Skip to content

Venovo® VENUM20100Recalls & reported problems

Angiomed GmbH & Co. Medizintechnik KG · Venovo® Venous Stent System 20 mm x 100 mm (80 cm delivery catheter) · Class 3 · Implant

All models of Venovo (84)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801741103865), received between Dec 9, 2022 and Aug 12, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 14, 2020Class IIOpen, Classified

    Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.

    Z-1897-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure5
  • Difficult or Delayed Positioning2
  • Difficult to Remove1
  • Material Deformation1
  • Misfire1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Packaging Problem; Packaging Problem

    A PATIENT PREPPED FOR A STENT PLACEMENT PROCEDURE USING THE VENOVO VENOUS STENT. IT WAS REPORTED THAT PRIOR TO THE PROCEDURE; THE DELIVERY SYSTEM WAS ALLEGEDLY DAMAGED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22780630 ↗
  • MalfunctionApr 21, 2023Difficult to Remove; Material Deformation; Activation Failure; Difficult to Remove; Material Deformation; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT WAS ALLEGEDLY STUCK FOR A COUPLE OF SECONDS. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16789934 ↗
  • MalfunctionMar 8, 2023Difficult or Delayed Positioning; Activation Failure; Difficult or Delayed Positioning; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC VEIN ABOVE THE GROIN VIA IPSILATERAL APPROACH, THE STENT ALLEGEDLY GOT STUCK TO THE DELIVERY SYSTEM AND TOOK A LITTLE TO GET IT OFF BEFORE IT FULLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16501681 ↗
  • MalfunctionMar 8, 2023Difficult or Delayed Positioning; Activation Failure; Difficult or Delayed Positioning; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN THE ILIAC VEIN ABOVE THE GROIN VIA IPSILATERAL APPROACH, THE STENT ALLEGEDLY GOT STUCK TO THE DELIVERY SYSTEM AND TOOK A LITTLE TO GET IT OFF BEFORE IT FULLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16501680 ↗
  • MalfunctionDec 26, 2022Misfire; Activation Failure; Misfire; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE DISTAL PORTION OF THE STENT ALLEGEDLY WOULD NOT COME OUT OF THE DEPLOYMENT CATHETER. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16046585 ↗
  • MalfunctionDec 9, 2022Activation Failure; Activation Failure

    IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE IN ILIAC VEIN VIA GROIN, THE STENT ALLEGEDLY FAILED TO EXPAND. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15949119 ↗

Other Stent, iliac vein devices

All Stent, iliac vein devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page