ACCOLADE™ EL DR L321Recalls & reported problems

Boston Scientific Corporation · Pacemaker · Class 3 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific Accolade pacemaker has 1,827 adverse event reports on record, including 2 deaths and 1,086 injuries. Many reports describe the battery running down faster than expected, and some describe the pacemaker switching into a protective fallback mode, sometimes leading to its removal. Two open recalls name this device, both about high battery resistance that may trigger that fallback mode, with a software update available for one. Reports rose to 346 in the last 12 months, from 201 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,911
Deaths reported
2
Injuries reported
1,096
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1,911 adverse event reports that list this exact device (DI 00802526559242), received between Apr 22, 2016 and Sep 30, 2026 — classified as 2 deaths, 1,096 injuries, 813 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (904 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 12 reportsJan 2024: 22 reportsFeb 2024: 9 reportsMar 2024: 9 reportsApr 2024: 12 reportsMay 2024: 11 reportsJun 2024: 15 reportsJul 2024: 8 reportsAug 2024: 10 reportsSep 2024: 17 reportsOct 2024: 15 reportsNov 2024: 19 reportsDec 2024: 8 reportsJan 2025: 17 reportsFeb 2025: 23 reportsMar 2025: 14 reportsApr 2025: 11 reportsMay 2025: 22 reportsJun 2025: 18 reportsJul 2025: 17 reportsAug 2025: 18 reportsSep 2025: 13 reportsOct 2025: 21 reportsNov 2025: 31 reportsDec 2025: 42 reportsJan 2026: 44 reportsFeb 2026: 39 reportsMar 2026: 49 reportsApr 2026: 52 reportsMay 2026: 35 reportsJun 2026: 32 reportsJul 2026: 35 reportsAug 2026: 26 reportsSep 2026: 45 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202652
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 20, 2025Class IOpen, Classified

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

    Z-0096-2026 · Root cause: Other · Recall details · FDA record ↗

  2. Dec 12, 2024Class IOpen, Classified

    A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

    Z-0822-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Premature Discharge of Battery904
  • Over-Sensing274
  • High impedance270
  • Battery Problem214
  • Defective Device195
  • Adverse Event Without Identified Device or Use Problem141
  • Signal Artifact/Noise141
  • Pacing Problem134

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,163
  • No Known Impact Or Consequence To Patient338
  • Unspecified Infection116
  • No Code Available86
  • No Consequences Or Impact To Patient71
  • Insufficient Information64
  • Syncope34
  • Sepsis33

Recent adverse event reports

The 25 most recent of 1,911 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948571 ↗
  • MalfunctionSep 30, 2026High impedance

    It was reported that this pacemaker and the right ventricular (RV) lead exhibited a Red Call Doctor Icon (RCD). The RCD was recorded due to a high out of range pacing impedance measurement, greater than 3000 Ohms. This pacemaker remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946995 ↗
  • MalfunctionSep 30, 2026Premature Discharge of Battery; Battery Problem

    It was reported that the implantable pacemaker device suspected premature battery depletion (PBD) and longevity dropped from 4.5 years to 2.5 years in a span of 3 months. Boston Scientific Technical Services (TS) discussed device appears to be operating as expected and the battery measurements are rounded down which can make it appear to be depleting faster than expected. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943566 ↗
  • MalfunctionSep 29, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941392 ↗
  • InjurySep 28, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, which states that the voltage is too low for projected remaining capacity. This pacemaker was subsequently explanted. A new device was implanted. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926607 ↗
  • InjurySep 28, 2026Premature Discharge of Battery

    It was reported that this device recorded a Code 1003, indicating that the voltage is too low for the projected remaining capacity. The device was found to be in storage mode with no therapy available. This pacemaker was subsequently explanted. Another pacemaker was implanted to complete this procedure. This pacemaker has not been returned at this time. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918987 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917388 ↗
  • MalfunctionSep 28, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915008 ↗
  • MalfunctionSep 25, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50896963 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    It was reported that this pacemaker exhibited a nonspecific product performance anomaly. Subsequently, this device was explanted and replaced. This product has not been returned for analysis. No additional adverse patient effects were reported.\\Additional information was received stating this patient had initially presented to the emergency room due to syncope and intermittent loss of capture (LOC) on the right ventricular (RV) lead. The patient also reported to feel fatigue. Subsequently, a pocket revision was performed and the RV lead was noted to have fractured. The device was interrogated and 1.5 years of battery were left remaining, due to this, the physician elected to replace the device at pocket revision and avoid a further replacement intervention in the near future. The RV lead was capped and a new lead was implanted with adequate sensing and impedance measurements obtained. No additional adverse patient effects were reported.

    Patient problems: Fatigue; No Clinical Signs, Symptoms or Conditions; Syncope/Fainting

    MAUDE report 50893447 ↗
  • MalfunctionSep 24, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage was too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring until radio frequency (RF) telemetry disabled was recommended. The device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884199 ↗
  • MalfunctionSep 24, 2026Nonstandard Device

    It was reported that this pacemaker was returned with no reported product performance issues or adverse patient effects. Analysis completed in our Post Market Quality Assurance laboratory identified a product performance issue. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876651 ↗
  • MalfunctionSep 23, 2026Over-Sensing

    It was reported that this pacemaker had a history of exhibiting crosstalk. Technical Services (TS) reviewed device data and recommended programming changes. The health care professional (HCP) indicated that the patient would be further evaluated. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50871814 ↗
  • MalfunctionSep 22, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage was too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continue monitoring until a radio frequency (RF) telemetry is disabled was recommended. The device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859981 ↗
  • MalfunctionSep 18, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50831065 ↗
  • MalfunctionSep 18, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819781 ↗
  • MalfunctionSep 18, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50813074 ↗
  • MalfunctionSep 17, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805832 ↗
  • MalfunctionSep 17, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50804000 ↗
  • MalfunctionSep 16, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50790272 ↗
  • MalfunctionSep 15, 2026Signal Artifact/Noise; Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775863 ↗
  • InjurySep 15, 2026Insufficient Device Problem Information

    It was reported that this device was explanted due to an unknown product anomaly. A new device was successfully implanted. No additional adverse patient effects were reported. This device is no longer in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771531 ↗
  • MalfunctionSep 11, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Device replacement was recommended. No adverse patient effects were reported. This device remains in service at this time.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50736654 ↗
  • MalfunctionSep 9, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50703211 ↗
  • InjurySep 9, 2026Premature Discharge of Battery

    It was reported that this pacemaker recorded a Code 1003, indicating the voltage was too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Device replacement was recommended. Subsequently, this pacemaker was explanted and replaced. this device has been received for instigation. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50702252 ↗

FDA clearance and classification

Approved through premarket approval (PMA) N970003 on Jun 3, 1999. Submitted by Boston Scientific Corp as “Pulsar/Pulsar Max Implantable Pulse Generator System With Consult Software”. The FDA review took 633 days. The FDA has since approved 160 supplements (changes to the device or its labeling), most recently on Oct 3, 2026.

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

The FDA lists this type of device as life-sustaining or life-supporting.

Common questions about the ACCOLADE™ EL DR L321

Has the ACCOLADE™ EL DR L321 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0096-2026 (Class I), began on Aug 20, 2025: Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ACCOLADE™ EL DR L321?

The FDA's MAUDE database lists 1,911 adverse event reports for this device, received between Apr 22, 2016 and Sep 30, 2026. In those reports, the most common specific device problems are “Premature Discharge of Battery” (904), “Over-Sensing” (274) and “High impedance” (270); the most common patient problems are “Unspecified Infection” (116), “Syncope” (34) and “Sepsis” (33). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ACCOLADE™ EL DR L321?

2 reports are classified as deaths, 1,096 as injuries and 813 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ACCOLADE™ EL DR L321 FDA approved?

It was approved through premarket approval (PMA) N970003 on Jun 3, 1999.

Is the ACCOLADE™ EL DR L321 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the ACCOLADE™ EL DR L321, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand ACCOLADE™ EL DR, model L321). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Implantable pulse generator, pacemaker (non-CRT) devices

All Implantable pulse generator, pacemaker (non-CRT) devices →