ACUITY™ X4 Spiral S 4674Recalls & reported problems

Boston Scientific Corporation · Pace/Sense Lead · Class 3 · Implant

All models of ACUITY X4 Spiral S (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Acuity X4 Spiral S lead is a wire that connects a pacemaker to the heart. Many reports describe the lead moving out of place, failing to pace the heart, or giving high electrical resistance readings, and many mention infection, sometimes leading to surgery to replace the lead. The records include 1,952 reports, 1,298 injuries and 15 deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,978
Deaths reported
16
Injuries reported
1,317
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,978 adverse event reports that list this exact device (DI 00802526524554), received between Apr 12, 2017 and Sep 30, 2026 — classified as 16 deaths, 1,317 injuries, 645 malfunctions.

The most frequently reported specific device problem is “Failure to Capture” (247 reports).

19 other recalls cover devices cleared under the same FDA submission (P010012); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 26 reportsJan 2024: 32 reportsFeb 2024: 31 reportsMar 2024: 37 reportsApr 2024: 24 reportsMay 2024: 22 reportsJun 2024: 33 reportsJul 2024: 19 reportsAug 2024: 30 reportsSep 2024: 21 reportsOct 2024: 33 reportsNov 2024: 19 reportsDec 2024: 12 reportsJan 2025: 14 reportsFeb 2025: 19 reportsMar 2025: 31 reportsApr 2025: 17 reportsMay 2025: 27 reportsJun 2025: 23 reportsJul 2025: 25 reportsAug 2025: 19 reportsSep 2025: 21 reportsOct 2025: 20 reportsNov 2025: 23 reportsDec 2025: 24 reportsJan 2026: 17 reportsFeb 2026: 19 reportsMar 2026: 27 reportsApr 2026: 16 reportsMay 2026: 17 reportsJun 2026: 11 reportsJul 2026: 16 reportsAug 2026: 13 reportsSep 2026: 24 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem974
  • Failure to Capture247
  • Device Dislodged or Dislocated222
  • High impedance176
  • High Capture Threshold175
  • Defective Device145
  • No Apparent Adverse Event101
  • Material Integrity Problem78

Patient problems

  • Unspecified Infection848
  • No Clinical Signs, Symptoms or Conditions709
  • Sepsis230
  • No Known Impact Or Consequence To Patient110
  • No Code Available107
  • No Consequences Or Impact To Patient82
  • Device Overstimulation of Tissue68
  • Bacterial Infection57

Recent adverse event reports

The 25 most recent of 1,978 reports, in the reporter's own words as published by the FDA.

  • DeathSep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient passed away on September 9th, 2026. During a routine cardiac resynchronization therapy defibrillator (CRT-D) system implant, the patient experienced cardiac arrest. The cardiac arrest occurred while this left ventricular (LV) lead was being implanted and after the right ventricular (RV) lead was implanted. Backup pacing was provided on the RV lead. Cardiopulmonary resuscitation (CPR) was given to the patient, and the patient experienced ventricular tachycardia (VT) and ventricular fibrillation (VF) several times. The field representative noted the patient was stable after the procedure. It was later reported by the field representative that the patient passed away overnight. It was noted the CRT-D was never implanted in the patient. This LV lead was an attempted implant. No additional adverse patient effects were reported.

    Patient problems: Cardiac Arrest; Tachycardia; Ventricular Fibrillation

    MAUDE report 50950617 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50941339 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50941282 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50923949 ↗
  • InjurySep 26, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. A temporary pacing system was implanted until the infection anomaly clears. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50907234 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to a fall and landed on device which resulted to a pocket infection. After discussion, the physician opted to wash out the pocket. There were no additional adverse patient effects reported. This LV lead remained in service.

    Patient problems: Unspecified Infection

    MAUDE report 50903015 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) system exhibited infection. Subsequently, the device and leads were explanted. No additional adverse patient effects were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50889337 ↗
  • InjurySep 23, 2026Material Integrity Problem

    It was reported that this left ventricular (LV) lead was not successfully implanted due to the pigtail portion not straightening as intended when a Whisper wire was advanced through the distal tip, resulting in perforation of the lead insulation and exit through the insulation. This LV lead was not implanted, was replaced with a new LV lead of the same model and returned for analysis. No adverse patient effects were reported.

    Patient problems: Perforation

    MAUDE report 50871438 ↗
  • InjurySep 21, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to an allergic reaction causing an infection at the implant site. The patient had swelling and brownish drainage was coming out of the incision. The physician noted that the patient had some sort of hypersensitivity to the recent implant. There was no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Hypersensitivity/Allergic reaction; Purulent Discharge; Swelling/ Edema; Unspecified Infection

    MAUDE report 50849002 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was explanted due to the patient suffering an infection. No additional adverse patient effects were reported.

    Patient problems: Unspecified Infection

    MAUDE report 50818134 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There was no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50777466 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There was no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50774390 ↗
  • MalfunctionSep 11, 2026Positioning Problem; Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50744419 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There were no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50739872 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50725377 ↗
  • MalfunctionSep 10, 2026Failure to Capture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50724868 ↗
  • InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50711054 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There was no additional adverse patient effects reported. The LV lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50611794 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50606761 ↗
  • InjurySep 4, 2026High Capture Threshold

    It was reported that this left ventricular (LV) lead was explanted and replaced due to high thresholds. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602118 ↗
  • MalfunctionSep 4, 2026Failure to Capture; High Capture Threshold

    It was reported that this cardiac resynchronization therapy defibrillator (CRT-D) exhibited high capture thresholds and loss of capture (LOC) for its left ventricular (LV) lead. Technical Services (TS) was consulted and discussed programming options and settings. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50601423 ↗
  • InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50594221 ↗
  • InjurySep 3, 2026Device Dislodged or Dislocated

    It was reported that this left ventricular (LV) lead was a part of t an entire system was extraction due to non-compliant patient and lead dislodgement. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50594218 ↗
  • InjurySep 2, 2026No Apparent Adverse Event

    It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50566125 ↗
  • MalfunctionAug 28, 2026Adverse Event Without Identified Device or Use Problem; Positioning Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50509676 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P010012 on May 2, 2002. Submitted by Boston Scientific Corp as “Contak Cd,Easytrak, Livian, Cognis And Acuity Spirial Automatic Implantable Cardiac Resynchronization Therapy Defibrilll”. The FDA review took 429 days. The FDA has since approved 581 supplements (changes to the device or its labeling), most recently on Nov 11, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

The FDA lists this type of device as life-sustaining or life-supporting.

Recalls under the same FDA submission

These recalls involve devices cleared under P010012. One submission often covers many models, so check the product and lot details in each record.

  • Z-1671-2022 · Jul 7, 2022 · Open, Classified
    The PRM is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The software in use controls all
  • Z-0081-2018 · Oct 4, 2017 · Terminated
    RESONATE X4 CRT-D, Model G447, Sterile.
  • Z-0080-2018 · Oct 4, 2017 · Terminated
    MOMENTUM X4 CRT-D, Model G138, Sterile.
  • Z-0079-2018 · Oct 4, 2017 · Terminated
    VIGILANT X4 CRT-D, Model G247, Sterile. Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) are indicated for patients with heart failure who receive stable optimal pharma
  • Z-1556-2010 · Mar 15, 2010 · Terminated
    Boston Scientific LIVIAN Cardiac Resynchronization Therapy Defibrillator (CRT-D), models H220, H225, H227, H229. The LIVIAN family of cardiac resynchronization therapy defibrillat
  • Z-0609-2010 · Dec 1, 2009 · Terminated
    Boston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. the following models are not available in the US: models N106, N107, P106, P107,
  • Z-1053-2007 · Mar 23, 2007 · Terminated
    VITALITY 2 (model T165, T175), VITALITY DS (model T125, T135), VITALITY EL (model T127), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventric
  • Z-1049-2007 · Mar 23, 2007 · Terminated
    CONTAK RENEWAL 3 RF (model H210, H215), CONTAK RENEWAL 3 RF HE (model H217, H219) cardiac resynchronization therapy defibrillator (CRT-D).
  • Z-1055-2007 · Mar 23, 2007 · Terminated
    VITALITY 2 EL DR/VR (model T167, T177). contents have been sterilized with gaseous ethylene oxide. The Guidant Vitality 2 ICD, implantable cardioverter defibrillators, are design
  • Z-1052-2007 · Mar 23, 2007 · Terminated
    VITALITY AVT (model A155). Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. P
  • Z-1047-2007 · Mar 23, 2007 · Terminated
    CONTAK RENEWAL 3 (model H170, H175) CONTAK RENEWAL 3 HE (model H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North,
  • Z-1054-2007 · Mar 23, 2007 · Terminated
    VITALITY DR HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 U
  • Z-0047-2008 · Mar 16, 2007 · Terminated
    Guidant VITALITY VR/DR, Implantable Cardioverter Defibrillator, ICD, Models 1870 and 1871. Guidant VITALITY DR+, Implantable Cardioverter Defibrillator, ICD, Model 1872.
  • Z-1288-06 · May 15, 2006 · Terminated
    Guidant CONTAK RENEWAL 3 (models H170, H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Ave
  • Z-0826-06 · Apr 7, 2006 · Terminated
    Guidant CONTAK RENEWAL 4 (models H190, H195) CONTAK RENEWAL; 4 HE (model H197). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Ha

…and 4 more.

Common questions about the ACUITY™ X4 Spiral S 4674

Has the ACUITY™ X4 Spiral S 4674 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00802526524554), based on the FDA recall file dated Nov 27, 2026. However, 19 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the ACUITY™ X4 Spiral S 4674?

The FDA's MAUDE database lists 1,978 adverse event reports for this device, received between Apr 12, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Capture” (247), “Device Dislodged or Dislocated” (222) and “High impedance” (176); the most common patient problems are “Unspecified Infection” (848), “Sepsis” (230) and “Device Overstimulation of Tissue” (68). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ACUITY™ X4 Spiral S 4674?

16 reports are classified as deaths, 1,317 as injuries and 645 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ACUITY™ X4 Spiral S 4674 FDA approved?

It was approved through premarket approval (PMA) P010012 on May 2, 2002.

Is the ACUITY™ X4 Spiral S 4674 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the ACUITY™ X4 Spiral S 4674, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand ACUITY™ X4 Spiral S, model 4674). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.