PowerPICC CK000757Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00801741141133), received between May 3, 2024 and Dec 7, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Pain1
- Swelling/ Edema1
- Thrombosis/Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 7, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY MEDWATCH THAT POWERPICC SV CATHETER WITH SHERLOCK TIP LOCATION SYSTEM WAS INSERTED PRIOR TO PATIENT'S DISCHARGE HOME WITH HOME ANTIBIOTIC INFUSIONS. PATIENT RETURNED WITH COMPLAINS OF LEAKING AND TENDERNESS AT PICC SITE, PICC LINE WAS PULLED OUT 7 CM IN COMPARISON TO ORIGINAL INSERTION MEASUREMENTS AT 2 CM. PICC WAS REMOVED AND A SLIT WAS FOUND IN THE PICC LINE TUBING AT JUST OUTSIDE THE 7 CM MARK. DUE TO LAPS IN TIME FROM INSERTION TO PATIENT RETURN, PACKAGING WAS DISCARDED AND THE LOT NUMBER IS NOT AVAILABLE. UNKNOWN HOW PATIENT CARED FOR DEVICE AT HOME PRIOR TO RETURN. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Pain
MAUDE report 20873944 ↗InjuryMay 3, 2024Fluid/Blood Leak; Adverse Event Without Identified Device or Use Problem; Fluid/Blood Leak; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, "ON (B)(6) 2024 A PATIENT CAME TO THE ER FOR PICC ASSESSMENT. THE PICC LINE WAS PLACED ON (B)(6) 2024 AND WAS LEFT EXPOSED AT 0 CM. THE PATIENT STATED THAT HE HAD NOTICED SWELLING UNDER THE SKIN NEAR THE INSERTION SITE WHEN FLUSHING. THE HOME HEALTH RN DECIDED TO PULL THE CATHETER OUT 3 CM AND THEN NOTICED THE CRACK. HE ARRIVED IN THE ER, AND AT THAT TIME I REMOVED THE PICC LINE. WHEN I ATTEMPTED TO FLUSH THE CATHETER POST REMOVAL, A BLOOD CLOT WAS DISLODGED FROM THE TIP OF THE CATHETER MEASURING AROUND AN INCH. THE PATIENT ALSO STATED TO ME THAT HIS LINE HAD BECOME DIFFICULT TO FLUSH ¿ BUT STATED THAT HE DID NOT THINK HE APPLIED EXCESSIVE PRESSURE WHILE ATTEMPTING TO FLUSH." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Swelling/ Edema; Thrombosis/Thrombus
MAUDE report 19239835 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page