PowerPICC SOLO CK000289CRecalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801741161148), received between Dec 22, 2020 and Dec 19, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Open, Classified

    Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

    Z-2666-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

  2. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Material Twisted/Bent2
  • Break1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED PATIENT PRESENTED TO ED REPORTING LEAKING FROM INSERTION SITE OF PICC LINE. LINE WAS REMOVED AND EVALUATED BY VAS RN. FOUND TRANSVERSE FRACTURE AT 7CM MARK. PICC LINE REMOVED, REPLACED WITH MST MIDLINE TO FINISH REMAINDER OF IV ANTIBIOTICS TREATMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23851348 ↗
  • MalfunctionNov 10, 2025Material Frayed; Material Twisted/Bent; Material Frayed; Material Twisted/Bent

    IT WAS REPORTED THAT ON 10/20/2025, WHILE INSERTING A 4FR, SINGLE LUMEN PICC, I NOTICED THE SHERLOCK 3CG TIP CONFIRMATION COPPER TIP, WAS ABOUT HALF THE SIZE IT USUALLY IS. IT ALSO SEEMED BENT A BIT. I DID NOT CUT THIS TIP. THE WIRE WAS PULLED BACK QUITE A BIT, PRIOR TO TRIMMING, AS ALWAYS. I DID NOT INSERT THIS PICC BUT HAD TO OPEN ANOTHER KIT FOR A NEW PICC, SAME LOT # AND THE TIP WAS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23512631 ↗
  • MalfunctionJun 9, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED BY CUSTOMER THAT CRACKED PICC LINE. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22170564 ↗
  • MalfunctionMay 5, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED VIA MEDWATCH, PICC LINE LEAKING AT THE INSERTION SITE. REMOVED AND INSPECTED. TRANSVERSE CRACK NOTED ALONG PICC LINE AT THE 6CM MARK. NEW PICC LINE WAS INSERTED FOR PATIENT TO CONTINUE IV THERAPY. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21964592 ↗
  • MalfunctionDec 22, 2020Break; Material Twisted/Bent

    IT WAS REPORTED FAILED ATTEMPT ON PICC PLACEMENT AS THE 3CG WIRE WAS FOUND BENT/BROKEN AT THE DISTAL TIP. NO OTHER INFORMATION WAS PROVIDED. ON (B)(6) 2020-- DEVICE RETURNED FOR EVALUATION WAS FOUND KINKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11060954 ↗