Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00801741069703), received between Jun 21, 2017 and Jul 10, 2019 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect Measurement5
- Material Integrity Problem3
- Material Twisted/Bent2
- Short Fill2
- Component Missing1
- Device Damaged Prior to Use1
- Difficult to Remove1
- Filling Problem1
Patient problems
- No Known Impact Or Consequence To Patient8
- No Consequences Or Impact To Patient7
- Discomfort2
- No Patient Involvement2
- Pain2
- Burning Sensation1
- Irritation1
- No Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 10, 2019Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED AFTER INVESTIGATION OF A SAMPLE RECEIVED THAT THE CATHETER SAMPLES OUTER DIAMETER WAS MEASURED, AND WERE FOUND TO BE LOW OUT OF SPECIFICATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8777871 ↗MalfunctionJun 24, 2019Short Fill; Material Rupture; Short Fill; Moisture or Humidity Problem
IT WAS REPORTED THAT THE BURST PACKETS DID NOT HAVE ENOUGH SOLUTION TO LUBRICATE THE CATHETERS.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8727922 ↗MalfunctionMay 14, 2019Difficult to Remove
IT WAS REPORTED THAT THE CATHETER WAS DIFFICULT TO REMOVE. THE PATIENT ADVISED THAT THE LAST TWO INCHES OF THE CATHETER TOOK A LOT OF PRESSURE TO REMOVE. NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: Discomfort; No Known Impact Or Consequence To Patient; Pain
MAUDE report 8608023 ↗MalfunctionFeb 26, 2019Nonstandard Device; Material Integrity Problem; Material Twisted/Bent; Nonstandard Device
IT WAS REPORTED THAT THE MAGIC 3 CATHETER WAS THIN, AND CAUSED THE CATHETER TO BEND UPON INSERTION.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 8371796 ↗MalfunctionJan 31, 2019Component Missing
IT WAS REPORTED THAT THE WATER PACKET WAS MISSING FROM THE MAGIC 3 CATHETER PACKAGE.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 8293492 ↗MalfunctionSep 17, 2018Incorrect Measurement; Material Integrity Problem; Incorrect Measurement
IT WAS REPORTED THAT THE DIAMETER OF THE CATHETER SEEMED LARGER THAN USUAL. THE PATIENT ALLEGEDLY EXPERIENCED DISCOMFORT DURING USE. NO MEDICAL INTERVENTION REPORTED.
Patient problems: Discomfort
MAUDE report 7882549 ↗MalfunctionSep 6, 2018Incorrect Measurement; Inadequacy of Device Shape and/or Size; Incorrect Measurement
IT WAS REPORTED THAT THE CATHETER DIAMETER WAS LARGER THAN 12FR, AT LEAST 14FR.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7852643 ↗MalfunctionAug 6, 2018Filling Problem; Short Fill; Short Fill
IT WAS REPORTED THAT THERE WAS NO WATER IN THE BURST PACKET.
Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 7754200 ↗MalfunctionJul 19, 2018Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement
IT WAS REPORTED THAT THE CATHETER WAS THICKER THAN USUAL.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7704128 ↗MalfunctionJul 18, 2018Device Damaged Prior to Use; Positioning Problem; Device Damaged Prior to Use
IT WAS REPORTED THAT THE SURFACE OF THE MAGIC 3 CATHETER WAS ROUGH, WHICH ALLEGEDLY MADE THE CATHETER DIFFICULT TO INSERT AND IRRITATED THE URETHRA.
Patient problems: Irritation
MAUDE report 7698381 ↗MalfunctionJul 18, 2018Product Quality Problem; Material Integrity Problem; Product Quality Problem
IT WAS REPORTED THAT THE MAGIC 3 CATHETER DID NOT HAVE ENOUGH LUBRICATION.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7697966 ↗MalfunctionJul 5, 2018Incorrect Measurement; Use of Device Problem; Incorrect Measurement
IT WAS REPORTED THAT THE CATHETER WAS 14FR INSTEAD OF 12FR.
Patient problems: No Consequences Or Impact To Patient; No Information
MAUDE report 7664471 ↗InjuryJun 21, 2017Patient-Device Incompatibility
IT WAS REPORTED THAT THE PATIENT ALLEGEDLY DEVELOPED AN URINARY TRACT INFECTION AFTER USE. ALLEGEDLY, THERE WAS BURNING AND PAIN DURING USE.
Patient problems: Burning Sensation; Pain; Urinary Tract Infection
MAUDE report 6658968 ↗
Other CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES devices
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119314C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1 recall3,889 reports
- Bardex® I.C. Foley Catheter 0165SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,994 reports
- Bardex® I.C. Foley Catheter 0165SI16C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES1,052 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter 1758SI14C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES778 reports
- Bardex® I.C. Temperature Sensing Foley Catheter 129414MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES436 reports
- Bardex® Lubri-Sil® I.C. All-Silicone Temp Sensing Foley Catheter 119316MC. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES231 reports
- Bardex® Lubri-Sil® I.C. Temperature-Sensing All-Silicone Foley Catheter 119316C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES198 reports
- Bardex® I.C. Foley Catheter 0165SI18C. R. Bard, Inc. · CATHETER, UROLOGICAL (ANTIMICROBIAL) AND ACCESSORIES181 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page