Skip to content

Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ EN0046140Recalls & reported problems

C. R. Bard, Inc. · Bard® Nasogastric Sump Tube with PreVent® Anti-Reflux Filter with ENFit™ · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741113246), received between May 31, 2018 and Feb 14, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Defective Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 20, 2023Class IIOpen, Classified

    BD (C.R. Bard) has received user complaints reporting that during use of BD Bard NG Sump Tubes with ENFit Connectors, inadequate or reduced suction, decompression, and drainage occurred.

    Z-0798-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Component1
  • Increase in Suction1
  • Insufficient Device Problem Information1
  • Suction Failure1
  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 14, 2024Defective Component; Defective Component

    IT WAS REPORTED THAT THE NASOGASTRIC TUBE HAD UNKNOWN ISSUE. (EN0046140, EN0046160)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18702035 ↗
  • MalfunctionJul 10, 2023Increase in Suction; Suction Failure; Increase in Suction; Suction Failure

    IT WAS REPORTED THAT THE CUSTOMER HAD AN INCIDENT WITH THE NEW ENFIT NG TUBES. DESPITE BARD HAVING IN THEIR PACKAGE INSERT THAT THEY WERE APPROVED FOR GASTRIC DECOMPRESSION, IT FAILED THAT PURPOSE. THE PATIENT HAD THE NG FOR GASTRIC DECOMPRESSION. IT WAS SET ON LOW INTERMITTENT SUCTION WHILE THE PATIENT WAS ON CICU, AND CHARTING INDICATED THAT IT WAS WORKING APPROPRIATELY AT THAT POINT WITH EXPECTED AMOUNTS OF GASTRIC OUTPUT. WHEN THE PATIENT WAS TAKEN TO SURGERY, THE SUCTION WAS INCREASED TO HIGH (NOT SURE IF CONTINUOUS OR INTERMITTENT) AND IT NO LONGER DRAINED GASTRIC OUTPUT. THE SUCTION TUBING THE NG WAS ATTACHED TO WAS REPORTED TO COLLAPSE IN. THE NG TUBE HAD TO BE REPLACED, AND WAS REPLACED WITH A NON-ENFIT VERSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17289178 ↗
  • MalfunctionSep 29, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE NASOGASTRIC TUBE DID NOT DO WHAT INTENDED IN MEDICAL SETTING. THE ISSUE WITH THE NASOGASTRIC TUBE WAS UNKNOWN. THE PRODUCT CATALOG NUMBERS PROVIDED BY CUSTOMER WERE EN0046100, EN0046120, EN0046140, EN0046160 AND EN0046180.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508466 ↗
  • MalfunctionMay 31, 2018Suction Problem

    IT WAS REPORTED THAT THE NG TUBE DO NOT SUCTION WELL. THE COMPLAINANT REPORTED THAT THE NURSING STAFF HAD TO BREAK THE "INNER MOAT" SECTION OUT OF THE ENFIT CONNECTOR TO IMPROVE THE SUCTION.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7557780 ↗

Other TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices

All TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page