Skip to content

Venovo® VENUL14060Recalls & reported problems

Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 14 mm x 60 mm (120 cm delivery catheter) · Class 3 · Implant

All models of Venovo (84)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741103490), received between Dec 18, 2020 and Apr 27, 2021 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 14, 2020Class IIOpen, Classified

    Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.

    Z-1899-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure8
  • Positioning Failure2
  • Loss of or Failure to Bond1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2021Positioning Failure; Activation Failure; Positioning Failure; Activation Failure

    IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO DEPLOY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11730581 ↗
  • MalfunctionApr 7, 2021Loss of or Failure to Bond; Positioning Failure; Loss of or Failure to Bond; Positioning Failure

    IT WAS REPORTED, THAT DURING TREATMENT THE DEVICE ALLEGEDLY, FAILED TO DEPLOY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11625966 ↗
  • MalfunctionFeb 4, 2021Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT DEPLOYMENT PROCEDURE THROUGH LEFT COMMON FEMORAL VEIN VIA ILIAC VEIN, THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11277840 ↗
  • MalfunctionJan 4, 2021Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11106942 ↗
  • MalfunctionDec 29, 2020Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY FAILED TO EXPAND FROM THE SYSTEM. THE STENT INTERSTICES FINALLY OPENED AFTER ADVANCING, TWISTING, AND TURNING THE SYSTEM. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11085713 ↗
  • MalfunctionDec 22, 2020Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE DISTAL END OF THE DEVICE DID NOT SELF EXPAND. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11053738 ↗
  • MalfunctionDec 21, 2020Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE DEVICE ALLEGEDLY FAILED TO EXPAND IN THE PROXIMAL END. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11044689 ↗
  • MalfunctionDec 21, 2020Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT, THE DEVICE ALLEGEDLY FAILED TO EXPAND. IT WAS FURTHER REPORTED THAT AFTER MANIPULATION, THE STENT FINALLY DEPLOYED AFTER FIVE MINUTES. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11044686 ↗
  • MalfunctionDec 18, 2020Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11035801 ↗

Other Stent, iliac vein devices

All Stent, iliac vein devices →

Sources and updates