Skip to content

PowerPICC SOLO CK000501ARecalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC SOLO (78)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741140815), received between Jul 31, 2019 and Apr 29, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Class IOpen, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Difficult to Advance1
  • Fluid/Blood Leak1
  • Obstruction of Flow1

Patient problems

  • Extravasation1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Swelling/ Edema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER "CT SCAN WAS PERFORMED WITH HIGH-PRESSURE INJECTION. THE CT TECH RECOGNIZED THAT THE IMAGE WAS LACKING CONTRAST RESOLUTION AND CHECK THE PATIENT'S ARM AND NOTICED LARGE AMOUNT OF SWELLING ABOVE THE INSERTION SITE. WE WERE CALLED TO REPLACE THE LINE, I REQUESTED THE NURSE SAVE THE CATHETER AND NOT PERFORM A GUIDEWIRE EXCHANGE. THE PATIENT NEEDED A CT SCAN. MY RECOMMENDATION WAS TO REPLACE THE LINE IN THE OTHER ARM WITH A DIFFERENT LOT NUMBER. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Extravasation; Swelling/ Edema

    MAUDE report 21929145 ↗
  • MalfunctionJan 14, 2021Break; Difficult to Advance

    IT WAS REPORTED "THE 3CG STYLET BROKE OFF THE PICC. RN STATED HE HAD TROUBLE ADVANCING THE PICC AND DECIDED TO RETRACT IT, HE THEN DISCOVERED THE 3CG STYLET HAD BROKEN OFF." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11170423 ↗
  • MalfunctionJul 31, 2019Obstruction of Flow

    IT WAS REPORTED VIA MS&S "NURSE STATES HAS A PATIENT WITH A POWERPICC THAT WILL NOT FLUSH. MS&S DISCUSSED POSSIBLE CAUSES OF OCCLUSION. RECOMMENDED SHE SPEAK WITH THE MD ABOUT THE USE OF CATHFLO OR A DYE STUDY."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8848987 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →