Nezhat-Dorsey 5202770Recalls & reported problems
Davol Inc. · Nezhat-Dorsey SmokEvac Trumpet Valve, with 5mmx33cm Probe Tip with Holes, Suction Tubing, and Dual Spike Adapter (Series 4A) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741065392), received between Nov 21, 2022 and Nov 21, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2022Break; Break
AS REPORTED, DURING LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2022, THREE BARD/DAVOL NEZHAT-DORSEY SMOKEVAC IRRIGATION TUBING SETS WERE USED WITH A MANUAL BULB PRESSURE BAG SYSTEM. AS REPORTED, THE PRESS ATTACHMENT FOR THE LAPAROSCOPIC TIP CRACKED/BROKE WITH INITIAL IRRIGATION FLOW ON ALL THREE DEVICES. AS REPORTED, THE PROCEDURE WAS COMPLETED BY USING A REUSABLE SCREW-IN LAPAROSCOPIC SUCTION IRRIGATION TIP ON FINAL TRUMPET, WITH PRESS FITTING REMOVED. THERE WAS NO REPORTED PATIENT INJURY/HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15835503 ↗MalfunctionNov 21, 2022Break; Break
AS REPORTED, DURING LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2022, THREE BARD/DAVOL NEZHAT-DORSEY SMOKEVAC IRRIGATION TUBING SETS WERE USED WITH A MANUAL BULB PRESSURE BAG SYSTEM. AS REPORTED, THE PRESS ATTACHMENT FOR THE LAPAROSCOPIC TIP CRACKED/BROKE WITH INITIAL IRRIGATION FLOW ON ALL THREE DEVICES. AS REPORTED, THE PROCEDURE WAS COMPLETED BY USING A REUSABLE SCREW-IN LAPAROSCOPIC SUCTION IRRIGATION TIP ON FINAL TRUMPET, WITH PRESS FITTING REMOVED. THERE WAS NO REPORTED PATIENT INJURY/HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15835409 ↗MalfunctionNov 21, 2022Break; Break
AS REPORTED, DURING LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2022, THREE BARD/DAVOL NEZHAT-DORSEY SMOKEVAC IRRIGATION TUBING SETS WERE USED WITH A MANUAL BULB PRESSURE BAG SYSTEM. AS REPORTED, THE PRESS ATTACHMENT FOR THE LAPAROSCOPIC TIP CRACKED/BROKE WITH INITIAL IRRIGATION FLOW ON ALL THREE DEVICES. AS REPORTED, THE PROCEDURE WAS COMPLETED BY USING A REUSABLE SCREW-IN LAPAROSCOPIC SUCTION IRRIGATION TIP ON FINAL TRUMPET, WITH PRESS FITTING REMOVED. THERE WAS NO REPORTED PATIENT INJURY/HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15835396 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page