Skip to content

Midline Catheter Kit P4254108DRecalls & reported problems

Bard Access Systems, Inc. · 4Fr DL PowerMidline catheter Max Kit · Class 2

All models of Midline Catheter Kit (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741129353), received between Jun 9, 2017 and Jul 26, 2024 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Class IIOpen, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2663-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Break1
  • Component Incompatible1
  • Hole In Material1
  • Leak/Splash1
  • Material Fragmentation1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Consequences Or Impact To Patient2
  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 26, 2024Component Incompatible; Component Incompatible

    IT WAS REPORTED TRYING TO PUT IN MIDLINE BUT NURSE UNABLE TO SEPARATE THE LINE AND NEEDLE. UNKNOWN IF PATIENT WAS HARMED. THERE WAS PATIENT IMPACT, MIDLINE COULD NOT BE INSERTED AND PROCEDURE WAS CANCELLED. THE PROBLEM OCCURRED WITH THE FLEXIBLE END OF THE GUIDEWIRE, IT GOT CAUGHT ON THE END OF THE INSERTION NEEDLE BEVEL AND COULD NOT BE REMOVED OR ADVANCED. NO OTHER INFORMATION WAS PROVIDED. 07/03/2024 - THE GUIDEWIRE RETURNED FOR EVALUATION EXHIBITED A KINK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19837644 ↗
  • MalfunctionApr 22, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE MIDLINE ¿LEAKED FROM AROUND INSERTION SITE¿.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9985739 ↗
  • MalfunctionApr 22, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE MIDLINE CATHETER WAS ¿LEAKING AT SITE¿.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9985735 ↗
  • MalfunctionMar 20, 2019Leak/Splash

    IT WAS REPORTED MIDLINE WAS SPILT AT SUTURE WING AND WAS LEAKING. THE MIDLINE WAS REMOVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8437276 ↗
  • InjurySep 7, 2018Break; Material Fragmentation

    IT WAS REPORTED THAT DURING THE INSERTION PROCEDURE WHEN THE PROVIDER WAS REMOVING THE GUIDEWIRE, IT BROKE AND A SEGMENT OF THE GUIDEWIRE REMAINED IN THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7856061 ↗
  • MalfunctionAug 31, 2017Hole In Material

    FACILITY REPORTED TO THE SALES REP THAT SMALL HOLES WERE FOUND ON EITHER SIDE OF THE MICROINTRODUCER THAT ALLOWED BLOOD TO COME OUT EVENT WHILE THE END CAP WAS ON. NO OTHER INFORMATION WAS PROVIDED. NO PATIENT HARM REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6836708 ↗
  • MalfunctionJun 9, 2017Fluid/Blood Leak

    PER SALES REP, THE FACILITY REPORTED THAT A DUAL LUMEN MIDLINE LEAKED AT THE CONNECTION OF BIFURCATION AND CATHETER. THE CATHETER WAS REMOVED. THERE WAS NO HARM TO THE PATIENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6632417 ↗

Other Midline catheter devices

All Midline catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page