER REPORTED THE ARV BROKE WHEN TRYING TO REMOVE IT WHEN ORDERED FOR THE NG TO DRAIN AFTER BEING CLAMPED. PER ADDITIONAL INFORMATION PROVIDED ON 17NOV2025, THE NG WAS CLAMPED AND THE WHITE PART OF THE ARV WAS IN THE SUCTION LUMEN, UPON ATTEMPT TO REMOVE IT TO REATTACH TO SUCTION, IT BROKE AT THE WHITE PART OF THE ARV.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609796 ↗MalfunctionNov 20, 2025Break; Break
THE CUSTOMER REPORTED THE ARV BROKE WHEN TRYING TO REMOVE IT WHEN ORDERED FOR THE NG TO DRAIN AFTER BEING CLAMPED. PER ADDITIONAL INFORMATION PROVIDED ON 17NOV2025, THE NG WAS CLAMPED AND THE WHITE PART OF THE ARV WAS IN THE SUCTION LUMEN, UPON ATTEMPT TO REMOVE IT TO REATTACH TO SUCTION IT BROKE AT THE BLUE PART OF THE ARV. THE DEVICE WAS REMOVED FROM THE PATIENT AND A NEW ONE HAS TO BE RE-INSERTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609645 ↗MalfunctionJun 9, 2025Difficult to Remove; Difficult to Remove
THE CUSTOMER REPORTED THAT THE BLUE AND WHITE PIECE IS UNABLE TO BE REMOVED FROM TUBE. UNABLE TO USE ITEM DUE TO THIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22169283 ↗MalfunctionJun 9, 2025Difficult to Remove; Difficult to Remove
THE CUSTOMER REPORTED THAT THE CAP WILL NOT COME OFF THE ANTI-REFLUX VALVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22168114 ↗MalfunctionJun 9, 2025Difficult to Remove; Difficult to Remove
THE CUSTOMER REPORTED THAT THE SALEM SUMP ANTI-REFLUX VALVE WAS UNABLE TO BE REMOVED FROM THE BLUE PORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167498 ↗MalfunctionMay 8, 2025Break; Break
THE CUSTOMER REPORTED THAT THE BLUE/WHITE FILTERED PLUG END FOR NG TUBE PULLED APART AT THE WHITE END WITH WHITE PORTION STUCK IN TUBE. REMOVED WHITE PORTION WITH USE OF KELLY CLAMP. STILL ABLE TO UTILIZE NG. ADDITIONAL DETAILS ABOUT INCIDENT: WHEN THE FILTER/VALVE HAD BEEN ATTEMPTED TO BE REMOVED FROM THE TUBING IT HAD GOTTEN STUCK. MANUAL REMOVAL WAS UNSUCCESSFUL AND THE WHITE PORTION OF THE FILTER SNAPPED OFF. THE BLUE PORTION WAS ABLE TO BE REMOVED FROM THE TUBING WITH THE ASSIST OF A KELLY CLAMP. THE NG TUBING WAS UNAFFECTED AND DID NOT HAVE TO BE REPLACED, ONLY THE FILTER/VALVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21990404 ↗MalfunctionMar 5, 2025Separation Failure; Separation Failure
THE CUSTOMER REPORTED THEY CAN'T REMOVE THE ARV FROM THE SALEM SUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21526398 ↗MalfunctionJan 24, 2025Separation Failure; Separation Failure
THE CUSTOMER REPORTED THEY HAD NO LUCK GETTING THE ANTI-REFUX VALVE OFF BY TWISTING, PULLING, ETC. THIS ISSUE OCCURRED STRAIGHT OUT OF THE PACKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21233174 ↗
Other TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices
- Salem Sump 8888266114Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION2 recalls0 reports
- CARDINAL HEALTH 8888266130ECardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall7 reports
- Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ EN0046160C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall7 reports
- Salem Sump 8888266148Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall5 reports
- Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ EN0046180C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall5 reports
- Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter 0046160C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION54 reports
- Bard® Nasogastric Sump Tube and PreVent® Filter and Lopez Valve® with ENFit™ EN0056140C. R. Bard, Inc. · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall4 reports
- Salem Sump 266148Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page