Venovo® VENUM20080Recalls & reported problems
Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 20 mm x 80 mm (80 cm delivery catheter) · Class 3 · Implant
All models of Venovo (84)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741103858), received between Feb 4, 2021 and Aug 26, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 14, 2020Class IIOpen, Classified
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
Z-1897-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure5
- Material Deformation2
- Material Twisted/Bent2
- Difficult or Delayed Activation1
- Difficult or Delayed Positioning1
- Incomplete or Missing Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2025Incomplete or Missing Packaging; Incomplete or Missing Packaging
ON (B)(6) 2025, A PATIENT PREPPED FOR A STENT PLACEMENT PROCEDURE USING VENOVO VENOUS STENT. PRIOR TO THE PROCEDURE, IT WAS REPORTED THAT THE ACTUAL PRODUCT DID NOT COME IN PROPER PACKAGING. IT WAS FURTHER REPORTED THAT THE PRODUCT WAS NOT PACKED IN A BOX BUT ONLY IN PLASTIC. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22892843 ↗MalfunctionJun 16, 2023Difficult or Delayed Positioning; Activation Failure; Difficult or Delayed Positioning; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN SUPERIOR VENA CAVA VIA RIGHT BRACHIAL VEIN ACCESS, THE PROXIMAL END OF THE STENT ALLEGEDLY FAILED TO EXPAND UPON DEPLOYMENT. THE PROCEDURE WAS COMPLETED USING SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17141895 ↗MalfunctionMay 8, 2023Difficult or Delayed Activation; Activation Failure; Difficult or Delayed Activation; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN SUPERIOR VENA CAVA VIA RIGHT BRACHIAL VEIN ACCESS, THE PROXIMAL END OF THE STENT ALLEGEDLY FAILED TO EXPAND UPON DEPLOYMENT. THE PROCEDURE WAS COMPLETED USING SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16884147 ↗InjuryMay 20, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT SOME TIME POST STENT PLACEMENT THE STENT ALLEGEDLY BENT. REPORTEDLY, AN ADDITIONAL PROCEDURE WAS REQUIRED. THE CURRENT PATIENT STATUS IS UNKNOWN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11853947 ↗MalfunctionMar 8, 2021Material Twisted/Bent; Activation Failure; Material Twisted/Bent; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE THROUGH COMMON AND EXTERNAL ILIAC VEIN, THE STENT ALLEGEDLY GOT TWISTED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11432212 ↗MalfunctionFeb 4, 2021Material Deformation; Activation Failure; Material Deformation; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE SYSTEM WOULD NOT RELEASE THE END OF THE STENT WHICH CAUSED A STRETCHING OF THE STENT AT THE PROXIMAL END. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11277838 ↗MalfunctionFeb 4, 2021Material Deformation; Activation Failure; Material Deformation; Activation Failure
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE END OF THE STENT WOULD NOT RELEASE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11277339 ↗
Other Stent, iliac vein devices
- Abre™ AB9U14080090Medtronic, Inc. · Stent, iliac vein2 recalls9 reports
- Venovo® VENUM16100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall55 reports
- Venovo® VENUM14080Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall36 reports
- Venovo® VENUM14120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall33 reports
- Venovo® VENUM14140Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall31 reports
- Venovo® VENUM14100Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall30 reports
- Venovo® VENUM16120Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall29 reports
- Venovo® VENUM14160Angiomed GmbH & Co. Medizintechnik KG · Stent, iliac vein1 recall27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page