PowerPICC Provena S1173108D5Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC Provena (116)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00801741155130), received between Oct 9, 2020 and Sep 2, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2026Class IIOpen, Classified
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Z-2664-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation5
- Difficult to Insert2
- Break1
- Difficult to Remove1
- Fracture1
- Material Frayed1
- Material Too Rigid or Stiff1
- Structural Problem1
Patient problems
- Pain4
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- Swelling1
- Unspecified Tissue Injury1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjurySep 2, 2026Break
Patient problems: Foreign Body In Patient
MAUDE report 50565729 ↗MalfunctionJan 19, 2024Material Frayed; Material Frayed
IT WAS REPORTED BY THE CUSTOMER THE CLINICIAN INSERTED A PICC ON (B)(6) 2023. HE WAS ABLE TO PUNCTURE THE VEIN, BUT UNABLE TO ADVANCE THE GUIDEWIRE, SO HE PULLED IT BACK TOGETHER W/ THE NEEDLE, BUT MET RESISTANCE AND NOTED A SHEAR AT THE ELASTIC COIL PART OF THE GUIDEWIRE. HE THEN DECIDED TO NICK THE PATIENT'S SKIN AND WAS ABLE TO SUCCESSFULLY REMOVE BOTH THE NEEDLE AND THE GUIDEWIRE. I DROPPED ANOTHER NEEDLE AND HE WAS ABLE TO SUCCESSFULLY INSERT THE PICC LINE. RESPONSE RECEIVED 01/09/2024: AS FAR AS WE KNOW, THERE ARE NO HEALTH CONSEQUENCES RELATED TO THE INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18545582 ↗MalfunctionAug 9, 2023Difficult to Remove; Difficult to Remove
IT WAS REPORTED BY CUSTOMER, THAT GUIDEWIRE WOULD NOT ADVANCE. WHEN ATTEMPTING TO PULL OUT IT, GOT STUCK. A TISSUE SUBSTANCE APPEARS TO BE STUCK ON WIRE CLOSE TO TIP. PATIENT NEEDED GUIDEWIRE REMOVED BY TEAM OF PHYSICIANS. PATIENT NOT HARMED. NO OTHER INFORMATION PROVIDED.
Patient problems: Unspecified Tissue Injury
MAUDE report 17490895 ↗MalfunctionAug 10, 2022Material Deformation; Material Deformation
IT WAS REPORTED "THE INTRODUCER TOO EASILY BUCKLING AND BECOMING BENT OR THE GREY PEEL-A-WAY END BEGINS TO CURL BACK ON ITSELF PREVENTING THE INTRODUCER FROM ENTERING THE VEIN WALL. WE WERE ABLE TO EASILY INSERT THE INTRODUCER THROUGH THE SKIN OVER THE GUIDE WIRE FOR ABOUT 1 INCH AFTER MAKING A SKIN NICK. HOWEVER, WHEN WE TRIED TO ADVANCE THE INTRODUCER THROUGH THE VEIN, WE MET RESISTANCE. WHEN WE TRIED TO ADVANCE THE INTRODUCER THROUGH THE VEIN, WE MET RESISTANCE. WE MADE A LARGER NICK BUT HAD THE SAME OUTCOME. WHEN WE REMOVED THE INTRODUCER I COULD SEE AND FEEL THE GRAY EDGE HAD ROLLED BACK ON ITSELF. WE HAD TO OPEN AN ADDITIONAL PICC KITS TO RETRIEVE ANOTHER INTRODUCER. WHEN WE RETRIEVED A NEW INTRODUCER, IT ADVANCED EASILY LIKE IT'S SUPPOSED TO."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15204199 ↗MalfunctionJun 8, 2021Material Deformation; Material Deformation
IT WAS REPORTED AN INTRODUCER BUNCHING UP. WE HAD TO USE THE INTRODUCER FROM THE MICROINTRODUCER KIT TO COMPLETE LINE INSERTION. THIS RESULTED IN BLEEDING AT THE SITE WHERE PRESSURE HAD TO HELD FOR 5-10 MINUTES BEFORE DRESSING COULD BE APPLIED. WE WERE PLACING A SINGLE LUMEN AT THE TIME. ADD INFO RCVD 01/25/2021: UNABLE TO USE PRODUCT AFTER FIRST ATTEMPT, PRODUCT SECURED, INTRODUCER FROM MICROINTRODUCER KIT USED TO PLACE LINE. NO PATIENT HARM WAS NOTED OR REPORTED, PRESSURE HAD TO BE HELD AT SITE FOR GREATER THAN 10 MIN BEFORE DRESSING COULD BE APPLIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11960301 ↗MalfunctionNov 4, 2020Difficult to Insert; Material Deformation
IT WAS REPORTED THE MICROINTRODUCER WAS NOT SMOOTH AND HAD BULGED EDGES WHICH CAUSED THE PATIENT PAIN UPON INSERTION. ADDITIONAL INFORMATION RECEIVED 10-06-2020: "DIAGNOSIS: BRAIN ABSCESS, NO INFECTIOUS DISEASE, 'WHEN PLACING PICC HAD TROUBLE WITH MICROINTRODUCER; WHEN INSPECTING INTRODUCER NOTICED BULGED EDGES. DUE TO THE BULGED EDGE UNABLE TO GET INTRODUCER THROUGH THE SKIN OF PATIENT. WE HAD TO DROP A MICROEZ MICROINTRODUCER KIT TO FINISH THE PICC LINE PLACEMENT. THIS HAS HAPPENED ON SEVERAL OCCASIONS."
Patient problems: Pain
MAUDE report 10787836 ↗MalfunctionOct 29, 2020Structural Problem
IT WAS REPORTED THE MICROINTRODUCER IS NOT SMOOTH AND HAS BULGED EDGES. IT WAS STATED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES THE SECOND DEVICE.
Patient problems: Pain
MAUDE report 10758422 ↗MalfunctionOct 29, 2020Material Too Rigid or Stiff
IT WAS REPORTED THE MICRO INTRODUCER IS NOT SMOOTH AND HAS BULGED EDGES. IT WAS STATED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES THE FIRST DEVICE.
Patient problems: Pain
MAUDE report 10758415 ↗MalfunctionOct 29, 2020Material Deformation
IT WAS REPORTED THE MICROINTRODUCER IS NOT SMOOTH AND HAS BULGED EDGES. IT WAS STATED THIS OCCURRED WITH THREE DEVICES. THIS REPORT ADDRESSES THE THIRD DEVICE.
Patient problems: Pain
MAUDE report 10758307 ↗MalfunctionOct 23, 2020Difficult to Insert; Material Deformation
IT WAS REPORTED "INTRODUCER WILL NOT ENTER THE SKIN UPON INSPECTION THERE IS A PROBLEM WITH THIS IT HAS A BEVELED EDGE INSTEAD OF A FLAT EDGE".
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10731392 ↗MalfunctionOct 9, 2020Fracture
IT WAS REPORTED "POSSIBLE MICRO FRACTURE INTERNAL LENGTH 1-3 CM PROXIMAL TO INSERTION SITE WITH 3FR SL POWERPICC". ADDITIONAL INFO. RECEIVED 09/18/2020: "9/16-RIGHT BASILIC 3FR SL PICC PLACED @ BEDSIDE WITHOUT DIFFICULTY, NO ADVERSE EVENTS DURING PROCEDURE, NO COMPLAINTS OF DISCOMFORT FROM PATIENT DURING LINE PLACEMENT. ONE HOUR POST LINE PLACEMENT, IV ANTIBIOTIC INFUSION WAS STARTED USING INFUSION PUMP. PATIENTS RIGHT ARM 6-8 CM DISTAL TO INSERTION SITE BEGAN TO SWELL WHICH CREATED SOME PATIENT DISCOMFORT. NURSE WAS NOTIFIED BY PATIENT, AS PUMP DID NOT ALARM. NURSE NOTED THE SWELLING AND STOPPED INFUSION. INFUSION WAS RESTARTED AT A SLOWER RATE. SWELLING APPEARED EVEN WHEN INFUSION WAS DECREASED. 9/17-VAT NOTIFIED AND ASKED TO ASSESS SITE. US DOPPLER ORDERED TO R/O THROMBUS WHICH WAS NEGATIVE, PRIOR TO CONTACTING VAT.THERE WAS NOTED SWELLING COMPARED TO LEFT ARM EVEN AFTER INFUSION HAD BEEN ON HOLD FOR 4+ HOURS. VAT COLLABORATED WITH PROVIDER, BEDSIDE NURSE, AND PATIENT. LINE REMOVED AND SAVED FOR EVALUATION. " "WAS THERE PATIENT HARM REPORTED?: NO HARM BUT DELAY IN DISCHARGE."
Patient problems: Swelling
MAUDE report 10659357 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page