PowerPICC CK000819Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741152566), received between Mar 1, 2023 and Jan 14, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Material Deformation2
- Material Discolored2
- Defective Component1
- Failure to Infuse1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 14, 2025Material Deformation; Material Deformation
IT WAS REPORTED, "MOM UNABLE TO FLUSH PICC AT HOME. DRESSING CHANGED 2 DAYS PRIOR TO ISSUE BUT USED YESTERDAY WITHOUT ISSUES. 2 DOSES OF TPA UNSUCCESS FOR BRISK BLOOD RETURN (ABLE FLUSH AFTER FIRST DOSE). X-RAY SHOWED CATHETER IS NOW BRACHIOCEPHALIC (POSSIBLE KINK IN ARM). CATHETER REMOVED AND NEW LINE PLACED. CATHETER SECURED WITH SECURACATH, 14CM ABOVE ANTECUBITAL FOSSA, 5CM EXTERNAL. PLACED ON (B)(6) 2024 AND REMOVED ON (B)(6) 2024. PATIENT REQUIRED ED VISIT OVER THE WEEKEND. FORTUNATELY, WE WERE ABLE TO REPLACE IN THE ED AND DID NOT REQUIRE AN ADMISSION. CATHETER REMOVED AND NEW LINE PLACED. NO PATIENT HARM REPORTED." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21148100 ↗MalfunctionSep 14, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT 4 FR SINGLE LUMEN PICC RUPTURE AT 40CM MARK. RUPTURE FOUND ON LINE REMOVAL THERAPY NO LONGER NEEDED. NO OTHER INFORMATION PROVIDED. ADDITIONAL INFO FROM CUSTOMER: ((B)(6) 2023) CATHETER WAS REMOVED (B)(6) 2023 NO NEW CATHETER PLACED AS THERAPY COMPLETE. PATIENT MENTIONED TO NURSE REMOVING THE LINE THAT SHE HAD A LOT OF ISSUES WITH IT SO RN CLOSELY INSPECTED CATHETER AND FOUND RUPTURE AT 40 CM MARK. IT WAS SECURED WITH SECUR-A-CATH. NO PT HARM REPORTED. WAS BEING TREATED FOR PERSISTENT HEADACHES AND DOUBLE VISION ADJUSTMENT OF PROGRAMMABLE SHUNT. THEY DO NOT HAVE AN INFECTION DISEASE. UNKOWN WHEN CATHETER RUPTURE OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17752916 ↗MalfunctionMar 28, 2023Material Discolored; Defective Component; Material Discolored; Defective Component
IT WAS REPORTED BY THE CUSTOMER, "BROWN DISCOLORATION AFTER USING SCISSORS. THE DISCOLORATION ON THE RING DISAPPEARED WITHIN A COUPLE HOURS. WHILE TRIMMING STATSEAL DISC BROWN DISCOLORATION WAS NOTED TO COME OFF SCISSORS. NEW RING AND SCISSORS OBTAINED BUT PICC HAD ALREADY BEEN PLACED AND SCISSORS USED TO TRIM CATHETER. STATLOCK WAS USED TO SECURE PICC. NO ISSUES AS OF 3/8/2023 WILL CONTINUE TO MONITOR. THERE WAS NO CONCOMITANT THERAPY AS EVENT OCCURRED WITH PICC PLACEMENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16632026 ↗MalfunctionMar 1, 2023Material Discolored; Material Discolored
IT WAS REPORTED BY THE CUSTOMER THAT "SOME SCISSORS FROM THE KIT THAT LEFT DISCOLORATION ON THE STAT SEAL DISC WHEN IT WAS TRIMMED. " ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER: "THE DISCOLORATION WAS BROWN. A NEW DISC AND SCISSORS WERE OBTAINED BUT PICC HAD ALREADY BEEN PLACED AND THE SCISSORS WERE USED TO TRIM CATHETER. PATIENT IS BEING MONITORED FOR POTENTIAL COMPLICATIONS. PATIENT IS BEING TREATED FOR MSSA BACTEREMIA AND PULMONARY VALVE ENDOCARDITIS."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16461378 ↗MalfunctionMar 1, 2023Failure to Infuse; Material Deformation; Failure to Infuse; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "12/22/2022 PT REPORTED THAT HER PICC WOULD OCCLUDE FREQUENTLY AND THAT SHE WOULD HAVE TO ADDUCT HER RIGHT ARM WHILE KEEPING THE PALMS UP TOWARDS THE CEILING. DRESSING CHANGED RN STARTED NS BOLUS AND NO ISSUES OBSERVED. 12/29 REPORTS PICC OCCLUSION ISSUES. X-RAY REQUESTED. ATTEMPTED PULLING LINE BACK TO ELIMINATE LOOP X 2. PICC WAS REMOVED ON (B)(6) AND NEW PICC ORDER PLACED." NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16461011 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page