PowerPICC CK000986Recalls & reported problems
Bard Access Systems, Inc. · Catheter Placement Kit · Class 2 · Implant
All models of PowerPICC (118)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801741216008), received between Oct 7, 2022 and Mar 17, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 11, 2025Open, Classified
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
Z-1507-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Material Deformation2
- Activation Problem1
- Contamination /Decontamination Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED VIA MEDWATCH "PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) LINE WAS FRACTURED ABOUT 2-3CM FROM THE CATHETER HUB. PATIENT NEEDED FOR 4-6 WEEKS OF ANTIBIOTICS (ABX) THERAPY AND LINE HAD TO BE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21625390 ↗MalfunctionMar 7, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED PICC TEAM WENT TO ASSESS A 4FR SL LINE AND FOUND A FRACTURE AROUND THE 5CM MARK. MEDICATION LEAKING IN VESSEL. IT IS UNKNOWN IF ANY HARM WAS CAUSED. PATIENT HAD TO BE STUCK AGAIN, LINE HAD TO REMOVED AND REPLACED CAUSING DELAYS IN MEDICATION ADMINISTRATION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21547833 ↗MalfunctionNov 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED BY CUSTOMER THAT PATIENT'S PICC WAS LEAKING AT THE INSERTION SITE OF A SINGLE LUMEN 4FR PICC. PATIENT WAS RECEIVING CHEMOTHERAPY THROUGH THE LINE. WHEN THEY PULLED THE PICC THEY NOTICED IT WAS FRACTURED 3 CM FROM THE HUB. THE VAT INSERTED THE LINE ON (B)(6) 2024. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20767765 ↗MalfunctionOct 11, 2024Activation Problem; Activation Problem
IT WAS REPORTED NURSE INSERTED INTRODUCER FOR PICC PLACEMENT AND THE SAFETY WOULD NOT ACTIVATE ON THE INTRODUCER. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20433257 ↗MalfunctionMay 15, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED WHEN PLACING A PICC, THE CLINICIAN NOTICED AN UNIDENTIFIABLE SPECK ON THE END OF THE DEAD END CAP INSIDE THE STERILE KIT. NO PATIENT IMPACT REPORTED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19319994 ↗MalfunctionApr 16, 2024Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT TRANSITION FROM THE PEEL AWAY SHEATH WILL ROLL UP ON TISSUE. IT SPLITS IT AND HAVE TO GET A NEW ONE. IT IS SOMETHING WITH THE TRANSITION THAT CATCHES. HAVE TO GO BACK IN AND KNICK A SECOND TIME TO TRY TO ADVANCE THE INTRODUCER. THERE WAS NO ADVERSE OUTCOME OR SERIOUS INJURY TO THE PATIENT OR HEALTHCARE PROVIDER. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19114953 ↗MalfunctionOct 7, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER VIA MEDWATCH THE "PICC LINE INSERTED UNABLE TO PASS, BEGAN TO PULL CATHETER OUT TO END PROCEDURE AND BEGIN TO MEET RESISTANCE AT 37 CM. MASSAGED MUSCLE AND INJECTED REST OF LIDOCAINE TO AREA. CONTINUED TO MASSAGE AREA AND LINE WAS THEN ABLE TO BECOME DISLODGED AND A POP SOUND. AT THE END OF THE TIP WAS A SMALL KNOT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15564369 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page