Skip to content

PowerPICC CK000757Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741141133), received between May 3, 2024 and Dec 7, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Pain1
  • Swelling/ Edema1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 7, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY MEDWATCH THAT POWERPICC SV CATHETER WITH SHERLOCK TIP LOCATION SYSTEM WAS INSERTED PRIOR TO PATIENT'S DISCHARGE HOME WITH HOME ANTIBIOTIC INFUSIONS. PATIENT RETURNED WITH COMPLAINS OF LEAKING AND TENDERNESS AT PICC SITE, PICC LINE WAS PULLED OUT 7 CM IN COMPARISON TO ORIGINAL INSERTION MEASUREMENTS AT 2 CM. PICC WAS REMOVED AND A SLIT WAS FOUND IN THE PICC LINE TUBING AT JUST OUTSIDE THE 7 CM MARK. DUE TO LAPS IN TIME FROM INSERTION TO PATIENT RETURN, PACKAGING WAS DISCARDED AND THE LOT NUMBER IS NOT AVAILABLE. UNKNOWN HOW PATIENT CARED FOR DEVICE AT HOME PRIOR TO RETURN. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Pain

    MAUDE report 20873944 ↗
  • InjuryMay 3, 2024Fluid/Blood Leak; Adverse Event Without Identified Device or Use Problem; Fluid/Blood Leak; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED, "ON (B)(6) 2024 A PATIENT CAME TO THE ER FOR PICC ASSESSMENT. THE PICC LINE WAS PLACED ON (B)(6) 2024 AND WAS LEFT EXPOSED AT 0 CM. THE PATIENT STATED THAT HE HAD NOTICED SWELLING UNDER THE SKIN NEAR THE INSERTION SITE WHEN FLUSHING. THE HOME HEALTH RN DECIDED TO PULL THE CATHETER OUT 3 CM AND THEN NOTICED THE CRACK. HE ARRIVED IN THE ER, AND AT THAT TIME I REMOVED THE PICC LINE. WHEN I ATTEMPTED TO FLUSH THE CATHETER POST REMOVAL, A BLOOD CLOT WAS DISLODGED FROM THE TIP OF THE CATHETER MEASURING AROUND AN INCH. THE PATIENT ALSO STATED TO ME THAT HIS LINE HAD BECOME DIFFICULT TO FLUSH ¿ BUT STATED THAT HE DID NOT THINK HE APPLIED EXCESSIVE PRESSURE WHILE ATTEMPTING TO FLUSH." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Swelling/ Edema; Thrombosis/Thrombus

    MAUDE report 19239835 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →

Sources and updates