Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741154737), received between Nov 6, 2019 and Oct 22, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Activation Problem1
  • Difficult or Delayed Separation1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THEY HAD STRANGE BREAK IN A PICC LINE THAT OCCURRED INSIDE A PATIENT'S ARM. THE LINE AT INSERTION WAS AT 0 AND THE BREAK IN THE LINE WAS AT 2. WHEN HE PRESENTED BACK FROM HOME, THE LINE WAS STILL AT 0, BUT IT DID OCCUR AT HOME SO WE CANNOT ACCOUNT FOR THE LINE THE ENTIRE TIME. THE LINE NEEDED TO BE REMOVED AND REPLACED DUE TO LEAKING INTO PATIENT'S ARM. PATIENT EXPERIENCED PAIN WHEN INFUSING OR LINE WAS FLUSHED. LINE HAS TWO SMALL SLITS, PARTIAL BREAK, WHERE LINE WAS BENT IN ARM. THE CATHETER WAS SUCCESSFULLY REMOVED. NO OTHER INTERVENTION WAS NECESSARY. THE LINE WAS DRESSED AND COVERED WITH A TEGADERM OVER THE INSERTION SITE. THE PATIENT CAME INTO THE ER WITH PAIN ON FLUSHING AND INFUSION BECAUSE OF INABILITY TO TOLERATE USING THE LINE. HE MISSED A DOSE OF ANTIBIOTICS AS A RESULT.

    Patient problems: Pain

    MAUDE report 23359976 ↗
  • MalfunctionApr 24, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY CUSTOMER THAT THEY HAD A POWERPICC THAT HAD A BREAK IN THE LINE NEAR THE HUB. THE PATIENT'S LINE NEEDED TO BE REPLACED. NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21903854 ↗
  • MalfunctionFeb 24, 2025Activation Problem; Activation Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE GREEN SAFETY MECHANISM BECOME UNATTACHED DURING SAFETY DEPLOYMENT. NO OTHER INFORMATION WAS PROVIDED AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21453070 ↗
  • MalfunctionSep 23, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT PICC WAS PLACED ON AN EXTREMELY LARGE PATIENT BACK ON (B)(6). IT WAS DISCONTINUED AND AROUND THE 8CM MARK THE PICC WAS FRACTURED. I SUSPECT THE LINE WAS KINKED IN THE ARM DUE TO PATIENT'S HABITUS. IT IS POSSIBLE THAT A CLINICIAN ATTEMPTED TO RESTORE PATENCY WITH AN INAPPROPRIATELY SIZED SYRINGE IF/WHEN RESISTANCE WAS MET. THE LINE WAS REMOVED PRIOR TO COMPLETION OF THERAPY AND A SHORT PIVC WAS PLACED TO FACILITATE THE LAST DOSE OF ABX. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20287916 ↗
  • MalfunctionApr 22, 2022Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation

    IT WAS REPORTED "LEAKING AT THE HUB." ADDITIONAL INFORMATION RECEIVED 03/31/2022 "THERE WASN T A CRACK IN THE HUB THAT WAS CAUSING A LEAK. IT WAS A MISALIGNMENT OF THE TWO HALVES OF THE HUB THAT CAUSED THE NEEDLELESS CONNECTOR TO NOT SIT CORRECTLY. NO MATTER HOW TIGHT YOU TRIED TO PUT THE CAP ON, THE DEVICE LEAKED WHEN IT WAS FLUSHED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14190392 ↗
  • MalfunctionFeb 16, 2022Difficult or Delayed Separation; Difficult or Delayed Separation

    IT WAS REPORTED "UPON PEELING SHEATH, INNER LUMEN DID NOT COMPLETELY PEEL AND UNABLE TO RELEASE CATHETER" NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13544352 ↗
  • MalfunctionNov 6, 2019Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT PICC WAS FOUND CONSISTENTLY LEAKING INFUSATE AT SITE, NOTED FROM DAY ONE AND HOME HEALTH NURSES HAD TO DO PRN DRESSING CHANGES ALL THE TIME. CXR WAS DONE AND IT WAS POSITIONED IN LOWER SVC. THE LINE WAS PULLED AND A NEW LINE PLACED IN THE OPPOSITE ARM. NO PATIENT HARM REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9285487 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →