PowerPICC CK000819Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741152566), received between Mar 1, 2023 and Jan 14, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Material Deformation2
  • Material Discolored2
  • Defective Component1
  • Failure to Infuse1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 14, 2025Material Deformation; Material Deformation

    IT WAS REPORTED, "MOM UNABLE TO FLUSH PICC AT HOME. DRESSING CHANGED 2 DAYS PRIOR TO ISSUE BUT USED YESTERDAY WITHOUT ISSUES. 2 DOSES OF TPA UNSUCCESS FOR BRISK BLOOD RETURN (ABLE FLUSH AFTER FIRST DOSE). X-RAY SHOWED CATHETER IS NOW BRACHIOCEPHALIC (POSSIBLE KINK IN ARM). CATHETER REMOVED AND NEW LINE PLACED. CATHETER SECURED WITH SECURACATH, 14CM ABOVE ANTECUBITAL FOSSA, 5CM EXTERNAL. PLACED ON (B)(6) 2024 AND REMOVED ON (B)(6) 2024. PATIENT REQUIRED ED VISIT OVER THE WEEKEND. FORTUNATELY, WE WERE ABLE TO REPLACE IN THE ED AND DID NOT REQUIRE AN ADMISSION. CATHETER REMOVED AND NEW LINE PLACED. NO PATIENT HARM REPORTED." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21148100 ↗
  • MalfunctionSep 14, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY CUSTOMER THAT 4 FR SINGLE LUMEN PICC RUPTURE AT 40CM MARK. RUPTURE FOUND ON LINE REMOVAL THERAPY NO LONGER NEEDED. NO OTHER INFORMATION PROVIDED. ADDITIONAL INFO FROM CUSTOMER: ((B)(6) 2023) CATHETER WAS REMOVED (B)(6) 2023 NO NEW CATHETER PLACED AS THERAPY COMPLETE. PATIENT MENTIONED TO NURSE REMOVING THE LINE THAT SHE HAD A LOT OF ISSUES WITH IT SO RN CLOSELY INSPECTED CATHETER AND FOUND RUPTURE AT 40 CM MARK. IT WAS SECURED WITH SECUR-A-CATH. NO PT HARM REPORTED. WAS BEING TREATED FOR PERSISTENT HEADACHES AND DOUBLE VISION ADJUSTMENT OF PROGRAMMABLE SHUNT. THEY DO NOT HAVE AN INFECTION DISEASE. UNKOWN WHEN CATHETER RUPTURE OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17752916 ↗
  • MalfunctionMar 28, 2023Material Discolored; Defective Component; Material Discolored; Defective Component

    IT WAS REPORTED BY THE CUSTOMER, "BROWN DISCOLORATION AFTER USING SCISSORS. THE DISCOLORATION ON THE RING DISAPPEARED WITHIN A COUPLE HOURS. WHILE TRIMMING STATSEAL DISC BROWN DISCOLORATION WAS NOTED TO COME OFF SCISSORS. NEW RING AND SCISSORS OBTAINED BUT PICC HAD ALREADY BEEN PLACED AND SCISSORS USED TO TRIM CATHETER. STATLOCK WAS USED TO SECURE PICC. NO ISSUES AS OF 3/8/2023 WILL CONTINUE TO MONITOR. THERE WAS NO CONCOMITANT THERAPY AS EVENT OCCURRED WITH PICC PLACEMENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16632026 ↗
  • MalfunctionMar 1, 2023Material Discolored; Material Discolored

    IT WAS REPORTED BY THE CUSTOMER THAT "SOME SCISSORS FROM THE KIT THAT LEFT DISCOLORATION ON THE STAT SEAL DISC WHEN IT WAS TRIMMED. " ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER: "THE DISCOLORATION WAS BROWN. A NEW DISC AND SCISSORS WERE OBTAINED BUT PICC HAD ALREADY BEEN PLACED AND THE SCISSORS WERE USED TO TRIM CATHETER. PATIENT IS BEING MONITORED FOR POTENTIAL COMPLICATIONS. PATIENT IS BEING TREATED FOR MSSA BACTEREMIA AND PULMONARY VALVE ENDOCARDITIS."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16461378 ↗
  • MalfunctionMar 1, 2023Failure to Infuse; Material Deformation; Failure to Infuse; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER "12/22/2022 PT REPORTED THAT HER PICC WOULD OCCLUDE FREQUENTLY AND THAT SHE WOULD HAVE TO ADDUCT HER RIGHT ARM WHILE KEEPING THE PALMS UP TOWARDS THE CEILING. DRESSING CHANGED RN STARTED NS BOLUS AND NO ISSUES OBSERVED. 12/29 REPORTS PICC OCCLUSION ISSUES. X-RAY REQUESTED. ATTEMPTED PULLING LINE BACK TO ELIMINATE LOOP X 2. PICC WAS REMOVED ON (B)(6) AND NEW PICC ORDER PLACED." NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16461011 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →