ARROW IPN000149Recalls & reported problems
Teleflex Incorporated · Super Arrow-Flex(R) Percutaneous Sheath Introducer Set with Integral Hemostasis Valve/Side Port (accepts .035" - .038" wire) · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801902001979), received between Jun 15, 2023 and Aug 13, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K924607); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation2
- Crack1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED WHEN THE IABP WAS PLACED ON 7/15, THE TELEFLEX SHEATH WAS LEAKING AFTER IT WAS INSERTED, AND A CRACK WAS NOTED AT THE HUB. THE PHYSICIAN HAD TO EXCHANGE THE SHEATH FOR A NEW ONE. THERE WAS NO PATIENT INJURY OR HARM. THE SHEATH WAS EXCHANGED WITHOUT ANY ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17538024 ↗MalfunctionAug 13, 2023Material Separation; Material Separation
DURING AN IABP REMOVAL, THE SHEATH BROKE APART AS THE PHYSICIAN WAS REMOVING THE BALLOON CATHETER. NO FOREIGN BODIES WERE LEFT IN THE PATIENT. IT WAS REPORTED THE SHEATH BROKE AT THE HUB AND THE BROKEN PIECE REMAINED IN THE IABP CATHETER, NOTHING BROKE INSIDE THE PATIENT'S VESSEL. NO INTERVENTION WAS NEEDED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17538017 ↗MalfunctionJun 15, 2023Material Separation; Material Separation
IT WAS REPORTED THE PHYSICIAN WAS REMOVING A 40CC BALLOON CATHETER FROM A BALLOON PUMP THROUGH THE 8F SHEATH AND THE SHEATH BROKE AT THE HUB/SHAFT OF THE SHEATH. NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE. ADDITIONALLY A MEDWATCH REPORT WAS RECEIVED AS FOLLOWS: "COULD NOT REPOSITION THE ITEM CORRECTLY AND THE BALLOON PUMP WAS REMOVED. MD PULLED BACK ON THE IABP TO REPOSITION AND WAS UNABLE TO GET IN CORRECT POSITION, MD REMOVED IABP AND WHEN REMOVING IABP THE 8F ARROWFLEX SHEATH COLLAPSED AND SNAPPED AT THE HUB. REINSTALLED THIS MONTH."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17134927 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K924607. One submission often covers many models, so check the product and lot details in each record.
- Z-1064-2016 · Jan 12, 2016 · TerminatedSuper Arrow-Flex Percutaneous Sheath Introducer Sets REF CL-07035 or CL-07635 or CL-07735 or CL-07835 Super Arrow-Flex Percutaneous Sheath Introducer Set with Integral Hemostasis
- Z-1491-2009 · Feb 11, 2009 · TerminatedArrow Super Arrow-Flex Percutaneous Sheath Introducers
- Z-0088-04 · Oct 22, 2003 · Terminated10 FR Super Arrow-Flex Percutaneous Sheath Introducer Set
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page