Venovo® VENUM16080Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 16 mm x 80 mm (80 cm delivery catheter) · Class 3 · Implant
All models of Venovo (84)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741103575), received between May 16, 2019 and Oct 28, 2024 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 14, 2020Class IIOpen, Classified
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
Z-1897-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure6
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Improper or Incorrect Procedure or Method1
- Malposition of Device1
- Migration or Expulsion of Device1
- Misfire1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Injury1
- Intimal Dissection1
- No Consequences Or Impact To Patient1
- Thrombosis/Thrombus1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 28, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE EXPIRED DEVICE WAS ALLEGEDLY IMPLANTED IN THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20548177 ↗MalfunctionFeb 23, 2024Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC VEIN, THE PROXIMAL END OF THE STENT ALLEGEDLY DID NOT FULLY DETACH FROM THE DELIVERY SYSTEM. IT WAS FURTHER REPORTED THAT EVEN AFTER ONE MINUTE, THE STENT ALLEGEDLY DID NOT RELEASE. REPORTEDLY THE PHYSICIAN ROTATED THE DEPLOYMENT HANDLE AND COMPLETED THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18765754 ↗MalfunctionJun 12, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE COMMON ILIAC VEIN VIA POPLITEAL ACCESS, THE STENT ALLEGEDLY GOT STUCK ON THE PROXIMAL END. IT WAS FURTHER REPORTED THAT THE STENT WAS RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17105510 ↗InjuryApr 4, 2023Migration or Expulsion of Device; Obstruction of Flow; Migration or Expulsion of Device; Obstruction of Flow
IT WAS REPORTED THAT SOMETIME POST STENT PLACEMENT PROCEDURE, THE STENT HAD ALLEGEDLY SHIFTED AND HAD A CLOT WHICH CAUSED SOME OBSTRUCTION OF FLOW ISSUES. IT WAS FURTHER REPORTED THAT THE CLOT WAS REMOVED AND TWO ADDITIONAL STENTS WERE PLACED TO PREVENT THE PREVIOUS STENT FROM SHIFTING FURTHER AND RESTORING BLOOD FLOW. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Thrombosis/Thrombus
MAUDE report 16676052 ↗MalfunctionFeb 20, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE IN LEFT COMMON ILIAC VEIN VIA FEMORAL ACCESS, THE STENT PROXIMAL END ALLEGEDLY WAS DIFFICULT TO DEPLOY. IT WAS FURTHER REPORTED THAT STENT STUCK TO CUSHION FOR SEVERAL MINUTES AND DIDN'T FULLY COME OFF OF CUSHION WITHOUT TWISTING THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16404075 ↗MalfunctionMar 15, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT PROCEDURE, THE STENT PROXIMAL END ALLEGEDLY HAD DIFFICULT TO DEPLOY. IT WAS FURTHER REPORTED THAT STENT WAS DEPLOYED AFTER COUPLE OF RATIONS BY THE HANDLE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11482052 ↗MalfunctionFeb 24, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT PLACEMENT PROCEDURE, THE PROXIMAL END OF STENT FAILED TO EXPAND. IT WAS FURTHER REPORTED THAT THE PHYSICIAN WAITED AND ROTATED HANDLE AND STENT APPEARED TO RELEASE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11371328 ↗MalfunctionDec 14, 2020Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT THROUGH COMMON ILIAC VESSEL, THE DEVICE ALLEGEDLY FAILED TO EXPAND IN THE DISTAL END. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11000958 ↗MalfunctionNov 19, 2019Break; Use of Device Problem; Misfire; Malposition of Device; Break; Use of Device Problem; Misfire; Malposition of Device
IT WAS REPORTED THAT DURING A STENT PLACEMENT IN THE LEFT COMMON ILIAC VEIN VIA LEFT COMMON FEMORAL APPROACH, THE STENT ALLEGEDLY FAILED TO DEPLOY MORE THAN 50%. REPORTEDLY, A GLOVE WAS CAUGHT IN TRANSITION WHEN THE SECOND CATHETER WENT INTO THE THIRD CATHETER. IT WAS FURTHER REPORTED THAT A HEMOSTAT WAS USED TO OPEN THE DEVICE AND MANUALLY DEPLOY THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9344741 ↗InjuryMay 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH THE RESULTS OF A CLINICAL TRIAL THAT AT 24-MONTH FOLLOW-UP VISIT, INTIMAL HYPERPLASIA WAS IDENTIFIED VIA DUPLEX ULTRASOUND. NO ACTION IS BEING TAKEN AT THIS TIME.
Patient problems: Injury; Intimal Dissection
MAUDE report 8616584 ↗
Other Stent, iliac vein devices
- Abre™ AB9U14080090Medtronic, Inc. · Stent, iliac vein2 recalls9 reports
- Venovo® VENUM16100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall55 reports
- Venovo® VENUM14080Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall36 reports
- Venovo® VENUM14120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall33 reports
- Venovo® VENUM14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall31 reports
- Venovo® VENUM14100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall30 reports
- Venovo® VENUM16120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall29 reports
- Venovo® VENUM14160Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page