Channel Drain V072228Recalls & reported problems

C. R. Bard, Inc. · Novaplus Channel Drain, 15Fr. Round Hubless Full Fluted · Class 1

All models of Channel Drain (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741119163), received between Oct 8, 2024 and Nov 7, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2
  • Material Too Soft/Flexible1
  • Patient Device Interaction Problem1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Foreign Body In Patient1
  • Seroma1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE WOUND DRAINS ARE TOO COMPRESSIBLE AND STOP WORKING. THEY HAVE BROKEN OFF IN PATIENTS, AND THEY HAVE CAUSED SEROMAS IN PATIENTS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Seroma

    MAUDE report 23496071 ↗
  • MalfunctionNov 6, 2025Material Too Soft/Flexible; Material Too Soft/Flexible

    IT WAS REPORTED THAT THE WOUND DRAINS ARE TOO COMPRESSIBLE AND STOP WORKING. THEY HAVE BROKEN OFF IN PATIENTS, AND THEY HAVE CAUSED SEROMAS IN PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23488048 ↗
  • MalfunctionNov 6, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE WOUND DRAINS ARE TOO COMPRESSIBLE AND STOP WORKING. THEY HAVE BROKEN OFF IN PATIENTS, AND THEY HAVE CAUSED SEROMAS IN PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23488039 ↗
  • MalfunctionOct 14, 2024Suction Failure; Suction Failure

    IT WAS REPORTED THAT WHEN CUSTOMER TRIALED THE WOULD DRAINS, THEY SEEMED SUBSTANTIALLY EQUIVALENT TO THEIR CURRENT DRAINS AND PERHAPS EVEN BETTER DUE TO THEIR SOFTER SILICONE. HOWEVER, THEY HAD NOTICED A PATTERN OF CONCERNING FINDINGS WHICH HAS LED THEM TO BELIEVE THEY NEED TO SWITCH BACK TO THEIR OLD VENDOR AS SOON AS POSSIBLE. SPECIFICALLY, CUSTOMER HAD ONE DRAIN BROKE OFF IN A PATIENT NECESSITATING OPERATIVE EXPLORATION, THE DRAINS ROUTINELY FAIL TO MAINTAIN SUCTION, INCREASED THE RISK FOR SEROMA AND INFECTION AND THEY OFTEN BECOME DISCONNECTED FROM THE BULB RESERVOIR, WHICH WAS MADE BY A DIFFERENT MANUFACTURER. CUSTOMER USED HUNDREDS OF DRAINS PER YEAR AND THEIR PERFORMANCE WAS CRITICAL TO THE SUCCESS OF THEIR PROCEDURES. CUSTOMER BELIEVED THAT THEIR CURRENT DRAINS WERE POSING A SIGNIFICANT ONGOING THREAT TO THE WELL-BEING OF THEIR PATIENTS. PER ADDITIONAL INFORMATION RECEIVED ON 27SEP2024, IT WAS REPORTED THAT THE NOVAPLUS 10 FR CHANNEL DRAIN WAS CLINICALLY UNACCEPTABLE BECAUSE TWICE IT HAD BROKEN OFF AT THE SKIN WHEN THEY INSERTED THE DRAIN. THIS INCIDENT OCCURRED AT THE (B)(6) HOSPITAL).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20440287 ↗
  • InjuryOct 8, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT WHEN CUSTOMER TRIALED THE WOULD DRAINS, THEY SEEMED SUBSTANTIALLY EQUIVALENT TO THEIR CURRENT DRAINS AND PERHAPS EVEN BETTER DUE TO THEIR SOFTER SILICONE. HOWEVER, THEY HAD NOTICED A PATTERN OF CONCERNING FINDINGS WHICH HAS LED THEM TO BELIEVE THEY NEED TO SWITCH BACK TO THEIR OLD VENDOR AS SOON AS POSSIBLE. SPECIFICALLY, CUSTOMER HAD ONE DRAIN BROKE OFF IN A PATIENT NECESSITATING OPERATIVE EXPLORATION, THE DRAINS ROUTINELY FAIL TO MAINTAIN SUCTION, INCREASED THE RISK FOR SEROMA AND INFECTION AND THEY OFTEN BECOME DISCONNECTED FROM THE BULB RESERVOIR, WHICH WAS MADE BY A DIFFERENT MANUFACTURER. CUSTOMER USED HUNDREDS OF DRAINS PER YEAR AND THEIR PERFORMANCE WAS CRITICAL TO THE SUCCESS OF THEIR PROCEDURES. CUSTOMER BELIEVED THAT THEIR CURRENT DRAINS WERE POSING A SIGNIFICANT ONGOING THREAT TO THE WELL-BEING OF THEIR PATIENTS. PER ADDITIONAL INFORMATION RECEIVED ON 27SEP2024, IT WAS REPORTED THAT THE NOVAPLUS 10 FR CHANNEL DRAIN WAS CLINICALLY UNACCEPTABLE BECAUSE TWICE IT HAD BROKEN OFF AT THE SKIN WHEN THEY INSERTED THE DRAIN. THIS INCIDENT OCCURRED AT THE SKOKIE HOSPITAL.

    Patient problems: Foreign Body In Patient

    MAUDE report 20398103 ↗

Other CATHETER, IRRIGATION devices

All CATHETER, IRRIGATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page