Skip to content

ARROW IPN000158Recalls & reported problems

Teleflex Incorporated · Super Arrow-Flex(R) Percutaneous Sheath Introducer Set with Integral Hemostasis Valve/Side Port and Radiopaque Tip Marker Band · Class 2

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00801902002075), received between Jan 11, 2024 and Sep 3, 2025 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Material Separation” (8 reports).

3 other recalls cover devices cleared under the same FDA submission (K924607); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation8
  • Break2
  • Detachment of Device or Device Component2
  • Fluid/Blood Leak2
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED "WHEN THE USER REMOVED THE DILATOR FROM THE SHEATH BEFORE USE, THE HEMOSTASIS VALVE GOT DETACHED WITH THE DILATOR. T HEREFORE, A NEW KIT WAS USED INSTEAD." THIS WAS FOUND PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22951908 ↗
  • MalfunctionAug 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED "WHEN THE USER WAS REMOVING THE DILATOR DURING USE, THE HEMOSTASIS VALVE GOT DETACHED WITH THE DILATOR. THEREFORE, THE SHEATH WAS REMOVED AND REPLACED WITH A NEW KIT. NO INJURY TO THE PATIENT OCCURRED." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22730342 ↗
  • MalfunctionFeb 7, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT "THE DOCTOR FOUND SHEATH HUB LEAK DURING USED ON THE PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21332035 ↗
  • MalfunctionSep 18, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT "PRIOR TO THE PROCEDURE ACCORDING TO IFU, THE MD FOUND SHEATH VALVE RUBBER OUT PHENOMENON OCCURS." ASSOCIATED COMPLAINTS: 9680794-2024-00927.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20246093 ↗
  • MalfunctionSep 13, 2024Material Separation; Material Separation

    IT WAS REPORTED "PRIOR TO THE PROCEDURE ACCORDING TO IFU, THE MD FOUND SHEATH VALVE RUBBER OUT PHENOMENON OCCURS." THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20212295 ↗
  • MalfunctionSep 3, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT: "DURING THE PROCEDURE ACCORDING TO IFU, THE MD FOUND AIR IS GENERATED DUE TO DEFECTIVE PROXIMAL HUB." THERE WAS NO REPORTED PATIENT HARM OR CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20123558 ↗
  • MalfunctionMay 1, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED "BLOOD LEAKAGE FROM A THREE-WAY VALVE WAS FOUND." THE USER CHANGED TO A NEW SET. NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENTS CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219943 ↗
  • MalfunctionApr 23, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED "DURING THE PROCEDURE ACCORDING TO IFU, THE MD IDENTIFIED THAT VALVE BREAKAGE OCCURRED. SO THE MD OPEND UP THE NEW KIT TO FINISH THE PROCEDURE." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATEINT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19164756 ↗
  • MalfunctionApr 23, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED "DURING THE PROCEDURE ACCORDING TO IFU, THE MD IDENTIFIED THAT VALVE BREAKAGE OCCURRED. SO THE MD OPENED UP THE NEW KIT TO FINISH THE PROCEDURE." THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19164717 ↗
  • MalfunctionApr 8, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: THE DILATOR BODY AND THE HUB SEPARATED DURING USE. THERE WAS NO PART LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETE AND THERE WAS NO INJURY OR HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19057684 ↗
  • MalfunctionJan 11, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: THE HEMOSTASIS VALVE OF THE SHEATH SNAPPED OFF AND GOT DETACHED DURING USE. THEREFORE, THE SHEATH WAS REMOVED AND REPLACED WITH A NEW KIT. NO INJURY TO THE PATIENT OCCURRED. ADDITIONAL INFORMATION: WHEN THE USER REMOVED THE CATHETER FROM THE SHEATH, A PART OF THE HEMOSTASIS VALVE GOT DETACHED AND REMOVED WITH THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491976 ↗
  • MalfunctionJan 11, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: THE HEMOSTASIS VALVE OF THE SHEATH SNAPPED OFF AND GOT DETACHED DURING USE. THEREFORE, THE SHEATH WAS REMOVED AND REPLACED WITH A NEW KIT. NO INJURY TO THE PATIENT OCCURRED. ADDITIONAL INFORMATION: WHEN THE USER REMOVED THE CATHETER FROM THE SHEATH, A PART OF THE HEMOSTASIS VALVE GOT DETACHED AND REMOVED WITH THE CATHETER. ASSOCIATED TO 9680794-2024-00052.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491919 ↗
  • MalfunctionJan 11, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: THE HEMOSTASIS VALVE OF THE SHEATH SNAPPED OFF AND GOT DETACHED DURING USE. THEREFORE, THE SHEATH WAS REMOVED AND REPLACED WITH A NEW KIT. NO INJURY TO THE PATIENT OCCURRED. ADDITIONAL INFORMATION: WHEN THE USER REMOVED THE CATHETER FROM THE SHEATH, A PART OF THE HEMOSTASIS VALVE GOT DETACHED AND REMOVED WITH THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491857 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K924607. One submission often covers many models, so check the product and lot details in each record.

  • Z-1064-2016 · Jan 12, 2016 · Terminated
    Super Arrow-Flex Percutaneous Sheath Introducer Sets REF CL-07035 or CL-07635 or CL-07735 or CL-07835 Super Arrow-Flex Percutaneous Sheath Introducer Set with Integral Hemostasis
  • Z-1491-2009 · Feb 11, 2009 · Terminated
    Arrow Super Arrow-Flex Percutaneous Sheath Introducers
  • Z-0088-04 · Oct 22, 2003 · Terminated
    10 FR Super Arrow-Flex Percutaneous Sheath Introducer Set

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page