TRUE DILATATION 0264513Recalls & reported problems
Bard Peripheral Vascular, Inc. · TRUE™ DILATATION Balloon Valvuloplasty Catheter, 26 mm x 4.5 cm,110 cm catheter · Class 2
All models of TRUE DILATATION (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741096631), received between Mar 27, 2017 and Aug 6, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash5
- Material Rupture5
- Deflation Problem3
- Break2
- Retraction Problem2
- Aspiration Issue1
- Detachment of Device or Device Component1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient2
- No Patient Involvement2
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50156373 ↗MalfunctionJun 18, 2026Break; Leak/Splash
ON (B)(6) 2026, A PATIENT UNDERWENT A VALVULOPLASTY PROCEDURE USING THE TRUE PTA BALLOON. IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON ALLEGEDLY LEAKED AT THE PROXIMAL BALLOON JOINT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25545284 ↗MalfunctionMar 25, 2026Difficult to Remove; Retraction Problem; Detachment of Device or Device Component
ON (B)(6) 2026, A 43-YEAR-OLD MALE PATIENT UNDERWENT A VALVULOPLASTY PROCEDURE USING A TRUE CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, WHEN THEY ATTEMPTED TO REMOVE THE CATHETER, THERE WAS RESISTANCE. IT WAS FURTHER REPORTED THAT THE BALLOON CATHETER ALLEGEDLY GOT SEPARATED FROM ITS INNER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24684741 ↗MalfunctionJan 27, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING A VALVULOPLASTY PROCEDURE, THE BALLOON ALLEGEDLY HAD A PINHOLE RUPTURE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13364906 ↗MalfunctionMay 21, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH FEMORAL ARTERY, THE DEVICE ALLEGEDLY LEAKED AT 2 ATM AT THE PROXIMAL END WHICH WAS CONNECTED TO THE SHAFT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11861463 ↗MalfunctionMay 21, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH FEMORAL ARTERY, THE BALLOON ALLEGEDLY LEAKED AT THE PROXIMAL END WHICH CONNECTS TO THE SHAFT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11861449 ↗MalfunctionMar 1, 2021Inflation Problem; Leak/Splash; Material Rupture; Inflation Problem; Leak/Splash; Material Rupture
IT WAS REPORTED THAT DURING A PREPARATION OF AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY LEAKED. IT WAS FURTHER REPORTED THAT THE BALLOON DID NOT EXPAND PROPERLY. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11395186 ↗MalfunctionJan 9, 2020Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL 0264513 PTA BALLOON DILATATION CATHETER ALLEGEDLY EXPERIENCED MATERIAL RUPTURE. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED PATIENT INJURY. THE PATIENT'S AGE, WEIGHT, AND GENDER WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9572519 ↗InjuryAug 1, 2018Deflation Problem; Retraction Problem; Material Rupture; Deflation Problem; Retraction Problem; Material Rupture
IT WAS REPORTED THAT DURING A VALVULOPLASTY PROCEDURE, THE PTV BALLOON ALLEGEDLY RUPTURED AND COULD NOT BE DEFLATED TO FOLD INTO ITS ORIGINAL SHAPE. IT WAS FURTHER REPORTED THAT THERE WERE ALLEGED DIFFICULTIES RETRACTING THE BALLOON FROM THE PATIENT, WHICH LED TO DETACHMENT OF THE PTV BALLOON IN THE RIGHT COMMON ILIAC ARTERY. A SNARE WAS USED TO ATTEMPT TO REMOVE THE DETACHED BALLOON MATERIAL, BUT WAS UNSUCCESSFUL. THE DETACHED BALLOON MATERIAL WAS LEFT IN THE ARTERY AND A FEMORAL-FEMORAL BYPASS WAS DONE, AS A RESULT OF THE BLOCKED ARTERY. NO FURTHER TREATMENT WAS NEEDED AFTER THE BYPASS PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7741390 ↗MalfunctionJun 29, 2017Break; Deflation Problem; Aspiration Issue; Break; Deflation Problem; Aspiration Issue
IT WAS REPORTED THAT DURING PREPARATION FOR AN AORTIC VALVULOPLASTY PROCEDURE THE BALLOON ALLEGEDLY HAD DIFFICULTY DEFLATING. REPORTEDLY, THE BALLOON ALLEGEDLY WAS ALSO ASPIRATING AIR DURING THE DEFLATION ATTEMPT. ANOTHER BALLOON WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 6678002 ↗MalfunctionMar 27, 2017Deflation Problem; Deflation Problem
IT WAS REPORTED THAT DURING PREPARATION FOR AN AORTIC VALVULOPLASTY PROCEDURE THE BALLOON ALLEGEDLY COULD NOT BE DEFLATED. REPORTEDLY, DURING DEFLATION AN AIR BUBBLE WAS OBSERVED IN THE BALLOON RENDERING THE BALLOON UNUSABLE. ANOTHER BALLOON WAS USED TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Patient Involvement
MAUDE report 6436239 ↗
Other Balloon aortic valvuloplasty devices
- TRUE DILATATION 0224512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty23 reports
- TRUE DILATATION 0204511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty21 reports
- TRUE DILATATION 0244512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty2 deaths19 reports
- TRUE DILATATION 0184511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty12 reports
- TRUE DILATATION 0214512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty10 reports
- TRUE FLOW TF0203511Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty10 reports
- TRUE DILATATION 0234512Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty1 death10 reports
- TRUE DILATATION 0254513Bard Peripheral Vascular, Inc. · Balloon aortic valvuloplasty9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page