Midline Catheter Kit S4254108Recalls & reported problems
Bard Access Systems, Inc. · 4Fr DL Provena Midline Full Kit · Class 2
All models of Midline Catheter Kit (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741154096), received between Feb 23, 2026 and Feb 23, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Tear, Rip or Hole in Device Packaging” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 23, 2026Tear, Rip or Hole in Device Packaging
IT WAS REPORTED BY CUSTOMER THAT THE PRODUCT WAS DAMAGED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24423387 ↗
Other Midline catheter devices
- Midline Catheter Kit P4154108DBard Access Systems, Inc. · Midline catheter1 recall1 death126 reports
- ARROW IPN921959Teleflex Incorporated · Midline catheter2 recalls0 reports
- Midline Catheter Kit S4153108BDPBard Access Systems, Inc. · Midline catheter1 recall36 reports
- Midline Catheter Kit S4153108DBard Access Systems, Inc. · Midline catheter1 recall32 reports
- Midline Catheter Kit S4153108DGPBard Access Systems, Inc. · Midline catheter1 recall15 reports
- Midline Catheter Kit S4254108BDPBard Access Systems, Inc. · Midline catheter1 recall11 reports
- Midline Catheter Kit P4154108Bard Access Systems, Inc. · Midline catheter1 recall10 reports
- ARROW IPN037698Teleflex Incorporated · Midline catheter1 recall9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page