PowerPICC SOLO CK000687Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC SOLO (78)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741136535), received between Mar 21, 2019 and Sep 10, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Activation Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Problem3
- Device Contamination with Chemical or Other Material1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2025Activation Problem; Activation Problem
IT WAS REPORTED BY CUSTOMER THAT THE SECURLOC NEEDLE IN 4FR PICC KIT DID NOT SAFETY- NEEDLE EXPOSED. NO HARM RESULTED FROM THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23018902 ↗MalfunctionSep 9, 2025Activation Problem; Activation Problem
IT WAS REPORTED, THE SECURLOC NEEDLE IN 4FR PICC KIT DID NOT SAFETY. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23000120 ↗MalfunctionAug 13, 2025Activation Problem; Activation Problem
IT WAS REPORTED SECURLOC INTRODUCER NEEDLE FROM PICC KIT DID NOT SAFETY. GREEN SAFETY MECHANISM CAME OFF COMPLETELY. NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22793711 ↗MalfunctionJul 10, 2019Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT A HAIR WAS FOUND ON THE CHLORAPREP STICK IN STERILE CUSTOM KIT. DEVICE WAS NOT USED ON PATIENT.
Patient problems: No Patient Involvement
MAUDE report 8777742 ↗MalfunctionMar 21, 2019Material Twisted/Bent
IT WAS REPORTED THAT DURING INSERTION THE 3CG WIRE BENT IN A FEW PLACES AND DURING REMOVAL, THE WIRE COMPLETELY SEVERED. IT WAS STATED THE SAME PICC CATHETER WAS PLACED WITH A DIFFERENT WIRE. IT WAS STATED THERE WAS NO PATIENT HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8441847 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page