Venovo® VENUM16120Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 16 mm x 120 mm (80 cm delivery catheter) · Class 3 · Implant
All models of Venovo (84)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Angiomed Venovo venous stent is a metal mesh tube placed in the iliac vein to keep blood flowing. There are 29 adverse event reports, and most describe the stent failing to open or release properly during placement, sticking to the delivery catheter, or being hard to advance, with procedures sometimes completed using another device. No deaths and one injury are reported. One recall, opened in December 2020, says the stent's upper end sometimes did not expand right away and remained attached to the delivery system; it remains open.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00801741103599), received between Jul 12, 2019 and Oct 25, 2024 — classified as 1 injury, 28 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (23 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 14, 2020Class IIOpen, Classified
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
Z-1897-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure23
- Adverse Event Without Identified Device or Use Problem2
- Break2
- Difficult to Advance2
- Detachment of Device or Device Component1
- Device-Device Incompatibility1
- Entrapment of Device1
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- Hypersensitivity/Allergic reaction1
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT SOMETIME POST AN ANGIOPLASTY PROCEDURE, THE PATIENT ALLEGEDLY EXPERIENCED PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20531705 ↗MalfunctionAug 12, 2024Material Twisted/Bent; Activation Failure; Material Twisted/Bent; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY BECAME STUCK TO THE INNER CATHETER. IT WAS FURTHER REPORTED THAT THERE WAS A SLIGHT PANIC AND DISCOMFORT TO THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19958455 ↗MalfunctionJul 2, 2024Break; Premature Activation; Difficult to Advance; Break; Premature Activation; Difficult to Advance
IT WAS REPORTED THAT REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE RIGHT COMMON FEMORAL VEIN, THE STENT ALLEGEDLY HAD RESISTANCE WHILE ADVANCING AND THE STENT ALLEGEDLY STARTED TO DEPLOY ON ITS OWN. IT WAS FURTHER REPORTED THAT THE STENT WAS ALLEGEDLY REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19654026 ↗MalfunctionJun 18, 2024Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC VEIN, THE STENT ALLEGEDLY WOULD NOT FULLY EXPAND. IT WAS FURTHER REPORTED THAT THE STENT WAS ROTATED IN COUNTERCLOCKWISE AND THEN CLOCKWISE FOR ABOUT HALF A TURN AND, THEN THE STENT ALLEGEDLY GOT RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19557086 ↗MalfunctionSep 5, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY STICKING DURING DEPLOYMENT. IT WAS FURTHER REPORTED THAT THE CATHETER WAS PUSHED PROXIMALLY AND SLIGHTLY TWISTED IN AND THEN THE STENT WAS RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17682481 ↗MalfunctionJul 21, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN LEFT COMMON ILIAC VEIN VIA LEFT COMMON FEMORAL ACCESS, THE DISTAL PART OF THE STENT ALLEGEDLY FAILED TO OPEN. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17372448 ↗MalfunctionJul 10, 2023Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation
IT WAS REPORTED THAT DURING STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY EXTREMELY HARD TO ADVANCE. IT WAS FURTHER REPORTED THAT THE CATHETER WAS CRIMPED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17285601 ↗MalfunctionJun 8, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE COMMON ILIAC VEIN VIA POPLITEAL VEIN, THE PROXIMAL END OF THE STENT ALLEGEDLY REMAINED STUCK TO THE CATHETER AND DID NOT FULLY EXPAND. IT WAS FURTHER REPORTED THAT AFTER SEVERAL ROTATIONS WITH THE GRIP, THE STENT FINALLY EXPANDED FROM THE CATHETER AFTER SEVERAL MINUTES. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17085392 ↗MalfunctionMay 22, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC VEIN AND INTO THE INFERIOR VENA CAVA VIA BILATERAL GROIN ACCESS, STENT WAS ALLEGEDLY STUCK TO THE STENT CUSHION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16973675 ↗MalfunctionMay 16, 2023Fracture; Detachment of Device or Device Component; Activation Failure; Fracture; Detachment of Device or Device Component; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE COMMON ILIAC VEIN, THE STENT ALLEGEDLY FAILED TO EXPAND. IT WAS FURTHER REPORTED THAT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY BALLOON WAS USED TO POST DILATE WHICH CAUSED THE STENT TO FULLY RELEASE FROM THE CATHETER AND IT TOOK APPROXIMATELY A TOTAL OF TWENTY MINUTES FOR THE STENT TO FULLY EXPAND IN THE PATIENT. FURTHERMORE, THE PHYSICIAN REMOVED THE RETRACTING CATHETER FROM THE BODY. REPORTEDLY, THE STENT WAS ALLEGEDLY FRACTURED FROM ONE OF THE STRUTS AREA WHERE THE STENT WAS STUCK. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16937329 ↗MalfunctionMay 15, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE FOR ILIAC VEIN COMPRESSION IN RIGHT ILIAC VEIN VIA RIGHT GROIN ACCESS, THE STENT ALLEGEDLY STUCK TO THE PAD UPON DEPLOYMENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16928697 ↗MalfunctionMay 11, 2023Malposition of Device; Activation Failure; Malposition of Device; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN ILIAC VEIN, THE CATHETER WAS ALLEGEDLY STUCK TO THE SIDE WALL OF THE STENT. IT WAS FURTHER REPORTED THAT THE STENT WAS PULLED DOWN DURING REMOVAL. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16910363 ↗MalfunctionApr 19, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY STUCK ON THE INNER CATHETER FOR THIRTY SECONDS AND THE USER HAD TWIST THE CATHETER FOR COMPLETE RELEASE FROM THE INNER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16770045 ↗MalfunctionApr 6, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC FEMORAL ARTERY THROUGH POPLITEAL ACCESS, THE STENT WAS ALLEGEDLY STUCK ON CATHETER. IT WAS FURTHER REPORTED THAT THE STENT WAS NOT DEPLOYED AND RELEASED PROPERLY FROM THE CATHETER. REPORTEDLY, PHYSICIAN HAD TO TWIST AND PULL THE CATHETER OFF OF STENT FOR THE STENT TO RELEASE. THE PROCEDURE WAS COMPLETED USING SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16689918 ↗MalfunctionApr 6, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC FEMORAL ARTERY THROUGH POPLITEAL ACCESS, THE STENT WAS ALLEGEDLY STUCK ON CATHETER. IT WAS FURTHER REPORTED THAT THE STENT WAS NOT DEPLOYED AND RELEASED PROPERLY FROM THE CATHETER. REPORTEDLY, PHYSICIAN HAD TO TWIST AND PULL THE CATHETER OFF OF STENT FOR THE STENT TO RELEASE. THE PROCEDURE WAS COMPLETED USING SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16687556 ↗MalfunctionMar 23, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT PLACEMENT PROCEDURE IN RIGHT ILIAC VEIN, THE STENT ALLEGEDLY FAILED TO FULLY DEPLOY FOR AROUND THREE MINUTES ON THE PROXIMAL END AT THE RESIN CONTROL PAD. IT WAS FURTHER REPORTED THAT THE STENT WAS RELEASED FROM THE PAD ROUGHLY THREE MINUTES AFTER THE INITIATION OF THE DEPLOYMENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16596689 ↗MalfunctionMar 22, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN LEFT EXTERNAL ILIAC VIA RIGHT INTERNAL JUGULAR VEIN, THE STENT WAS ALLEGEDLY STUCK TO THE CATHETER AND ALLEGEDLY NOT OPENED FULLY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16589177 ↗MalfunctionMar 4, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN SUPERIOR VENACAVA BRACHIOCEPHALIC LESION SITE VIA LEFT FEMORAL VEIN APPROACH, THE STENT WAS DEPLOYED FINE WITH NO ISSUES UNTIL THE LAST TEN PERCENT AND IT WAS ALLEGEDLY ADHERED TO THE PAD. IT WAS FURTHER REPORTED THAT WHEN STENT HANDLE WAS ROTATED, THE STENT DEPLOYED OFF FROM THE CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16483382 ↗MalfunctionJan 31, 2023Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO EXPAND AND STUCK ON THE SMALL DISTAL PORTION OF THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16270198 ↗MalfunctionJan 5, 2023Misfire; Activation Failure; Misfire; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE VIA FEMORAL VEIN, THE STENT WAS ALLEGEDLY PARTIALLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16110282 ↗InjuryJul 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT POST STENT IMPLANT IN THE LEFT COMMON ILIAC VEIN, THE PATIENT ALLEGEDLY EXPERIENCED ALLERGIC REACTION. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 12165071 ↗MalfunctionApr 8, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN RIGHT COMMON ILIAC VEIN, THE STENT ALLEGEDLY HAD AN EXPANSION ISSUE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11633038 ↗MalfunctionMar 15, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING STENT DEPLOYMENT THE DEVICE ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11482700 ↗MalfunctionMar 9, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE PROXIMAL END OF THE STENT FAILED TO DEPLOY. IT WAS FURTHER REPORTED THAT THE STENT WAS COMPLETELY CAUGHT AND ROTATED THE HANDLE FOR 5 TIMES STENT GOT RELEASED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11441701 ↗MalfunctionMar 4, 2021Activation Failure; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY HAD EXPANSION ISSUE AT PROXIMAL END. IT WAS FURTHER REPORTED THAT AFTER ROTATION OF THE HANDLE THE STENT WAS SUCCESSFULLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11411213 ↗
Other Stent, iliac vein devices
- Abre™ AB9U14080090Medtronic, Inc. · Stent, iliac vein2 recalls9 reports
- Venovo® VENUM16100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall55 reports
- Venovo® VENUM14080Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall36 reports
- Venovo® VENUM14120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall33 reports
- Venovo® VENUM14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall31 reports
- Venovo® VENUM14100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall30 reports
- Venovo® VENUM14160Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall27 reports
- Venovo® VENUL14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall18 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page