PowerPICC CK000613Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741121395), received between Sep 26, 2023 and Sep 15, 2026 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Open, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Fluid/Blood Leak2
- Contamination /Decontamination Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Dyspnea2
- Anxiety1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Contamination /Decontamination Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50776464 ↗MalfunctionMar 6, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT A PICC CATHETER WHICH AFTER PLACEMENT WAS LEAKING AND WAS FOUND TO BE CRACKED UPON REMOVAL. THE ORIGINAL PICC PLACEMENT WAS PERFORMED ON (B)(6) 2024. THE PICC WAS LEAKING ON (B)(6) 2024 THE PICC LINE WAS REMOVED AND REPLACED. NO OTHER ADVERSE EVENTS. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED: THE EVENT DID NOT INVOLVE AN URGENT/LIFE THREATENING MEDICAL SITUATION. PATIENT HAD A LEAKING PICC FOR SEVERAL DAYS; THE PICC CATHETER NEEDED TO BE REMOVED AND REPLACED. NO OTHER ADVERSE EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18850227 ↗InjuryJan 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, ON 01-03-2024 TOLEDO HOSPITAL BD PICC WAS PLACED, AND THE PATIENT BECAME ANXIOUS AND SHORT OF BREATH. PATIENT CODE BLUE. NOW RECOVERED AND STABLE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Anxiety; Dyspnea
MAUDE report 18540109 ↗InjuryJan 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, ON 01-01-2024 TOLEDO HOSPITAL BD PICC WAS PLACED, AND RAPID RESPONSE WAS CALLED - THE PATIENT EXPERIENCED LIP TINGLING AND SHORTNESS OF BREATH. PICC WAS REMOVED IMMEDIATELY, AND NO FURTHER ISSUES. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Dyspnea
MAUDE report 18537118 ↗MalfunctionSep 26, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THAT THEY HAD REMOVED A PICC FROM A PATIENT YESTERDAY. THEY STATED THAT THEY WERE IN THE ED THE DAY PRIOR, AFTER FEELING A POP IN THE LINE AND PATIENT REPORTS FEELING FLUID INSIDE HER ARM. LINE WAS INTACT UPON REMOVAL BUT THERE IS AN OBVIOUS HOLE THAT IS ALMOST SEVERING THE LINE BETWEEN THE 8CM AND 9 CM MARKS. THIS HAPPENED AS PATIENT WAS FLUSHING THE LINE, THEY WENT TO THE ED AND WERE TOLD IT FLUSHES FINE, THEY WOULD BE OK UNTIL APPT TOMORROW (9/8). ONLY SALINE WAS FLUSHED AFTER POP. TO MY KNOWLEDGE, NO LINE WAS RETAINED WITHIN THE PATIENT. INSERTION NOTE STATES 38CM, AND 38CM WAS REMOVED. PATIENT WAS ON DAPTOMYCIN FOR TREATMENT WHEN THIS OCCURRED. THERE WAS A STATLOCK IN PLACE AS WELL AS A TEGADERM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17821287 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page