Venovo® VENUM14120Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 14 mm x 120 mm (80 cm delivery catheter) · Class 3 · Implant
All models of Venovo (84)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
The Venovo venous stent system is a mesh tube placed in the iliac vein to keep it open. Its 32 adverse event reports mostly describe the stent failing to expand during placement, sometimes stopping partway or staying attached to the delivery catheter. No deaths and no injuries were reported. One recall, started in December 2020 and still open, concerns reports that the stent's end did not immediately expand and remained connected to the delivery system.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00801741103452), received between Nov 10, 2020 and Sep 22, 2026 — classified as 33 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (27 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 14, 2020Class IIOpen, Classified
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
Z-1899-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure27
- Positioning Failure7
- Loss of or Failure to Bond4
- Misfire3
- Difficult to Remove2
- Break1
- Difficult or Delayed Positioning1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions33
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2026Loss of or Failure to Bond; Positioning Failure
On (b)(6) 2026, a patient underwent a stent placement procedure using a Venovo Venous Stent in the iliac vein via femoral access. During the procedure, it was reported that the stent failed to deploy as intended. The procedure was subsequently completed using another Venovo Venous Stent. The patient was reported to be in good condition, and there was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50850444 ↗MalfunctionJun 19, 2026Loss of or Failure to Bond; Positioning Failure
ON (B)(6), 2026, A PATIENT UNDERWENT STENT PLACEMENT PROCEDURE USING VENOVO STENT FOR ILIAC VEIN THROMBOSIS. IT WAS REPORTED THAT THE STENT FAILED TO DEPLOY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25562881 ↗MalfunctionMay 2, 2025Loss of or Failure to Bond; Positioning Failure; Loss of or Failure to Bond; Positioning Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY NOT DEPLOYED UPON CORRECT PLACEMENT AND BOTH THUMBWHEELS SPINNING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21950395 ↗MalfunctionMar 31, 2025Loss of or Failure to Bond; Misfire; Loss of or Failure to Bond; Misfire
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE LEFT ILIAC VEIN VIA LEFT FEMORAL VEIN, THE STENT WAS ALLEGEDLY STOPPED DEPLOYED MIDWAY THROUGH. IT WAS FURTHER REPORTED THAT PHYSICIAN ALLEGEDLY HAD TO OPEN UP DEPLOYMENT HANDLE AND DEPLOY MANUALLY.. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21729067 ↗MalfunctionJul 22, 2024Difficult to Remove; Difficult to Advance; Material Deformation; Difficult to Remove; Difficult to Advance; Material Deformation
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE LEFT COMMON ILIAC VEIN VIA LEFT POPLITEAL VEIN, THERE WAS ALLEGEDLY HARD TIME ADVANCING THE STENT UP FROM THE POP VEIN INTO THE ILIAC VEIN. IT WAS FURTHER REPORTED THAT THE STENT OPENED WITHOUT ANY ISSUES, BUT THE ENDS NEAR THE INFERIOR VENA ALLEGEDLY REMAIN UNFLARED. REPORTEDLY, WHEN THE STENT WAS PULLED OUT OF THE CATHETER, THE STENT ENDS ALLEGEDLY FOLDED ON THE INSIDE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19790767 ↗MalfunctionMay 31, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT THROUGH RIGHT COMMON FEMORAL VEIN, THE DEVICE ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11906192 ↗MalfunctionMay 20, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE, THE PROXIMAL END OF THE STENT ALLEGEDLY FAILED TO EXPAND. IT WAS FURTHER REPORTED THAT STENT WAS EXPANDED AFTER 30 SEC AND DEPLOYED SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11854179 ↗MalfunctionMay 20, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE, THE PROXIMAL END OF THE STENT ALLEGEDLY FAILED TO EXPAND. IT WAS FURTHER REPORTED THAT PHYSICIAN ADVANCED THE SHEATH INTO PROXIMAL END OF STENT TO RELEASE THE SENT. REPORTEDLY STENT WAS DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11854175 ↗MalfunctionMay 14, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO EXPAND. IT WAS FATHER REPORTED THAT STENT WAS EXPANDED AFTER THREE MINUTES. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11824807 ↗MalfunctionMay 11, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING THE STENT DEPLOYMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11805132 ↗MalfunctionApr 21, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY BENT AT TIP AND CRUSHED OVER ITSELF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11699240 ↗MalfunctionApr 19, 2021Positioning Failure; Activation Failure; Positioning Failure; Activation Failure
IT WAS REPORTED THAT DURING VENOVO STENT DEPLOYMENT PROCEDURE THROUGH RIGHT FEMORAL VEIN VIA EXTERNAL ILIAC VEIN, THE STENT ALLEGEDLY HAD EXPANSION ISSUE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11684300 ↗MalfunctionApr 19, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT ON BILATERAL ILIAC VIA BILATERAL POPLITEAL, THE STENT PROXIMAL END ALLEGEDLY FAILED TO EXPAND. IT WAS FURTHER REPORTED THAT STENT GOT DEPLOYED AFTER TWISTING THE DEPLOYMENT SYSTEM. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11683806 ↗MalfunctionApr 5, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE, THE PROXIMAL END OF THE STENT ALLEGEDLY HAD EXPANSION ISSUE. IT WAS FURTHER REPORTED THAT THE PHYSICIAN ROTATED HANDLE AND THE STENT WAS RELEASED AFTER 30 SECONDS. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11612974 ↗MalfunctionMar 31, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT THE PROXIMAL END OF THE STENT ALLEGEDLY HAD EXPANSION ISSUE. THE STENT WAS RELEASED AFTER 30 SECONDS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593829 ↗MalfunctionMar 31, 2021Difficult to Remove; Activation Failure; Difficult to Remove; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT THE PROXIMAL END OF THE STENT ALLEGEDLY HAD EXPANSION ISSUE. IT WAS FURTHER REPORTED THAT THE STENT ALLEGEDLY DIFFICULT TO REMOVE. THE STENT WAS RELEASED AFTER 30 SECONDS. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593781 ↗MalfunctionMar 29, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11580848 ↗MalfunctionMar 29, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT POST STENT DEPLOYMENT THROUGH COMMON FEMORAL ARTERY THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11580846 ↗MalfunctionMar 24, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT THROUGH LEFT EXTERNAL ILIAC VEIN, THE DEVICE ALLEGEDLY FAILED TO EXPAND IN THE PROXIMAL END. THE PROCEDURE WAS COMPLETED BY TWISTING AND TURNING THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11559071 ↗MalfunctionMar 16, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING PORT PLACEMENT THE DEVICE ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11494767 ↗MalfunctionMar 15, 2021Positioning Failure; Activation Failure; Positioning Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT DEPLOYMENT PROCEDURE, THE PROXIMAL END OF THE STENT ALLEGEDLY FAILED TO DEPLOY. IT WAS FURTHER REPORTED THAT THE HANDLE WAS ROTATED AND THE STENT GOT RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11479467 ↗MalfunctionMar 12, 2021Positioning Failure; Activation Failure; Positioning Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT DEPLOYMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO DEPLOY AT THE PROXIMAL END. IT WAS FURTHER REPORTED THAT THE HANDLE WAS ROTATED AND THE STENT GOT RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11473137 ↗MalfunctionMar 4, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT, THE DEVICE ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11408865 ↗MalfunctionFeb 24, 2021Difficult or Delayed Positioning; Positioning Failure; Activation Failure; Difficult or Delayed Positioning; Positioning Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO DEPLOY. IT WAS FURTHER REPORTED THAT AFTER ROTATION, THE STENT GOT RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11371601 ↗MalfunctionFeb 10, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE THROUGH THE ILIO FEMORAL AT THE BILATERAL POPLITEAL, THE DISTAL END OF THE STENT ALLEGEDLY FACED EXPANSION FAILURE. IT WAS FURTHER REPORTED THAT TWISTING THE DEPLOYMENT SYSTEM SUCCESSFULLY DEPLOYED THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11306591 ↗
Other Stent, iliac vein devices
- Abre™ AB9U14080090Medtronic, Inc. · Stent, iliac vein2 recalls9 reports
- Venovo® VENUM16100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall55 reports
- Venovo® VENUM14080Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall36 reports
- Venovo® VENUM14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall31 reports
- Venovo® VENUM14100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall30 reports
- Venovo® VENUM16120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall29 reports
- Venovo® VENUM14160Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall27 reports
- Venovo® VENUL14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page