PowerPICC SOLO 1194108D1Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC SOLO (78)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741154782), received between Feb 2, 2023 and Jun 26, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Defective Component” (6 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2666-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Mar 11, 2025Class IOpen, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Defective Component6
- Fluid/Blood Leak3
- Activation Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Infiltration into Tissue1
- Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2026Fluid/Blood Leak
IT WAS REPORTED PATIENT RETURNED TO HAVE PICC REPLACED DUE TO DIFFICULTY FLUSHING AND BEING "POSITIONAL" FOR USE. ONCE PICC WAS REMOVED WAS FOUND TO HAVE A FRACTURE AT 7 CM. THE COURSE OF TREATMENT WAS NOT ALTERED AND THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636233 ↗MalfunctionJul 28, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, PATIENT RECEIVED LEFT BRACHIAL SINGLE LUMEN PICC PLACEMENT AND WAS GETTING AT HOME TPN. DUE TO WAKING UP WITH TPN HAVING LEAKED OUT AND BED SIGNIFICANTLY WET. PATIENT ALSO NOTED NO BLOOD RETURN FOR A COUPLE OF DAYS PRIOR. PATIENT COMPLAINED OF PAIN WHEN FLUSHING IN UPPER ARM. PICC LINE WAS REMOVED IN EMERGENCY DEPARTMENT BY VAT RN AND ON INSPECTION NOTED A HOLE AT THE 8CM MARK OF THE CATHETER. LINE WAS REPLACED AND PATIENT DISCHARGED.
Patient problems: Infiltration into Tissue; Pain
MAUDE report 22630303 ↗MalfunctionMay 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PICC LINES WERE DEFECTIVE AND LEAKING. ADDITIONAL INFORMATION: THE PICC WAS REMOVED AND REPLACED 3/14/25. THE PICC WAS DISCARDED AFTER PHOTOS WERE TAKEN. THE PATIENT WAS ALREADY BEING ADMITTED FOR SURGERY AND THE PICC WAS ABLE TO BE REPLACED DURING THAT ADMISSION SO THERE DID NOT END UP BEING A DELAY IN TREATMENT THAT I AM AWARE OF. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22006772 ↗MalfunctionNov 20, 2024Activation Problem; Activation Problem
IT WAS REPORTED BY CUSTOMER THAT GREEN SAFETY COVER CAME OFF OF THE INNER SAFETY COMPONENT ON PICC KIT SAFETY INTRODUCER NEEDLE. AS A RESULT, THE NEEDLE BEVEL COULD NOT BE COVERED. ADDITIONAL INFORMATION RECEIVED ON OCTOBER 22, 2024: NO PATIENT HARM OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20733922 ↗MalfunctionFeb 15, 2023Defective Component; Defective Component
IT WAS REPORTED BY BD SALES REPRESENTATIVE ¿DURING CLINICAL VISIT, STAFF REPORTED THAT AIR IS BEING INTRODUCED INTO THE WIRE CONNECTOR ADAPTER OF THE PICC LINE. A 4 FRENCH SINGLE LUMEN PICC KIT WAS OPENED FOR DEMONSTRATION. THE LINE WAS PRIMED WITH SALINE AND AIR IS ENTERING EITHER FROM THE HOLE THE WIRE PASSES THROUGH OR AT THE CONNECTION TO THE PLASTIC (THE RUBBERIZED PIECE THE WIRE STYLET PASSES THROUGH).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16383547 ↗MalfunctionFeb 14, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER ¿ONE OF OUR PICC NURSES NOTED AIR BEING INTRODUCED INTO THE WIRE CONNECTOR ADAPTER OF THE LINE UPON ASPIRATION DURING ASSESSMENT OF BLOOD RETURN ON THREE SINGLE LUMEN 4 FRENCH PICCS.¿ THIS REPORT ADDRESSES THE THIRD DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16372088 ↗MalfunctionFeb 14, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER ¿ONE OF OUR PICC NURSES NOTED AIR BEING INTRODUCED INTO THE WIRE CONNECTOR ADAPTER OF THE LINE UPON ASPIRATION DURING ASSESSMENT OF BLOOD RETURN ON THREE SINGLE LUMEN 4 FRENCH PICCS.¿ THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16372085 ↗MalfunctionFeb 14, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER ¿ONE OF OUR PICC NURSES NOTED AIR BEING INTRODUCED INTO THE WIRE CONNECTOR ADAPTER OF THE LINE UPON ASPIRATION DURING ASSESSMENT OF BLOOD RETURN ON THREE SINGLE LUMEN 4 FRENCH PICCS.¿ THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16372072 ↗MalfunctionFeb 11, 2023Defective Component; Defective Component
IT WAS REPORTED BY THE CUSTOMER THAT THE YELLOW RUBBER SEAL ON THE T-CONNECTOR SET IS NOT MAKING A COMPLETE SEAL. AIR AND SALINE LEAK WHEN FLUSHING AFTER THE LINE HAS ALREADY BEEN PRIMED. NO OTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION PROVIDED. NO PATIENTS HAVE BEEN HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16359656 ↗MalfunctionFeb 2, 2023Defective Component; Material Deformation; Defective Component; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "YELLOW RUBBER SEAL ON T-CONNECTOR SET NOT MAKING COMPLETE SEAL, AIR AND SALINE LEAKAGE WHEN FLUSHING AFTER LINE HAS ALREADY BEEN PRIMED, USER HAS CONCERN ABOUT AIR ENTERING THE LINE." ADD INFO RCVD 01/18/2023: REGARDING THE DESCRIPTION, ESSENTIALLY ONE OF OUR PICC NURSES NOTED AIR BEING INTRODUCED INTO THE WIRE CONNECTOR ADAPTER OF THE LINE UPON ASPIRATION DURING ASSESSMENT OF BLOOD RETURN WITH THE LATEST BATCH OF SINGLE LUMEN 4 FRENCH PICCS. WE HAD DISCUSSED THIS AND OPENED ANOTHER KIT OF THE SAME LOT, PRIMED A LINE WITH SALINE AND TOOK A VIDEO TO ATTEMPT TO DETERMINE WHAT WAS HAPPENING. IT APPEARS EITHER THE CONNECTION OF THE RUBBERIZED PIECE THE WIRE STYLET PASSES THROUGH IS THE ISSUE, WITH AIR ENTERING FROM EITHER THE HOLE THE WIRE PASSES THROUGH OR AT THE CONNECTION TO THE PLASTIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16288027 ↗
FDA clearance and classification
The FDA regulates product code LJS (catheter,intravascular,therapeutic,long-term greater than 30 days) under 21 CFR 880.5970 as a Class II device. 21 CFR 880.5970 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PowerPICC SOLO 1194108D1
Has the PowerPICC SOLO 1194108D1 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2666-2026 (Class II), began on May 21, 2026: Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the PowerPICC SOLO 1194108D1?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Feb 2, 2023 and Jun 26, 2026. In those reports, the most common specific device problems are “Defective Component” (6), “Fluid/Blood Leak” (3) and “Activation Problem” (1); the most common patient problems are “Infiltration into Tissue” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PowerPICC SOLO 1194108D1?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PowerPICC SOLO 1194108D1 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the PowerPICC SOLO 1194108D1, and is it still sold?
The FDA's device database lists Bard Access Systems, Inc. as the company responsible for it (brand PowerPICC SOLO, model 1194108D1). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,692 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,507 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall398 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls244 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall263 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days276 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page