Skip to content

PowerPICC CK000870ARecalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC (118)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741184246), received between Aug 21, 2025 and Feb 19, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Separation” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 11, 2025Open, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Difficult or Delayed Separation2
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2026Product Quality Problem

    IT WAS REPORTED BY CUSTOMER THAT WHEN THEY WERE PLACING A PICC THE AMPULE FOR LIDOCAINE INJECTION SHATTERED WHEN THE NURSING WAS SNAPPING THE TOP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24389335 ↗
  • MalfunctionSep 18, 2025Difficult or Delayed Separation; Difficult or Delayed Separation

    IT WAS REPORTED VIA MEDWATCH THAT WHILE PLACING PICC MICRO EZ MICRO INTRODUCER 4.5F FAILED TO BREAK AWAY AT ORANGE HUB. SHEATH DID NOT BREAK AT ALL. PICC PULLED BACK AND EXCHANGED OVER GUIDEWIRE SUCCESSFULLY. PROCEDURE COMPLETED WITHOUT DIFFICULTY USING 4.5 MICRO EZ MICRO INTRODUCER KIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23085842 ↗
  • MalfunctionAug 21, 2025Difficult or Delayed Separation; Difficult or Delayed Separation

    IT WAS REPORTED THAT THE PICC WAS PLACED AS PER POLICY AND PROCEDURE AT BEDSIDE. UNABLE TO PEEL AWAY MICROEZ MICROINTRODUCER, 4.5F (1.6MM ID X 2.3 MM OD X 7CM LENGTH). MICROINTRODUCER DID NOT BREAK AT HUB AS IT IS SUPPOSED TO DO. INSTEAD, ONE OF THE ORANGE WINGS CRACKED OFF AND THE ORANGE PIECE CIRCLING THE TOP OF THE SHEATH STAYED INTACT MAKING IT IMPOSSIBLE TO PEEL AWAY THE SHEATH AND COMPLETE PICC INSERTION. PICC EXCHANGED OVER GUIDEWIRE WITHOUT DIFFICULTY, PROCEDURE COMPLETED. MAX BARRIER MAINTAINED THROUGHOUT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22866309 ↗

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page