Halo One™ Thin-Walled Guiding Sheath HLO51035Recalls & reported problems
Clearstream Technologies Limited · Halo One 5F guiding sheath diameter, 10 cm shaft length, including 035 Dilator · Class 2
All models of Halo One Thin-Walled Guiding Sheath (27)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00801741161889), received between Dec 16, 2021 and Sep 11, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation6
- Failure to Advance4
- Material Split, Cut or Torn2
- Unraveled Material2
- Connection Problem1
- Detachment of Device or Device Component1
- Device-Device Incompatibility1
- Difficult to Insert1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Failure to Advance; Detachment of Device or Device Component; Material Deformation
On (b)(6) 2026, a patient underwent a recanalization procedure using a Halo One device. It was reported that, during the procedure, the healthcare professional had difficulty inserting the 5 to 10 from the same side. After four or five attempts, the tip became frayed and unusable. There was a part of the sheath that detached. The procedure was completed using another device. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50728539 ↗MalfunctionMay 23, 2025Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
ON (B)(6) 2025, A PATIENT WAS UNDERGOING A RECANALIZATION PROCEDURE USING HALO ONE SHEATH. DURING THE PROCEDURE, WHEN ATTEMPTING TO INSERT THE PRODUCT INTO THE IPSILATERAL COMMON FEMORAL ARTERY IN A CASE OF CALCIFICATION, RESISTANCE WAS ENCOUNTERED AND INSERTION INTO THE BLOOD VESSEL WAS NOT POSSIBLE. THE PRODUCT WAS REMOVED FROM THE BODY AND THE TIP WAS FOUND TO BE BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22078458 ↗InjuryJan 10, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, IT WAS ALLEGEDLY DIFFICULT TO BE PUNCTURED AND INSERT THE SHEATH. IT WAS FURTHER REPORTED THAT THE PATIENT ALLEGEDLY EXPERIENCED MORE BLEEDING THAN USUAL. REPORTEDLY, THE BLEEDING DID NOT STOP EASILY, AFTER USING THE HEMOSTATIC DEVICE AND THE BLOOD PRESSURE DROPPED. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21122654 ↗MalfunctionDec 24, 2024Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE INTRODUCER SHEATH ALLEGEDLY COULD NOT BE INSERTED. IT WAS FURTHER REPORTED THAT THE TIP OF THE INTRODUCER SHEATH WAS ALLEGEDLY DAMAGED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21010202 ↗MalfunctionOct 21, 2024Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA IPSILATERAL APPROACH, THE INTRODUCER SHEATH ALLEGEDLY HAD RESISTANCE WHILE INSERTING. IT WAS FURTHER REPORTED THAT THE TIP OF THE INTRODUCER SHEATH WAS ALLEGEDLY TURNED OVER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20491020 ↗MalfunctionMar 5, 2024Connection Problem; Material Deformation; Connection Problem; Material Deformation
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE SUPERIOR FEMORAL ARTERY VIA AN IPSILATERAL ANTEGRADE APPROACH, THE DEVICE SHEATH ITSELF STOPPED THE BLEEDING AND THE VALVE CONNECTION HAD ALLEGEDLY CAME LOOSE. IT WAS FURTHER REPORTED THAT THE SCREW THREAD ON THE SHEATH SIDE WAS ALLEGEDLY CRUSHED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18843091 ↗MalfunctionMay 8, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE OF A HIGHLY CALCIFIED TARGET LESION, THE SHEATH WAS ALLEGEDLY FOUND TO BE TORN WHEN THE PHYSICIAN NOTICED THAT THE ABNORMAL FORM OF SHEATH UNDER FLUOROSCOPY AND REMOVED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16885594 ↗MalfunctionDec 13, 2022Unraveled Material; Material Split, Cut or Torn; Unraveled Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING PREPARATION OF A RECANALIZATION PROCEDURE, THE TIP OF THE SHEATH WAS ALLEGEDLY FOUND PEELED UPON OPENING THE PACKAGE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15966550 ↗MalfunctionDec 16, 2021Unraveled Material; Device-Device Incompatibility; Material Deformation; Unraveled Material; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT SUPERFICIAL FEMORAL ARTERY, THE SHEATH WAS ALLEGEDLY HARD TO BE INSERTED OVER THE GUIDEWIRE. IT WAS FURTHER REPORTED THAT THE TIP OF THE SHEATH WAS FOUND TO BE ALLEGEDLY DAMAGED AND PEELED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13015678 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page