CK000525B PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen recallZ-2664-2026 · Bard Access Systems, Inc. · initiated May 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Access Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt
Recall details
- Recall number
- Z-2664-2026
- Status
- Open, Classified
- Initiated by the company
- May 21, 2026
- Posted by the FDA
- Jul 2, 2026
- Recalling firm
- Bard Access Systems, Inc.
- Quantity
- 29,490
- Product code
- LJS
- FDA submissions
- K180548
- Distribution
- Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
Product description
CK000525B PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236129 CK000883 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188800 CK000884A PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741219177 CK001000 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741223020 CK001066 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741228544 CK001068 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741228728 S1173108D PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741130342 S1173108D1 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154720 S1173108D4 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155031 S1173108D5 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155130 S1173108D6 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741232190 S1193108D3 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154980 S1193108D4 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155086 S1193108D5 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155185 S1193108PD PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741134135 S1274108D PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130366 S1274108D4 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155055 S1274108D5 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155154 S1294108D PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130373 S1294108D2 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154898 S1294108D4 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155109 S1294108D5 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155208 S3173108 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741130144 S3274108D PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130403 S9274108D PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130441 The PowerPICC Provena Catheters and PowerPICC Provena Catheter with SOLO Valve Technology are indicated for short and long term peripheral access to central venous system for intravenous therapy, blood sampling, power injection of contrast media, and allow for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that a catheter lumen of 20 guage or larger be used.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
CK000525B PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236129 Lot Numbers: REKW2510 REKX3786 CK000883 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188800 Lot Number: REKX3767 CK000884A PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741219177 Lot Number: REKX0133 CK001000 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741223020 Lot Number: REKY1171 CK001066 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741228544 Lot Number: REKX0087 CK001068 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741228728 Lot Number: REKW1569 S1173108D PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741130342 Lot Number: REKX3449 S1173108D1 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154720 Lot Number: REKU2923 S1173108D4 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155031 Lot Numbers: REKV3208 REKW2482 S1173108D5 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155130 Lot Number: REKX0070 S1173108D6 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741232190 Lot Number: REKY0141 S1193108D3 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154980 Lot Number: REKV3188 S1193108D4 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155086 Lot Numbers: REKV3240 REKW0961 S1193108D5 PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741155185 Lot Number: REKW1468 S1193108PD PowerPICC Provena SOLO Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741134135 Lot Numbers: REKV3256 REKY1690 S1274108D PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130366 Lot Number: REKW1461 S1274108D4 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155055 Lot Number: REKW2520 S1274108D5 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155154 Lot Number: REKW1458 S1294108D PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130373 Lot Number: REKV2247 S1294108D2 PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741154898 Lot Numbers: REKW0952 REKX2229 S1294108D4 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155109 Lot Numbers: REKW1469 REKX0072 S1294108D5 PowerPICC Provena SOLO Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741155208 Lot Numbers: REKW1462 REKX0143 S3173108 PowerPICC Provena Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741130144 Lot Number: REKV3241 S3274108D PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130403 Lot Number: REKX1638 S9274108D PowerPICC Provena Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741130441 Lot Number: REKW0965
Devices named in this recall
25 devices in the FDA device registry whose identifier appears in the recall's code information.
- PowerPICC Provena S1274108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall120 reports
- PowerPICC Provena S1274108D4Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall87 reports
- PowerPICC Provena S1274108D5Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall66 reports
- PowerPICC Provena S1173108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall45 reports
- PowerPICC Provena S1294108D4Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall44 reports
- PowerPICC Provena S1294108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall34 reports
- PowerPICC Provena S1173108D4Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall28 reports
- PowerPICC Provena S1294108D2Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall26 reports
- PowerPICC Provena S1294108D5Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall1 death26 reports
- PowerPICC Provena S9274108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall23 reports
- PowerPICC Provena S1193108D5Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall13 reports
- PowerPICC Provena S1193108D4Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall11 reports
- PowerPICC Provena S1173108D5Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall11 reports
- PowerPICC Provena S3274108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall7 reports
- PowerPICC Provena S1193108D3Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall7 reports
- PowerPICC Provena CK000883Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall4 reports
- PowerPICC Provena S1193108PDBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall2 reports
- PowerPICC Provena S1173108D1Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall1 report
- PowerPICC Provena S3173108Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall0 reports
- PowerPICC Provena SOLO CK000884ABard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Provena SOLO CK001000Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Provena Sherlock 3CG CK001066Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Provena Sherlock 3CG CK001068Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Provena S1173108D6Bard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- PowerPICC Provena SOLO CK000525BBard Access Systems, Inc. · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99096).
- Z-2656-2026 · Open, ClassifiedCK000495A Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741121609 CK000566 Poly Midline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741108051 CK000566 Poly Midline C…
- Z-2657-2026 · Open, Classified5625240 DuoGlide Short-Term Straight Dialysis Catheter 13 Fr Dual-Lumen UDI-DI Code: 00801741042119 The intended use of the Short-Term Dialysis Catheter is attaining temporary/sh…
- Z-2658-2026 · Open, ClassifiedCK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide internal reinforcement to aid in catheter placement.
- Z-2659-2026 · Open, Classified1274108D PowerPICC SV Catheter 4 Fr Dual-Lumen REKW2449 00801741087110 3173108D PowerPICC SV Catheter 3 Fr Single-Lumen REKV2227 00801741027536 The PowerPlCC SV catheters are in…
- Z-2660-2026 · Open, Classified5605150 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086953 5605240 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Tr…
- Z-2661-2026 · Open, Classified5855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107177 5855300 Power-Trialysis Slim-Cath Straight Short-Term Dialys…
- Z-2662-2026 · Open, ClassifiedCK000727 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138478 CK000729 PowerMidline Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741138713 P4153108D PowerMidline C…
- Z-2663-2026 · Open, ClassifiedCK000980A PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236150 P4254108 PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741129278 P4254108D PowerMidline Cath…
- Z-2666-2026 · Open, Classified1174108 PowerPICC Catheter 4 Fr Single-Lumen UDI-DI Code: 00801741154935 1295108 PowerPICC SOLO Catheter 5 Fr Dual-Lumen UDI-DI Code: 00801741034602 1174108D3 PowerPICC Catheter …
- Z-2667-2026 · Open, ClassifiedCK000880 Provena Midline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741188770 S4153108BDP Provena Midline Catheter 3 Fr Single-Lumen UDI-DI Code: 00801741154003 S4153108BP Proven…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page