PowerPICC Provena S9274108DRecalls & reported problems
Bard Access Systems, Inc. · 4 Fr Dual Lumen · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Bard PowerPICC catheter is a long-term intravenous catheter. The 23 reports mostly describe leaks, kinking, and parts breaking, fraying, or separating; some mention guidewire fragments remaining in or moving through the patient. The reports include 3 injuries and no deaths. There is 1 open recall, begun in May 2026, because lidocaine ampoules supplied with the product were made under deficient manufacturing practices and were being recalled by the supplier.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00801741130441), received between Sep 21, 2017 and Jul 12, 2023 — classified as 3 injuries, 20 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- May 21, 2026Open, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2664-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak4
- Break3
- Detachment Of Device Component3
- Leak/Splash3
- Material Deformation3
- Material Frayed3
- Kinked2
- Material Twisted/Bent2
Patient problems
- No Known Impact Or Consequence To Patient12
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions2
- Cardiac Arrest1
- Foreign Body In Patient1
- No Code Available1
- No Information1
- Thrombus1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
InjuryJul 12, 2023Material Frayed; Material Frayed
IT WAS REPORTED BY CUSTOMER THAT APPROXIMATELY 3 CM OF THE TIP OF THE GUIDEWIRE BROKE OFF DURING A PICC PLACEMENT AND REQUIRED INTERVENTIONAL RADIOLOGY TO RETRIEVE THE BROKEN PORTION OF THE WIRE. ADDITIONAL INFO FROM CUSTOMER: 10-JULY-2023: CUSTOMER REPORTED "BROKEN GUIDEWIRE LEFT IN PATIENT THAT MIGRATED TO PULMONARY ARTERY. PATIENT TAKEN TO INTERVENTIONAL RADIOLOGY WHEN RADIOLOGIST ATTEMPTED TO REMOVE BROKEN WIRE. IN THE PROCESS OF ATTEMPTING RETRIEVAL, PATIENT'S AICD LEAD WAS DISLODGED CAUSING, PATIENT TO BECOME HEMODYNAMICALLY UNSTABLE. AT THIS TIME, MEDICATION WAS ADMINISTERED TO IMPROVE HEART RATE AND BLOOD PRESSURE. IN ADDITION, A TEMPORARY CARDIAC PACEMAKER HAD TO BE PLACED EMERGENTLY. A SECOND ATTEMPT WAS MADE 2 DAYS LATER AND BROKEN WIRE WAS SUCCESSFULLY RETRIEVED."
Patient problems: Foreign Body In Patient
MAUDE report 17313098 ↗InjuryJun 10, 2023Material Frayed; Material Frayed
THE CUSTOMER REPORTED A GUIDEWIRE TIP BREAK OFF DURING A PICC INSERTION WITH ABOUT 1.5 CM OF IT REMAINING IN THE PATIENT. THE CLINICIAN WHO PLACED THE LINE STATED THAT IT WAS A SMOOTH INSERTION, THEY MET RESISTANCE ONCE PULLED BACK AND WERE ABLE TO THREAD THE PICC WITHOUT ANY FURTHER RESISTANCE. THEY DID NOT DISCOVER THE WIRE TIP HAD BROKEN OFF UNTIL DAYS LATER AND THE WIRE WAS ALREADY DISCHARGED. WIRE MADE ITS WAY TO PATIENT'S LUNG. THEY WENT IN YESTERDAY TO RETRIEVE THE WIRE AND WERE UNSUCCESSFUL. DURING THE RETRIEVAL PROCESS THE PATIENT CODED. AT THIS POINT THEY ARE STILL DETERMINING THE BEST COURSE OF ACTION FOR THE PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Cardiac Arrest
MAUDE report 17102116 ↗MalfunctionOct 12, 2022Material Frayed; Material Frayed
IT WAS REPORTED BY THE CUSTOMER THAT "THEY HAD A TLS PICC STYLET BREAK DURING INSERTION. BOTH PIECES WERE ABLE TO BE REMOVED".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15587534 ↗MalfunctionJul 16, 2021Material Deformation; Material Deformation
IT WAS REPORTED "PRODUCT COMPLAINT". ADDITIONAL INFORMATION RECEIVED 06/28/2021: "ADMISSION OF 6/8-6/22 THREE CALLS FOR NO BLOOD RETURN; TPA INSTALLATION X2: PICC PULLED BACK X2 TO ESTABLISH BLOOD RETURN. ORIGINAL EXTERNAL 2 CM. PULLED BACK TO 6 CM AS KINK WAS FOUND AT APPROX 5CM. REMOVED ON (B)(6) AS TREATMENT COMPLETE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12182125 ↗MalfunctionNov 12, 2020Material Twisted/Bent
IT WAS REPORTED "AT LEAST 4 PATIENTS RECENTLY THAT HAVE PICCS THAT KINK RIGHT AT THE BEGINNING OF THE CATHETER. CUSTOMER FEELS SHE HAS HAD MULTIPLE PICCS KINK LATELY AT THE INSERTION SITE." THIS REPORT ADDRESSES THE FOURTH PATIENT WITH A KINKED CATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10829725 ↗MalfunctionNov 12, 2020Material Deformation
IT WAS REPORTED "AT LEAST 4 PATIENTS RECENTLY THAT HAVE PICCS THAT KINK RIGHT AT THE BEGINNING OF THE CATHETER. CUSTOMER FEELS SHE HAS HAD MULTIPLE PICCS KINK LATELY AT THE INSERTION SITE." THIS REPORT ADDRESSES THE THIRD PATIENT WITH A KINKED CATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10829724 ↗MalfunctionNov 12, 2020Material Deformation
IT WAS REPORTED "AT LEAST 4 PATIENTS RECENTLY THAT HAVE PICCS THAT KINK RIGHT AT THE BEGINNING OF THE CATHETER. CUSTOMER FEELS SHE HAS HAD MULTIPLE PICCS KINK LATELY AT THE INSERTION SITE." THIS REPORT ADDRESSES THE SECOND PATIENT WITH A KINKED CATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10829717 ↗MalfunctionNov 12, 2020Material Twisted/Bent
IT WAS REPORTED "AT LEAST 4 PATIENTS RECENTLY THAT HAVE PICCS THAT KINK RIGHT AT THE BEGINNING OF THE CATHETER. CUSTOMER FEELS SHE HAS HAD MULTIPLE PICCS KINK LATELY AT THE INSERTION SITE." THIS REPORT ADDRESSES THE FIRST PATIENT WITH A KINKED CATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10829713 ↗MalfunctionApr 23, 2018Break; Detachment Of Device Component; Nonstandard Device
THE FACILITY REPORTED TO THE SALES REP THAT THE HUB BROKE OFF OF THE PROVENA PICC. NO PATIENT INJURY WAS REPORTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7450677 ↗MalfunctionApr 19, 2018Leak/Splash; Nonstandard Device
IT WAS REPORTED THAT THERE WAS A LEAK AT THE JUNCTION BETWEEN THE PURPLE LUMEN AND THE TUBING WITH THE CLAMP. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7445547 ↗MalfunctionApr 6, 2018Disconnection; Component Missing
IT WAS REPORTED THAT A HOMECARE PATIENT THAT HAD A 4FR. PROVENA PICC CAME IN AND BOTH HUB AND CLAVE'S WERE MISSING; THE LINES WERE CLAMPED BUT HUB'S AND CLAVE'S HAD APPARENTLY DISCONNECTED FROM THE TUBING. THE PICC WAS REPLACED IN THE ED. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7403752 ↗MalfunctionMar 19, 2018Obstruction of Flow
THE FACILITY REPORTED THAT THE 4FR D/L CATHETERS ARE CLOTTING OFF EASILY ESPECIALLY WITH CANCER PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Thrombus
MAUDE report 7351004 ↗MalfunctionMar 19, 2018Detachment Of Device Component
THE FACILITY REPORTED THAT THE PATIENT'S LINE WAS PLACED ON (B)(6) 2018. IT WAS STATED THAT THE PURPLE PORT COMPLETELY SLID OFF THE LINE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7350178 ↗MalfunctionDec 20, 2017Fluid/Blood Leak
IT WAS REPORTED THAT THE PICC WAS LEAKING AT THE RED HUB AND TUBING CONNECTION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7133754 ↗MalfunctionDec 19, 2017Fluid/Blood Leak
IT WAS REPORTED THAT THE PICC WAS LEAKING AT THE RED HUB CONNECTION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7131003 ↗MalfunctionDec 19, 2017Fluid/Blood Leak
IT WAS REPORTED THAT THE PICC WAS LEAKING AT THE HUB CONNECTION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7130923 ↗MalfunctionNov 15, 2017Leak/Splash
FACILITY REPORTED TO THE SALES REP THAT THE PICC WAS PLACED ON (B)(6) 2017. RED PORT LEAKING AT THE EXTENSION HUB CONNECTION. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7034731 ↗MalfunctionNov 14, 2017Fluid/Blood Leak
THE FACILITY REPORTED TO THE SALES REPRESENTATIVE THAT THERE WAS A LEAK IN THE CATHETER AT THE RED HUB/EXTENSION LEG. NO OTHER INFORMATION WAS PROVIDED. THERE WAS NO HARM TO THE PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7031661 ↗MalfunctionNov 8, 2017Kinked
DURING THE INVESTIGATION OF THE SAMPLE, IT WAS NOTED THAT THAT THE PICC CATHETER WAS KINKED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7010299 ↗MalfunctionNov 3, 2017Leak/Splash
IT WAS REPORTED TO THE SALES REP FROM THE FACILITY THAT THE PROVENA PICC LEAKED JUST BELOW THE HUB ON THE PURPLE TUBING OF THE EXTENSION LEG. THE PICC WAS REMOVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7001458 ↗InjuryOct 19, 2017Detachment of Device or Device Component
IT WAS REPORTED BY THE SALES REP THAT THE FACILITY WAS PLACING A PICC, RIGHT SIDED PLACEMENT AND UPON INITIAL INSERTION MET RESISTANCE. THE PLACEMENT WAS PULLED BACK; MANIPULATED THE CATHETER AND THE PICC WENT IN FINE ON THE SECOND ATTEMPT. WHEN THE PLACER WENT TO RETRACT THE SHERLOCK STYLET, SHE FELT RESISTANCE WITH THE STYLET AND REMOVED THE STYLET FROM THE PICC. UPON REMOVAL, THE STYLET WAS MISSING THE TIP. THE CATHETER WAS THEN REMOVED AND FLUSHED TO ENSURE THE TIP WAS NOT IN THE CATHETER. CHEST X-RAY SHOWED APPROXIMATELY 3.5 CM OF THE SHERLOCK TIP IN THE PATIENT'S RIGHT AXILLARY VEIN. THE PATIENT HAS LEFT SIDE SWAN-GANZ CATHETER AND PACEMAKER. THE PICC WAS TRIMMED TO 40 CM. INTERVENTIONAL RADIOLOGY RETRIEVED THE WIRE.
Patient problems: No Code Available
MAUDE report 6963639 ↗MalfunctionOct 12, 2017Break; Kinked
REPORTED BY FACILITY THAT THIS PICC WAS PLACED (B)(6) 2017, REMOVED (B)(6) 2017 BECAUSE OF KINKS IN CATHETER AND A BREAK AT THE HUB OF THE PURPLE LUMEN HUB.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6945853 ↗MalfunctionSep 21, 2017Break; Detachment Of Device Component
PER SALES REP, THE FACILITY REPORTED THAT THE RED HUB ON THE CATHETER BROKE OFF. THE PICC WAS REMOVED. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Information
MAUDE report 6883786 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page