TRUE DILATATION 0224512Recalls & reported problems
Bard Peripheral Vascular, Inc. · TRUE™ DILATATION Balloon Valvuloplasty Catheter, 22 mm x 4.5 cm,110 cm catheter · Class 2
All models of TRUE DILATATION (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Bard True Dilatation balloon catheter is used to widen a narrowed aortic valve. The 23 reports that name its identifier mostly describe balloons leaking or tearing and trouble deflating or pulling the device back out, sometimes requiring another balloon to finish the procedure; they mention no deaths and 3 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00801741091001), received between Jan 12, 2017 and May 13, 2026 — classified as 3 injuries, 20 malfunctions.
The most frequently reported specific device problem is “Aspiration Issue” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Aspiration Issue7
- Leak/Splash5
- Material Rupture5
- Retraction Problem5
- Deflation Problem2
- Difficult to Insert2
- Break1
- Crack1
Patient problems
- No Consequences Or Impact To Patient10
- No Patient Involvement7
- No Clinical Signs, Symptoms or Conditions4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Injury1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026Material Rupture
ON (B)(6) 2026, A PATIENT UNDERWENT A VALVULOPLASTY PROCEDURE USING THE TRUE BALLOON DILATATION CATHETER. DURING THE PROCEDURE, THE PROXIMAL PART OF THE BALLOON ALLEGEDLY LEAKED DUE TO PINHOLE RUPTURE AT 4 ATM. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25157638 ↗MalfunctionFeb 18, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING PREPARATION OF A VALVULOPLASTY PROCEDURE, AIR WAS VENTING IN CONTINUOUSLY RESULTING IN LEAKAGE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21399410 ↗MalfunctionDec 27, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT PRIOR TO VALVULOPLASTY PROCEDURE, ALLEGEDLY THERE WAS LEAKAGE IN THE CONNECTION BETWEEN THE HUB AND THE SHAFT. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21024649 ↗MalfunctionNov 12, 2024Deflation Problem; Deflation Problem
IT WAS REPORTED THAT PRIOR TO AN ANGIOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY COULD NOT BE DEFLATED. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20664293 ↗InjuryAug 30, 2022Retraction Problem; Material Rupture; Detachment of Device or Device Component; Retraction Problem; Material Rupture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING VALVULOPLASTY PROCEDURE, THE PTA BALLOON ALLEGEDLY RUPTURED AND DETACHED FROM SHAFT. IT WAS FURTHER REPORTED THAT MULTIPLE ATTEMPTS MADE TO REMOVE DETACHED MATERIAL. REPORTEDLY THE DETACHED SEGMENT REMAINED IN THE RIGHT EXTERNAL ILIAC ARTERY CAUSING ARTERIAL INJURY. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15318244 ↗MalfunctionApr 15, 2020Leak/Splash; Leak/Splash
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATES THAT MODEL 0224512 VALVULOPLASTY BALLOON CATHETER ALLEGEDLY EXPERIENCED A LEAK. THE INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED PATIENT INJURY. THE PATIENT'S AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9965733 ↗InjuryOct 2, 2019Retraction Problem; Retraction Problem
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE AORTIC VALVE, THE PTA BALLOON ALLEGEDLY GOT STUCK INSIDE THE SHEATH DURING RETRACTION. IT WAS FURTHER REPORTED THAT MEDICAL INTERVENTION WAS REQUIRED TO CUT DOWN AND REMOVE THE BALLOON FROM THE PATIENT. REPORTEDLY, IT IS UNKNOWN HOW THE ANGIOPLASTY PROCEDURE WAS COMPLETED. THE STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Injury
MAUDE report 9148363 ↗MalfunctionNov 29, 2018Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING AN AORTIC VALVULOPLASTY PROCEDURE, THE BALLOON VALVULOPLASTY CATHETER ALLEGEDLY GOT CAUGHT ON THE GUIDEWIRE DURING REMOVAL FROM THE PATIENT. THE BALLOON AND GUIDEWIRE WERE REMOVED TOGETHER AS ONE UNIT. IT WAS FURTHER REPORTED THAT AFTER THE PROCEDURE, SIGNIFICANT RESISTANCE WAS FELT IN ORDER TO REMOVE THE BALLOON FROM THE GUIDEWIRE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8118516 ↗MalfunctionNov 14, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT PRIOR TO A VALVULOPLASTY PROCEDURE, THE PTV BALLOON ALLEGEDLY RUPTURED DURING A TEST INFLATION. ANOTHER DEVICE WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8071943 ↗MalfunctionMay 7, 2018Break; Material Rupture; Break; Material Rupture
IT WAS REPORTED THAT THE VALVULOPLASTY BALLOON ALLEGEDLY WOULD NOT HOLD THE PRESSURE THROUGHOUT INFLATION. IT WAS FURTHER REPORTED THAT THE BALLOON WAS REMOVED FROM THE PATIENT AND AN ATTEMPT TO INFLATE IT ON THE BACK TABLE ALLEGEDLY DISCOVERED A PIN HOLE RUPTURE AT THE DISTAL PORTION OF THE BALLOON. REPORTEDLY, THE PROCEDURE WAS COMPLETED WITH ANOTHER VALVULOPLASTY BALLOON. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7492609 ↗MalfunctionApr 3, 2018Aspiration Issue; Aspiration Issue
IT WAS REPORTED THAT DURING PREPARATION FOR A TAVI PROCEDURE, THE VALVULOPLASTY BALLOON ALLEGEDLY COULD NOT BE ASPIRATED WITH A LUER LOCK SYRINGE. THE BALLOON WAS EXCHANGED FOR ANOTHER TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 7394477 ↗MalfunctionApr 3, 2018Aspiration Issue; Aspiration Issue
IT WAS REPORTED THAT DURING PREPARATION FOR A TAVI PROCEDURE, THE VALVULOPLASTY BALLOON ALLEGEDLY COULD NOT BE ASPIRATED WITH A LUER LOCK SYRINGE. THE BALLOON WAS EXCHANGED FOR ANOTHER TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 7392938 ↗InjuryDec 27, 2017Retraction Problem; Retraction Problem
IT WAS REPORTED THAT POST DILATATION OF A 26MM TAVI, DURING RETRACTION THROUGH A 14 FR SHEATH, THE VALVULOPLASTY BALLOON ALLEGEDLY WOULD NOT RETRACT INTO THE SHEATH. IT WAS FURTHER REPORTED THAT THE SHEATH AND CATHETER WERE REMOVED TOGETHER AS A SINGLE UNIT RESULTING IN ALLEGED TRAUMA AT THE FEMORAL ARTERY ACCESS SITE, REQUIRING ADDITIONAL BALLOON ANGIOPLASTY TO TAMPONADE THE RESULTANT LEAK. PATIENT STATUS WAS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7149104 ↗MalfunctionDec 10, 2017Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING PREPARATION FOR A TAVI PROCEDURE, THE VALVULOPLASTY BALLOON ALLEGEDLY COULD NOT BE DEFLATED WITH A SYRINGE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7102611 ↗MalfunctionDec 6, 2017Leak/Splash; Aspiration Issue; Leak/Splash; Aspiration Issue
IT WAS REPORTED THAT DURING AN AORTIC VALVULOPLASTY PROCEDURE AN ALLEGED AIR BUBBLE WAS FOUND INSIDE THE BALLOON DURING INFLATION. THE BALLOON WAS DEFLATED AND REMOVED FROM THE PATIENT. A NEW SYRINGE AND STOPCOCK WAS USED BUT THE SAME ISSUE OCCURRED. A NEW BALLOON WAS PREPPED AND USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7088276 ↗MalfunctionOct 6, 2017Difficult to Insert; Retraction Problem; Difficult to Insert; Retraction Problem
IT WAS REPORTED THAT AFTER A BALLOON AORTIC VALVULOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY HAD DIFFICULTIES PASS THROUGH THE 12FR SHEATH BUT EVENTUALLY WOULD NOT RETRACT THROUGH THE SHEATH. IT WAS FURTHER REPORTED THAT THE VALVULOPLASTY BALLOON CATHETER AND SHEATH WERE REMOVED TOGETHER AS A SINGLE UNIT. PATIENT STATUS IS UNKNOWN AT THIS TIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6921005 ↗MalfunctionOct 6, 2017Difficult to Insert; Retraction Problem; Difficult to Insert; Retraction Problem
IT WAS REPORTED THAT AFTER A BALLOON AORTIC VALVULOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY HAD DIFFICULTIES PASSING THRU THE 12FR SHEATH BUT EVENTUALLY WOULD NOT RETRACT THROUGH THE SHEATH. IT WAS FURTHER REPORTED THAT THE VALVULOPLASTY BALLOON CATHETER AND SHEATH WERE REMOVED TOGETHER AS A SINGLE UNIT. PATIENT STATUS IS UNKNOWN AT THIS TIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6919639 ↗MalfunctionAug 15, 2017Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING A TRANSCATHETER AORTIC VALVE REPLACEMENT THE BALLOON ALLEGEDLY WOULD NOT INFLATE PAST 5 ATM. THE HEALTH CARE PROVIDER DEFLATED AND REMOVED THE BALLOON SUCCESSFULLY FROM THE PATIENT. UPON FURTHER INSPECTION OF THE BALLOON ON THE BACK TABLE, THE HEALTH CARE PROVIDER FOUND A PINHOLE RUPTURE NEAR THE PROXIMAL END OF THE BALLOON. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6795203 ↗MalfunctionJul 10, 2017Aspiration Issue; Aspiration Issue
IT WAS REPORTED THAT THE VALVULOPLASTY BALLOON CATHETER ALLEGEDLY ASPIRATED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6699704 ↗MalfunctionMay 30, 2017Peeled/Delaminated; Aspiration Issue; Peeled/Delaminated; Aspiration Issue
IT WAS REPORTED THAT DURING PREPARATION FOR AN AORTIC VALVULOPLASTY PROCEDURE THE BALLOON CATHETER ALLEGEDLY ASPIRATED AT THE BIFURCATION WHILE PULLING NEGATIVE PRESSURE. ANOTHER BALLOON WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 6599951 ↗MalfunctionMar 21, 2017Aspiration Issue; Aspiration Issue
IT WAS REPORTED THAT DURING PREPARATION FOR AN AORTIC VALVULOPLASTY PROCEDURE, AIR WAS ALLEGEDLY BEING ASPIRATED INTO THE BALLOON. ANOTHER BALLOON WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 6420624 ↗MalfunctionFeb 24, 2017Leak/Splash; Aspiration Issue; Leak/Splash; Aspiration Issue
IT WAS REPORTED THAT THE VALVULOPLASTY BALLOON ALLEGEDLY ASPIRATED WHILE PULLING NEGATIVE PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 6358768 ↗MalfunctionJan 12, 2017Crack; Crack
IT WAS REPORTED THAT UPON REMOVING THE DEVICE FROM THE PROTECTIVE SHEATH AN ALLEGED CRACK WAS OBSERVED ON THE PROXIMAL BALLOON HUB. ANOTHER BALLOON WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO REPORTED PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 6247915 ↗
Other Balloon aortic valvuloplasty devices
- TRUE DILATATION 0204511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty21 reports
- TRUE DILATATION 0244512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty2 deaths19 reports
- TRUE DILATATION 0184511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty12 reports
- TRUE DILATATION 0264513Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty11 reports
- TRUE FLOW TF0203511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty10 reports
- TRUE DILATATION 0214512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty10 reports
- TRUE DILATATION 0234512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty1 death10 reports
- TRUE DILATATION 0254513Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page