PowerPICC Provena S1193108D3Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741154980), received between Feb 12, 2020 and Dec 30, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2664-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation4
- Device Markings/Labelling Problem1
- Naturally Worn1
- Product Quality Problem1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2025Product Quality Problem; Product Quality Problem
IT WAS REPORTED BY CUSTOMER THAT THEY HAVE A DEFECTIVE PICC THAT WE WERE HOPING TO GET REPLACED. THE TAPER STARTED AT THE 6CM MARK, SO IT WOULD NOT ADVANCE INTO THE SHEATH PAST THAT POINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23931578 ↗MalfunctionMar 30, 2024Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED, INVERTED P-WAVE / QRS. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19009029 ↗MalfunctionJan 26, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "TWO MICROINTRODUCERS BENT WHEN TRYING TO INSERT." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE SECOND DEVICE. 1/20/2023 - THE CUSTOMER REPORTED THIS EVENT OCCURRED ON TWO DEVICES HOWEVER ONLY RETURNED ONE DEVICE FOR EVALUATION. THIS FILE WILL ADDRESS THE DEVICE THAT WAS NOT RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16250309 ↗MalfunctionJan 26, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "TWO MICROINTRODUCERS BENT WHEN TRYING TO INSERT." NO OTHER INFORMATION WAS PROVIDED. THIS REPORT ADDRESSES THE FIRST DEVICE. ON (B)(6) 2023 - THE CUSTOMER REPORTED THIS EVENT OCCURRED ON TWO DEVICES HOWEVER ONLY RETURNED ONE DEVICE FOR EVALUATION. THIS FILE WILL ADDRESS THE RETURNED MICROINTRODUCER TIP THAT EXHIBITED A SECTION WITH INWARD BUNCHING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16248339 ↗MalfunctionOct 12, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER "THE PEEL AWAY SHEATH IS "SERIOUSLY HANGING UP IN SKIN". ONE OCCURRED ON (B)(6) 2022. IT WAS REPLACED WITH A DROP-IN MICROINTRODUCER KIT AND "WORKED LIKE A CHARM."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15589385 ↗MalfunctionFeb 23, 2021Device Markings/Labelling Problem; Naturally Worn
IT WAS REPORTED THE PROVENA POWERPICCS THAT WE HAVE BEEN USING, IS THAT AFTER A FEW DRESSING CHANGES AND AFTER A FEW WEEKS, THE NUMBERS AND TICK MARKS ON THE OUTSIDE OF THE CATHETER START TO WEAR OFF. ADDITIONAL INFORMATION RECEIVED (B)(6) 2021: WHAT TYPE OF CATHETER SECUREMENT WAS UTILIZED: STAT-LOCK. WAS THERE PATIENT HARM REPORTED?: NO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11369265 ↗MalfunctionFeb 12, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT "ON (B)(6) 2020 THE OPERATOR INVOLVED IN A PICC LINE INSERTION UTILIZING SHERLOCK 3CG GUIDANCE. THE OPERATOR INDICATES THAT HE GAINED VASCULAR ACCESS, TRIMMED THE PICC AND WAS INSERTING UNDER SHERLOCK 3CG GUIDANCE WITH NORMAL TRACKING INDICATED. THE OPERATOR INDICATES THE PICC APPEARED TO BE ALMOST FULLY INSERTED WHEN RESISTANCE WAS FELT AND SHERLOCK TRACKING INDICATED POSSIBLE ¿LOOPING¿ OF CATHETER. THE OPERATOR FLUSHED THE PICC BRISKLY WITH SALINE AND THE ¿LOOP APPEARED TO RESOLVE¿. FINAL TIP LOCATION WAS CONFIRMED UTILIZING 3CG WITH MAX P-WAVE NOTED. UPON REMOVAL OF THE STYLETTE AT THE END OF THE PROCEDURE THE OPERATOR NOTICED A ¿KINK¿ IN THE STYLETTE 5 CM PROXIMAL TO THE NITINOL STYLETTE TIP. THE OPERATOR REPORTS THAT THE STYLETTE WAS INTACT AND THE PROCEDURE WAS COMPLETED WITH THE PICC LEFT IN PLACE. THE OPERATOR DID NOT RETAIN THE STYLETTE OR ANY OTHER COMPONENTS FOR EVALUATION."
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9700619 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record