Skip to content

PowerPICC Provena S1193108D3Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC Provena (116)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741154980), received between Feb 12, 2020 and Dec 30, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Class IIOpen, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2664-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation4
  • Device Markings/Labelling Problem1
  • Naturally Worn1
  • Product Quality Problem1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 30, 2025Product Quality Problem; Product Quality Problem

    IT WAS REPORTED BY CUSTOMER THAT THEY HAVE A DEFECTIVE PICC THAT WE WERE HOPING TO GET REPLACED. THE TAPER STARTED AT THE 6CM MARK, SO IT WOULD NOT ADVANCE INTO THE SHEATH PAST THAT POINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23931578 ↗
  • MalfunctionMar 30, 2024Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED, INVERTED P-WAVE / QRS. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19009029 ↗
  • MalfunctionJan 26, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER "TWO MICROINTRODUCERS BENT WHEN TRYING TO INSERT." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND DEVICE. 1/20/2023 - THE CUSTOMER REPORTED THIS EVENT OCCURRED ON TWO DEVICES HOWEVER ONLY RETURNED ONE DEVICE FOR EVALUATION. THIS FILE WILL ADDRESS THE DEVICE THAT WAS NOT RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16250309 ↗
  • MalfunctionJan 26, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER "TWO MICROINTRODUCERS BENT WHEN TRYING TO INSERT." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST DEVICE. ON (B)(6) 2023 - THE CUSTOMER REPORTED THIS EVENT OCCURRED ON TWO DEVICES HOWEVER ONLY RETURNED ONE DEVICE FOR EVALUATION. THIS FILE WILL ADDRESS THE RETURNED MICROINTRODUCER TIP THAT EXHIBITED A SECTION WITH INWARD BUNCHING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16248339 ↗
  • MalfunctionOct 12, 2022Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER "THE PEEL AWAY SHEATH IS "SERIOUSLY HANGING UP IN SKIN". ONE OCCURRED ON (B)(6) 2022. IT WAS REPLACED WITH A DROP-IN MICROINTRODUCER KIT AND "WORKED LIKE A CHARM."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15589385 ↗
  • MalfunctionFeb 23, 2021Device Markings/Labelling Problem; Naturally Worn

    IT WAS REPORTED THE PROVENA POWERPICCS THAT WE HAVE BEEN USING, IS THAT AFTER A FEW DRESSING CHANGES AND AFTER A FEW WEEKS, THE NUMBERS AND TICK MARKS ON THE OUTSIDE OF THE CATHETER START TO WEAR OFF. ADDITIONAL INFORMATION RECEIVED (B)(6) 2021: WHAT TYPE OF CATHETER SECUREMENT WAS UTILIZED: STAT-LOCK. WAS THERE PATIENT HARM REPORTED?: NO.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11369265 ↗
  • MalfunctionFeb 12, 2020Material Deformation; Material Deformation

    IT WAS REPORTED THAT "ON (B)(6) 2020 THE OPERATOR INVOLVED IN A PICC LINE INSERTION UTILIZING SHERLOCK 3CG GUIDANCE. THE OPERATOR INDICATES THAT HE GAINED VASCULAR ACCESS, TRIMMED THE PICC AND WAS INSERTING UNDER SHERLOCK 3CG GUIDANCE WITH NORMAL TRACKING INDICATED. THE OPERATOR INDICATES THE PICC APPEARED TO BE ALMOST FULLY INSERTED WHEN RESISTANCE WAS FELT AND SHERLOCK TRACKING INDICATED POSSIBLE ¿LOOPING¿ OF CATHETER. THE OPERATOR FLUSHED THE PICC BRISKLY WITH SALINE AND THE ¿LOOP APPEARED TO RESOLVE¿. FINAL TIP LOCATION WAS CONFIRMED UTILIZING 3CG WITH MAX P-WAVE NOTED. UPON REMOVAL OF THE STYLETTE AT THE END OF THE PROCEDURE THE OPERATOR NOTICED A ¿KINK¿ IN THE STYLETTE 5 CM PROXIMAL TO THE NITINOL STYLETTE TIP. THE OPERATOR REPORTS THAT THE STYLETTE WAS INTACT AND THE PROCEDURE WAS COMPLETED WITH THE PICC LEFT IN PLACE. THE OPERATOR DID NOT RETAIN THE STYLETTE OR ANY OTHER COMPONENTS FOR EVALUATION."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9700619 ↗

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page