PowerPICC CK000677Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741131790), received between Aug 3, 2023 and Dec 31, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Break1
- Material Deformation1
Patient problems
- Extravasation1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- Skin Burning Sensation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2024Break; Break
IT WAS REPORTED THAT PICC FRACTURED AT 5CM DURING REMOVAL. EXTENDED STAY IN ER. PT ARRIVED FOR A "SIMPLE" PICC LINE REMOVAL. ENDED UP WITH 38CM OF RETAINED CATHETER WHICH PER CXR WAS INTACT. PATIENT UNDERWENT FLURO/IR PROCEDURE TO REMOVE THE REMAINING CATHETER VIA ACCESS OF THE FEMORAL VEIN. PATIENT TOLERATED WELL AND WAS DISCHARGED TO HOME WITH HIS SPOUSE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 21047771 ↗MalfunctionDec 30, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY THE CUSTOMER THE DEVICE WAS LEAKING. THERE WAS A SMALL OPENING IN THE PICC LINE WHEN REMOVED. INFILTRATED INTO ARM, REQUIRED TREATMENT AS WELL AS CAUSING A DELAY IN CARE.
Patient problems: Extravasation
MAUDE report 21044732 ↗MalfunctionOct 9, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT THE PICC LINE ACTUALLY KNOT WHILE BEING INSERTED. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED 08 OCTOBER 2023: PICC WAS NOT KINKED OR BENT PRIOR TO INSERTION INTO VEIN. URGENT DUE TO BEING UNABLE TO REMOVE KNOTTED PICC LINE. PICC LINE KNOTTED IN PATIENT ARM WHICH REQUIRED INTERVENTIONAL RADIOLOGY TO REMOVE. DID NOT INTERRUPT THE ADMINISTRATION OF ANY MEDICATION AS PATIENT HAD OTHER IV ACCESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17896047 ↗MalfunctionAug 3, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT LEAKAGE FOUND AT THE INSERTION SITE. PATIENT COMPLAINED OF BURNING AT THE SITE. EVALUATED AND FOUND LEAK AT THE 2CM MARK JUST BELOW THE SKIN LEVEL AT THE INSERTION SITE. REMOVED. RESPONSE RECEIVED 27 JULY 2023: THE EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING MEDICAL SITUATION, CAUSE A CLINICALLY SIGNIFICANT DELAY IN TREATMENT, INTERRUPTION IN TREATMENT, AND CAUSED NO HARM TO THE PATIENT. PATIENT WAS RECEIVING OR PLANNED TO RECEIVE CEFAZOLIN.
Patient problems: Skin Burning Sensation
MAUDE report 17450402 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.