Skip to content

PowerPICC CK000677Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741131790), received between Aug 3, 2023 and Dec 31, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Break1
  • Material Deformation1

Patient problems

  • Extravasation1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Skin Burning Sensation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 31, 2024Break; Break

    IT WAS REPORTED THAT PICC FRACTURED AT 5CM DURING REMOVAL. EXTENDED STAY IN ER. PT ARRIVED FOR A "SIMPLE" PICC LINE REMOVAL. ENDED UP WITH 38CM OF RETAINED CATHETER WHICH PER CXR WAS INTACT. PATIENT UNDERWENT FLURO/IR PROCEDURE TO REMOVE THE REMAINING CATHETER VIA ACCESS OF THE FEMORAL VEIN. PATIENT TOLERATED WELL AND WAS DISCHARGED TO HOME WITH HIS SPOUSE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21047771 ↗
  • MalfunctionDec 30, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER THE DEVICE WAS LEAKING. THERE WAS A SMALL OPENING IN THE PICC LINE WHEN REMOVED. INFILTRATED INTO ARM, REQUIRED TREATMENT AS WELL AS CAUSING A DELAY IN CARE.

    Patient problems: Extravasation

    MAUDE report 21044732 ↗
  • MalfunctionOct 9, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY CUSTOMER THAT THE PICC LINE ACTUALLY KNOT WHILE BEING INSERTED. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED 08 OCTOBER 2023: PICC WAS NOT KINKED OR BENT PRIOR TO INSERTION INTO VEIN. URGENT DUE TO BEING UNABLE TO REMOVE KNOTTED PICC LINE. PICC LINE KNOTTED IN PATIENT ARM WHICH REQUIRED INTERVENTIONAL RADIOLOGY TO REMOVE. DID NOT INTERRUPT THE ADMINISTRATION OF ANY MEDICATION AS PATIENT HAD OTHER IV ACCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17896047 ↗
  • MalfunctionAug 3, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY CUSTOMER THAT LEAKAGE FOUND AT THE INSERTION SITE. PATIENT COMPLAINED OF BURNING AT THE SITE. EVALUATED AND FOUND LEAK AT THE 2CM MARK JUST BELOW THE SKIN LEVEL AT THE INSERTION SITE. REMOVED. RESPONSE RECEIVED 27 JULY 2023: THE EVENT DID NOT INVOLVE AN URGENT OR LIFE-THREATENING MEDICAL SITUATION, CAUSE A CLINICALLY SIGNIFICANT DELAY IN TREATMENT, INTERRUPTION IN TREATMENT, AND CAUSED NO HARM TO THE PATIENT. PATIENT WAS RECEIVING OR PLANNED TO RECEIVE CEFAZOLIN.

    Patient problems: Skin Burning Sensation

    MAUDE report 17450402 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.