ACUITY™ X4 Spiral S 4674Recalls & reported problems
Boston Scientific Corporation · Pace/Sense Lead · Class 3 · Implant
Quick summary
This Boston Scientific Acuity X4 Spiral S lead is a wire that connects a pacemaker to the heart. Many reports describe the lead moving out of place, failing to pace the heart, or giving high electrical resistance readings, and many mention infection, sometimes leading to surgery to replace the lead. The records include 1,952 reports, 1,298 injuries and 15 deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,975
- Deaths reported
- 16
- Injuries reported
- 1,317
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,975 adverse event reports that list this exact device (DI 00802526524554), received between Apr 12, 2017 and Sep 30, 2026 — classified as 16 deaths, 1,317 injuries, 642 malfunctions.
The most frequently reported specific device problem is “Failure to Capture” (245 reports).
19 other recalls cover devices cleared under the same FDA submission (P010012); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem972
- Failure to Capture245
- Device Dislodged or Dislocated223
- High Capture Threshold176
- High impedance176
- Defective Device145
- No Apparent Adverse Event101
- Material Integrity Problem79
Patient problems
- Unspecified Infection846
- No Clinical Signs, Symptoms or Conditions707
- Sepsis230
- No Known Impact Or Consequence To Patient110
- No Code Available107
- No Consequences Or Impact To Patient82
- Device Overstimulation of Tissue68
- Bacterial Infection57
Recent adverse event reports
The 25 most recent of 1,975 reports, in the reporter's own words as published by the FDA.
DeathSep 30, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient passed away on September 9th, 2026. During a routine cardiac resynchronization therapy defibrillator (CRT-D) system implant, the patient experienced cardiac arrest. The cardiac arrest occurred while this left ventricular (LV) lead was being implanted and after the right ventricular (RV) lead was implanted. Backup pacing was provided on the RV lead. Cardiopulmonary resuscitation (CPR) was given to the patient, and the patient experienced ventricular tachycardia (VT) and ventricular fibrillation (VF) several times. The field representative noted the patient was stable after the procedure. It was later reported by the field representative that the patient passed away overnight. It was noted the CRT-D was never implanted in the patient. This LV lead was an attempted implant. No additional adverse patient effects were reported.
Patient problems: Cardiac Arrest; Tachycardia; Ventricular Fibrillation
MAUDE report 50950617 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50941339 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50941282 ↗InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50923949 ↗InjurySep 26, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50907234 ↗InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this left ventricular (LV) lead was part of a system revision due to a fall and landed on device which resulted to a pocket infection. After discussion, the physician opted to wash out the pocket. There were no additional adverse patient effects reported. This LV lead remained in service.
Patient problems: Unspecified Infection
MAUDE report 50903015 ↗InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) system exhibited infection. Subsequently, the device and leads were explanted. No additional adverse patient effects were reported.
Patient problems: Unspecified Infection
MAUDE report 50889337 ↗InjurySep 23, 2026Material Integrity Problem
It was reported that this left ventricular (LV) lead was not successfully implanted due to the pigtail portion not straightening as intended when a Whisper wire was advanced through the distal tip, resulting in perforation of the lead insulation and exit through the insulation. This LV lead was not implanted, was replaced with a new LV lead of the same model and returned for analysis. No adverse patient effects were reported.
Patient problems: Perforation
MAUDE report 50871438 ↗InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this left ventricular (LV) lead was explanted due to the patient suffering an infection. No additional adverse patient effects were reported.
Patient problems: Unspecified Infection
MAUDE report 50818134 ↗InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There was no additional adverse patient effects reported. The LV lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50777466 ↗InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There was no additional adverse patient effects reported. The LV lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50774390 ↗MalfunctionSep 11, 2026Insufficient Device Problem Information
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50744419 ↗InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50739872 ↗InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50725377 ↗MalfunctionSep 10, 2026Failure to Capture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50724868 ↗InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50711054 ↗InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this left ventricular (LV) lead was part of a system revision due to infection. There was no additional adverse patient effects reported. The LV lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50611794 ↗InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50606761 ↗InjurySep 4, 2026High Capture Threshold
It was reported that this left ventricular (LV) lead was explanted and replaced due to high thresholds. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602118 ↗MalfunctionSep 4, 2026Failure to Capture; High Capture Threshold
It was reported that this cardiac resynchronization therapy defibrillator (CRT-D) exhibited high capture thresholds and loss of capture (LOC) for its left ventricular (LV) lead. Technical Services (TS) was consulted and discussed programming options and settings. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50601423 ↗InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this implantable lead was part of the system revision due to infection. No adverse patient effects were reported. The implantable lead was explanted.
Patient problems: Unspecified Infection
MAUDE report 50594221 ↗InjurySep 3, 2026Device Dislodged or Dislocated
It was reported that this left ventricular (LV) lead was a part of t an entire system was extraction due to non-compliant patient and lead dislodgement. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50594218 ↗InjurySep 2, 2026No Apparent Adverse Event
Patient problems: Unspecified Infection
MAUDE report 50566125 ↗MalfunctionAug 28, 2026Adverse Event Without Identified Device or Use Problem; Positioning Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50509676 ↗InjuryAug 20, 2026Insufficient Device Problem Information
It was reported that this left ventricular (LV) lead and the right ventricular (RV) lead were explanted due to an unknown product performance issue. This LV lead was successfully replaced. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50351155 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010012. One submission often covers many models, so check the product and lot details in each record.
- Z-1671-2022 · Jul 7, 2022 · Open, ClassifiedThe PRM is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The software in use controls all
- Z-0081-2018 · Oct 4, 2017 · TerminatedRESONATE X4 CRT-D, Model G447, Sterile.
- Z-0080-2018 · Oct 4, 2017 · TerminatedMOMENTUM X4 CRT-D, Model G138, Sterile.
- Z-0079-2018 · Oct 4, 2017 · TerminatedVIGILANT X4 CRT-D, Model G247, Sterile. Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) are indicated for patients with heart failure who receive stable optimal pharma
- Z-1556-2010 · Mar 15, 2010 · TerminatedBoston Scientific LIVIAN Cardiac Resynchronization Therapy Defibrillator (CRT-D), models H220, H225, H227, H229. The LIVIAN family of cardiac resynchronization therapy defibrillat
- Z-0609-2010 · Dec 1, 2009 · TerminatedBoston Scientific COGNIS CRT-D, Cardiac Resynchronization Therapy Defibrillator, models N118, N119. the following models are not available in the US: models N106, N107, P106, P107,
- Z-1049-2007 · Mar 23, 2007 · TerminatedCONTAK RENEWAL 3 RF (model H210, H215), CONTAK RENEWAL 3 RF HE (model H217, H219) cardiac resynchronization therapy defibrillator (CRT-D).
- Z-1052-2007 · Mar 23, 2007 · TerminatedVITALITY AVT (model A155). Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. P
- Z-1055-2007 · Mar 23, 2007 · TerminatedVITALITY 2 EL DR/VR (model T167, T177). contents have been sterilized with gaseous ethylene oxide. The Guidant Vitality 2 ICD, implantable cardioverter defibrillators, are design
- Z-1054-2007 · Mar 23, 2007 · TerminatedVITALITY DR HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 U
- Z-1047-2007 · Mar 23, 2007 · TerminatedCONTAK RENEWAL 3 (model H170, H175) CONTAK RENEWAL 3 HE (model H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North,
- Z-1053-2007 · Mar 23, 2007 · TerminatedVITALITY 2 (model T165, T175), VITALITY DS (model T125, T135), VITALITY EL (model T127), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ventric
- Z-0047-2008 · Mar 16, 2007 · TerminatedGuidant VITALITY VR/DR, Implantable Cardioverter Defibrillator, ICD, Models 1870 and 1871. Guidant VITALITY DR+, Implantable Cardioverter Defibrillator, ICD, Model 1872.
- Z-1288-06 · May 15, 2006 · TerminatedGuidant CONTAK RENEWAL 3 (models H170, H175) CONTAK RENEWAL 3 HE (models H177, H179) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Ave
- Z-0827-06 · Apr 7, 2006 · TerminatedGuidant CONTAK RENEWAL 4 AVT (model M170). Cardiac Resynchronization Therapy Defibrillator (CRT-D). Sterile EO. Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55
…and 4 more.
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall264 deaths16,996 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,283 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,513 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,906 reports
- ESSENTIO™ MRI DR L111Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls9 deaths1,790 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.