Skip to content

PowerPICC CK000285CRecalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741161117), received between Nov 11, 2021 and Nov 11, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Material Frayed” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Material Frayed1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2021Material Frayed; Material Frayed

    IT WAS REPORTED DURING PLACING A 4F SL PICC AND 2.7CM OF THE END OF THE WIRE BROKE OFF INTO THE PATIENT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 12797081 ↗

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates