Provena Midline CK000917Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2
All models of Provena Midline (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741187827), received between Mar 24, 2020 and May 6, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation6
- Difficult or Delayed Separation1
- Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2026Material Deformation
IT WAS REPORTED THE SHEATH DOES NOT ENTER THE SKIN AS SMOOTHLY AS BEFORE AND TENDS TO CRUMPLE UP. NO INJURY TO PATIENT HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094544 ↗MalfunctionMay 6, 2026Material Deformation
IT WAS REPORTED THE SHEATH DOES NOT ENTER THE SKIN AS SMOOTHLY AS BEFORE AND TENDS TO CRUMPLE UP. NO INJURY TO PATIENT HAS BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094541 ↗MalfunctionJan 22, 2026Material Deformation; Difficult or Delayed Separation
IT WAS REPORTED THAT THE DILATOR BROKE APART WHEN BEING USED IN THE PATIENT. NO PATIENT HARM WAS OBSERVED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24147969 ↗MalfunctionDec 9, 2025Misassembled; Misassembled
IT WAS REPORTED A NEEDLE OUTSIDE OF THE MIDLINE KIT. FORTUNATELY, NO ONE WAS POKED/INJURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23758397 ↗MalfunctionDec 12, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT 3FR DILATORS ARE BUCKLING WHEN TRYING TO TRANSITION THROUGH SKIN. ADDITIONAL INFORMATION 6/19 IN INABILITY TO ADVANCE DILATOR AND SHEATH THROUGH SKIN REQUIRED ADDITIONAL SKIN NICK AND OBTAINING NEW 3 FR. DILATOR. NO URGENT/LIFE THREATING MEDICAL SITUATION. 12/08/2023 - TWO SAMPLES WERE RETURNED FOR INVESTIGATION. THE LEADING EDGE OF EACH MICROINTRODUCER WAS DAMAGED AND PLASTICALLY DEFORMED ON BOTH SAMPLES. THIS REPORT ADDRESSES THE SECOND SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18311960 ↗MalfunctionJun 30, 2023Material Deformation; Material Deformation
IT WAS REPORTED BY CUSTOMER THAT 3FR DILATORS ARE BUCKLING WHEN TRYING TO TRANSITION THROUGH SKIN. ADDITIONAL INFORMATION 6/19, IN INABILITY TO ADVANCE DILATOR AND SHEATH THROUGH SKIN REQUIRED ADDITIONAL SKIN NICK AND OBTAINING NEW 3 FR. DILATOR. NO URGENT/LIFE THREATING MEDICAL SITUATION. ON 12/08/2023 - TWO SAMPLES WERE RETURNED FOR INVESTIGATION. THE LEADING EDGE OF EACH MICROINTRODUCER WAS DAMAGED AND PLASTICALLY DEFORMED ON BOTH SAMPLES. THIS REPORT ADDRESSES THE FIRST SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17239833 ↗MalfunctionMar 24, 2020Material Deformation; Material Deformation
IT WAS REPORTED "DILATOR PEEL APART SHEATH, ISSUES CONTINUE, I WAS ABLE TO SALVAGE THIS 3 FR. MIDLINE IV DILATATOR, UNABLE TO USE THIS DILATOR DESPITE SKIN NICK. I AVOID SKIN NICKING." ADDITIONAL INFORMATION RECEIVED: 03/04/2020 - ATTEMPTING TO ADVANCE DILATOR AND SHEATH OVER WIRE, DEVICE WOULD NOT PENETRATE SKIN, BUCKLED AFTER MAKING SKIN NICK CONTINUED DIFFICULTY ADVANCING AFTER 3RD ATTEMPT TO ADVANCE SHEATH BECAME ACCORDION LIKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9876077 ↗
Other Midline catheter devices
- Midline Catheter Kit P4154108DBard Access Systems, Inc. · Midline catheter1 recall1 death126 reports
- ARROW IPN921959Teleflex Incorporated · Midline catheter2 recalls0 reports
- Midline Catheter Kit S4153108BDPBard Access Systems, Inc. · Midline catheter1 recall36 reports
- Midline Catheter Kit S4153108DBard Access Systems, Inc. · Midline catheter1 recall32 reports
- Midline Catheter Kit S4153108DGPBard Access Systems, Inc. · Midline catheter1 recall15 reports
- Midline Catheter Kit S4254108BDPBard Access Systems, Inc. · Midline catheter1 recall11 reports
- Midline Catheter Kit P4154108Bard Access Systems, Inc. · Midline catheter1 recall10 reports
- ARROW IPN037698Teleflex Incorporated · Midline catheter1 recall9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page