Skip to content

PowerPICC SOLO CK000687Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC SOLO (78)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741136535), received between Mar 21, 2019 and Sep 10, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Activation Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Class IOpen, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Problem3
  • Device Contamination with Chemical or Other Material1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2025Activation Problem; Activation Problem

    IT WAS REPORTED BY CUSTOMER THAT THE SECURLOC NEEDLE IN 4FR PICC KIT DID NOT SAFETY- NEEDLE EXPOSED. NO HARM RESULTED FROM THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23018902 ↗
  • MalfunctionSep 9, 2025Activation Problem; Activation Problem

    IT WAS REPORTED, THE SECURLOC NEEDLE IN 4FR PICC KIT DID NOT SAFETY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23000120 ↗
  • MalfunctionAug 13, 2025Activation Problem; Activation Problem

    IT WAS REPORTED SECURLOC INTRODUCER NEEDLE FROM PICC KIT DID NOT SAFETY. GREEN SAFETY MECHANISM CAME OFF COMPLETELY. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22793711 ↗
  • MalfunctionJul 10, 2019Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A HAIR WAS FOUND ON THE CHLORAPREP STICK IN STERILE CUSTOM KIT. DEVICE WAS NOT USED ON PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 8777742 ↗
  • MalfunctionMar 21, 2019Material Twisted/Bent

    IT WAS REPORTED THAT DURING INSERTION THE 3CG WIRE BENT IN A FEW PLACES AND DURING REMOVAL, THE WIRE COMPLETELY SEVERED. IT WAS STATED THE SAME PICC CATHETER WAS PLACED WITH A DIFFERENT WIRE. IT WAS STATED THERE WAS NO PATIENT HARM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8441847 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page