Skip to content

PowerMidline CK000727Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2

All models of PowerMidline (15)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741138478), received between May 16, 2024 and Mar 17, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Defective Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Open, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2662-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Component2
  • Material Deformation2
  • Contamination /Decontamination Problem1
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Hematoma1
  • Hemorrhage/Blood Loss/Bleeding1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 17, 2026Material Deformation

    IT WAS REPORTED THREE BARD SL POWER KITS WHERE THE TOP OF THE GRAY SHEATH OF THE INTRODUCER HAS A SMALL SPLIT APPROXIMATELY 1 CM IN LENGTH. THIS IS BEFORE USE. THERE WAS NO NEGATIVE OUTCOME THAT OCCURRED AS A RESULT. THIS REPORT ADDRESSES ALL THREE KITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24617548 ↗
  • MalfunctionMar 17, 2026Material Deformation

    IT WAS REPORTED BARD SL POWER KIT WHERE THE TOP OF THE GRAY SHEATH OF THE INTRODUCER HAS A SMALL SPLIT APPROXIMATELY 1 CM IN LENGTH. THIS IS BEFORE USE. I NOTICED THE SPLIT UPON INSERTION WITH VERY LITTLE FORCE. NO NEGATIVE OUTCOME OCCURRED AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24617536 ↗
  • MalfunctionJun 3, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THERE HAVE BEEN SEVERAL REPORTS OF FIBERS ON THE WIRE. THE EXACT NUMBER OF OCCURRENCES WAS NOT PROVIDED BY THE COMPLAINANT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22132944 ↗
  • MalfunctionApr 17, 2025Defective Component; Defective Component

    IT WAS REPORTED "I WAS PLACING A MIDLINE TODAY AND NOTICED A WEIRD SPIRALING ON THE INTERNAL WIRE FOR THE POWER MIDLINE. WHEN I PULLED THE WIRE OUT OF THE MIDLINE AFTER THE MIDLINE WAS PLACED, I NOTICED FRAYING ON THE WIRE ITSELF. YOU CAN SEE SOME SMALL WHITE SPECKS AND FRAYS. I THEN NOTICED IT ON MY SECOND MIDLINE OF THE DAY. A COLLEGE ASKED ME IF I WOULD BE ABLE TO PULL ANY OF THOSE PIECES OFF IF I GRABBED ON TO THEM. I TRIED THIS AFTER AND I HAD ONE THAT FLAKED OFF RIGHT IN MY HAND. JUST LOOKED LIKE A LITTLE WHITE SPECK. THE FRAYING DOES GET WORSE IF YOU RUN YOUR FINGERS ACROSS IT., IT'S WAY MORE NOTICEABLE AT THAT POINT. THERE WAS NO ADVERSE EVENT DURING OR AFTER THE PROCEDURE." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES BOTH DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21858491 ↗
  • MalfunctionMay 16, 2024Difficult to Insert; Defective Component; Difficult to Insert; Defective Component

    IT WAS REPORTED, ¿THIS GUIDE WIRE DID NOT LOOK THIS WAY UNTIL AFTER I TRIED TO PUT IT IN THE PT., AT WHICH TIME I COULD NOT ADVANCE IT AND LITERALLY HAD TO FORCE IT OUT OF THE VEIN! THE VEIN WAS NO LONGER USABLE AFTER THAT. AFTER I FORCED THE GUIDE WIRE OUT, SHE DID HAVE SOME BLEEDING AND EXPRESSED SOME PAIN AT THE SITE, THE BLEEDING WAS EASILY CONTROLLED. IT DID LEAVE A VISIBLE HEMATOMA THAT RAPIDLY APPEARED, WHICH I NORMALLY DO NOT SEE IN MY PATIENTS, BUT AGAIN SHE WAS COMPROMISED, AND HER EXTREMITIES WERE COVERED WITH BRUISES.¿ NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 19338029 ↗

Other Midline catheter devices

All Midline catheter devices →

Sources and updates