Skip to content

PowerPICC Provena S1173108D5Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC Provena (116)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741155130), received between Oct 9, 2020 and Sep 2, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Class IIOpen, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2664-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation5
  • Difficult to Insert2
  • Break1
  • Difficult to Remove1
  • Fracture1
  • Material Frayed1
  • Material Too Rigid or Stiff1
  • Structural Problem1

Patient problems

  • Pain4
  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1
  • Swelling1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjurySep 2, 2026Break

    Patient problems: Foreign Body In Patient

    MAUDE report 50565729 ↗
  • MalfunctionJan 19, 2024Material Frayed; Material Frayed

    IT WAS REPORTED BY THE CUSTOMER THE CLINICIAN INSERTED A PICC ON (B)(6) 2023. HE WAS ABLE TO PUNCTURE THE VEIN, BUT UNABLE TO ADVANCE THE GUIDEWIRE, SO HE PULLED IT BACK TOGETHER W/ THE NEEDLE, BUT MET RESISTANCE AND NOTED A SHEAR AT THE ELASTIC COIL PART OF THE GUIDEWIRE. HE THEN DECIDED TO NICK THE PATIENT'S SKIN AND WAS ABLE TO SUCCESSFULLY REMOVE BOTH THE NEEDLE AND THE GUIDEWIRE. I DROPPED ANOTHER NEEDLE AND HE WAS ABLE TO SUCCESSFULLY INSERT THE PICC LINE. RESPONSE RECEIVED 01/09/2024: AS FAR AS WE KNOW, THERE ARE NO HEALTH CONSEQUENCES RELATED TO THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18545582 ↗
  • MalfunctionAug 9, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED BY CUSTOMER, THAT GUIDEWIRE WOULD NOT ADVANCE. WHEN ATTEMPTING TO PULL OUT IT, GOT STUCK. A TISSUE SUBSTANCE APPEARS TO BE STUCK ON WIRE CLOSE TO TIP. PATIENT NEEDED GUIDEWIRE REMOVED BY TEAM OF PHYSICIANS. PATIENT NOT HARMED. NO OTHER INFORMATION PROVIDED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17490895 ↗
  • MalfunctionAug 10, 2022Material Deformation; Material Deformation

    IT WAS REPORTED "THE INTRODUCER TOO EASILY BUCKLING AND BECOMING BENT OR THE GREY PEEL-A-WAY END BEGINS TO CURL BACK ON ITSELF PREVENTING THE INTRODUCER FROM ENTERING THE VEIN WALL. WE WERE ABLE TO EASILY INSERT THE INTRODUCER THROUGH THE SKIN OVER THE GUIDE WIRE FOR ABOUT 1 INCH AFTER MAKING A SKIN NICK. HOWEVER, WHEN WE TRIED TO ADVANCE THE INTRODUCER THROUGH THE VEIN, WE MET RESISTANCE. WHEN WE TRIED TO ADVANCE THE INTRODUCER THROUGH THE VEIN, WE MET RESISTANCE. WE MADE A LARGER NICK BUT HAD THE SAME OUTCOME. WHEN WE REMOVED THE INTRODUCER I COULD SEE AND FEEL THE GRAY EDGE HAD ROLLED BACK ON ITSELF. WE HAD TO OPEN AN ADDITIONAL PICC KITS TO RETRIEVE ANOTHER INTRODUCER. WHEN WE RETRIEVED A NEW INTRODUCER, IT ADVANCED EASILY LIKE IT'S SUPPOSED TO."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15204199 ↗
  • MalfunctionJun 8, 2021Material Deformation; Material Deformation

    IT WAS REPORTED AN INTRODUCER BUNCHING UP. WE HAD TO USE THE INTRODUCER FROM THE MICROINTRODUCER KIT TO COMPLETE LINE INSERTION. THIS RESULTED IN BLEEDING AT THE SITE WHERE PRESSURE HAD TO HELD FOR 5-10 MINUTES BEFORE DRESSING COULD BE APPLIED. WE WERE PLACING A SINGLE LUMEN AT THE TIME. ADD INFO RCVD 01/25/2021: UNABLE TO USE PRODUCT AFTER FIRST ATTEMPT, PRODUCT SECURED, INTRODUCER FROM MICROINTRODUCER KIT USED TO PLACE LINE. NO PATIENT HARM WAS NOTED OR REPORTED, PRESSURE HAD TO BE HELD AT SITE FOR GREATER THAN 10 MIN BEFORE DRESSING COULD BE APPLIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11960301 ↗
  • MalfunctionNov 4, 2020Difficult to Insert; Material Deformation

    IT WAS REPORTED THE MICROINTRODUCER WAS NOT SMOOTH AND HAD BULGED EDGES WHICH CAUSED THE PATIENT PAIN UPON INSERTION. ADDITIONAL INFORMATION RECEIVED 10-06-2020: "DIAGNOSIS: BRAIN ABSCESS, NO INFECTIOUS DISEASE, 'WHEN PLACING PICC HAD TROUBLE WITH MICROINTRODUCER; WHEN INSPECTING INTRODUCER NOTICED BULGED EDGES. DUE TO THE BULGED EDGE UNABLE TO GET INTRODUCER THROUGH THE SKIN OF PATIENT. WE HAD TO DROP A MICROEZ MICROINTRODUCER KIT TO FINISH THE PICC LINE PLACEMENT. THIS HAS HAPPENED ON SEVERAL OCCASIONS."

    Patient problems: Pain

    MAUDE report 10787836 ↗
  • MalfunctionOct 29, 2020Structural Problem

    IT WAS REPORTED THE MICROINTRODUCER IS NOT SMOOTH AND HAS BULGED EDGES. IT WAS STATED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: Pain

    MAUDE report 10758422 ↗
  • MalfunctionOct 29, 2020Material Too Rigid or Stiff

    IT WAS REPORTED THE MICRO INTRODUCER IS NOT SMOOTH AND HAS BULGED EDGES. IT WAS STATED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.

    Patient problems: Pain

    MAUDE report 10758415 ↗
  • MalfunctionOct 29, 2020Material Deformation

    IT WAS REPORTED THE MICROINTRODUCER IS NOT SMOOTH AND HAS BULGED EDGES. IT WAS STATED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES THE THIRD DEVICE.

    Patient problems: Pain

    MAUDE report 10758307 ↗
  • MalfunctionOct 23, 2020Difficult to Insert; Material Deformation

    IT WAS REPORTED "INTRODUCER WILL NOT ENTER THE SKIN UPON INSPECTION THERE IS A PROBLEM WITH THIS IT HAS A BEVELED EDGE INSTEAD OF A FLAT EDGE".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10731392 ↗
  • MalfunctionOct 9, 2020Fracture

    IT WAS REPORTED "POSSIBLE MICRO FRACTURE INTERNAL LENGTH 1-3 CM PROXIMAL TO INSERTION SITE WITH 3FR SL POWERPICC". ADDITIONAL INFO. RECEIVED 09/18/2020: "9/16-RIGHT BASILIC 3FR SL PICC PLACED @ BEDSIDE WITHOUT DIFFICULTY, NO ADVERSE EVENTS DURING PROCEDURE, NO COMPLAINTS OF DISCOMFORT FROM PATIENT DURING LINE PLACEMENT. ONE HOUR POST LINE PLACEMENT, IV ANTIBIOTIC INFUSION WAS STARTED USING INFUSION PUMP. PATIENTS RIGHT ARM 6-8 CM DISTAL TO INSERTION SITE BEGAN TO SWELL WHICH CREATED SOME PATIENT DISCOMFORT. NURSE WAS NOTIFIED BY PATIENT, AS PUMP DID NOT ALARM. NURSE NOTED THE SWELLING AND STOPPED INFUSION. INFUSION WAS RESTARTED AT A SLOWER RATE. SWELLING APPEARED EVEN WHEN INFUSION WAS DECREASED. 9/17-VAT NOTIFIED AND ASKED TO ASSESS SITE. US DOPPLER ORDERED TO R/O THROMBUS WHICH WAS NEGATIVE, PRIOR TO CONTACTING VAT.THERE WAS NOTED SWELLING COMPARED TO LEFT ARM EVEN AFTER INFUSION HAD BEEN ON HOLD FOR 4+ HOURS. VAT COLLABORATED WITH PROVIDER, BEDSIDE NURSE, AND PATIENT. LINE REMOVED AND SAVED FOR EVALUATION. " "WAS THERE PATIENT HARM REPORTED?: NO HARM BUT DELAY IN DISCHARGE."

    Patient problems: Swelling

    MAUDE report 10659357 ↗

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices