Skip to content

PowerPICC SOLO CK000882Recalls & reported problems

Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant

All models of PowerPICC SOLO (78)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741188794), received between Oct 14, 2022 and Oct 14, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 11, 2025Class IOpen, Classified

    Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

    Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 14, 2022Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER "SENTINEL EVENT ASSOCIATED WITH PICC INSERTION DUE TO THE PICC 'LOOPING AND TYING KNOT' IN PATIENT'S VASCULATURE". ADDITIONAL INFORMATION RECEIVED 10/12/2022: NO SECUREMENT DEVICE BECAUSE PRODUCT WAS BEING REMOVED DURING INSERTION. UNSUCCESSFUL INSERTION. PLACEMENT DATE:ON (B)(6) 2022. EXPLANT DATE: ON (B)(6) 2022- CUT DOWN PERFORMED IN OR PATIENT INFO: NO HARM- PATIENT HAD TO BE SENT TO OR FOR REMOVAL. PATIENT WAS BEING TREATED WITH ANTIBIOTICS FOR INFECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15611528 ↗

Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices

All Catheter, intravascular, therapeutic, long-term greater than 30 days devices →

Sources and updates