Skip to content

PowerPICC CK000549Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC (118)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741107313), received between Jun 25, 2020 and Jun 25, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Material Frayed” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Frayed1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 25, 2020Material Frayed; Material Frayed

    IT IS WAS REPORTED DURING RETRACTION OF STYLET, CLINICIAN NOTICED IT FELT "WEIRD". HE CALLED FOR A CXR TO CONFIRM PLACEMENT BUT ALSO TO REVIEW IF ANY OF STYLET HAD FRACTURED. CXR CONFIRMED FRACTURED STYLET AS WELL AS PLACEMENT OF PICC. CLINICIAN DID NOT SAVE STYLET. IT WAS PLACED IN SHARPS CONTAINER. THE TIP OF THE STYLET IS OBSERVED IN PULMONARY ARTERY. PATIENT REFUSED RETRACTION/REMOVAL OF STYLET.

    Patient problems: Foreign Body In Patient

    MAUDE report 10195833 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page