ACCOLADE™ MRI DR L311Recalls & reported problems
Boston Scientific Corporation · Pacemaker · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Boston Scientific Accolade MRI DR pacemaker is implanted to regulate heartbeat. Reports mostly describe batteries draining early, noise sensed as heartbeats, and trouble communicating by radio. There are 7,044 reports, including 59 deaths and 4,139 injuries. Two recalls are open, both tied to pacemakers entering a safety mode because of high battery resistance. Reports rose to 1,479 in the last 12 months from 955 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 7,283
- Deaths reported
- 59
- Injuries reported
- 4,160
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 7,283 adverse event reports that list this exact device (DI 00802526559228), received between Jul 16, 2016 and Sep 30, 2026 — classified as 59 deaths, 4,160 injuries, 3,064 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (2,020 reports).
2 FDA recall records name this device's identifier, 2 still open.
26 other recalls cover devices cleared under the same FDA submission (P150012); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 20, 2025Class IOpen, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Z-0095-2026 · Root cause: Other · Recall details · FDA record ↗
- Dec 12, 2024Class IOpen, Classified
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Z-0822-2025 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Premature Discharge of Battery2,020
- Adverse Event Without Identified Device or Use Problem1,703
- Over-Sensing991
- High impedance685
- Incorrect, Inadequate or Imprecise Result or Readings625
- Signal Artifact/Noise575
- Battery Problem539
- Pacing Problem528
Patient problems
- No Clinical Signs, Symptoms or Conditions4,106
- Unspecified Infection1,416
- No Known Impact Or Consequence To Patient521
- Sepsis338
- No Code Available292
- No Consequences Or Impact To Patient193
- Pocket Erosion185
- Syncope/Fainting155
Recent adverse event reports
The 25 most recent of 7,283 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950541 ↗MalfunctionSep 30, 2026Premature Discharge of Battery
It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948442 ↗MalfunctionSep 30, 2026Premature Discharge of Battery
It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947235 ↗InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this pacemaker was explanted and replaced. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946996 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941390 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941361 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50941359 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50940889 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937702 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with this pacemaker experienced pocket erosion and infection. This pacemaker was subsequently explanted and has been returned. No additional adverse patient effects were reported.
Patient problems: Pocket Erosion; Unspecified Infection
MAUDE report 50937651 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50937127 ↗MalfunctionSep 29, 2026Premature Discharge of Battery; Use of Device Problem; Interrogation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936902 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50936657 ↗MalfunctionSep 29, 2026Signal Artifact/Noise; Over-Sensing
It was reported that this pacemaker and the right atrial (RA) lead exhibited oversensing of non-physiological and mild myopotential noise. This pacemaker recorded an inappropriate Atrial Tachy Response (ATR) episode due to the oversensing. Technical Services (TS) discussed performing isometrics to test for noise and impedance fluctuations. Also, TS provided reprogramming options. This pacemaker remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935898 ↗MalfunctionSep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Tachycardia
MAUDE report 50935653 ↗MalfunctionSep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935569 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935445 ↗InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this pacemaker was explanted due to an unspecified reason. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934997 ↗MalfunctionSep 29, 2026Over-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934317 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934127 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932362 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. No adverse patient effects were reported. This device remains in service at this time.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931935 ↗MalfunctionSep 29, 2026High impedance
It was reported that this pacemaker exhibited high out of range pacing impedance measurements on the right ventricular (RV) lead. Subsequently, this pacemaker was explanted due to normal battery depletion. This pacemaker was successfully replaced, and no additional adverse patient effects were reported. This pacemaker has not been returned for analysis.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931303 ↗MalfunctionSep 29, 2026Premature Discharge of Battery
It was reported that this pacemaker recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50927809 ↗InjurySep 28, 2026Premature Discharge of Battery; Use of Device Problem; Interrogation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926084 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P150012. One submission often covers many models, so check the product and lot details in each record.
- Z-1770-2026 · Mar 19, 2026 · Open, ClassifiedBoston Scientific Pacemaker, Implantable Pulse Generator, Pacemaker (non- Cardiac Re-Synchronization Therapy) labeled as: 1. ESSENTIO SR SL, Model Number L100; 2. ESSENTIO DR SL
- Z-0084-2026 · Aug 20, 2025 · Open, ClassifiedACCOLADE DR EL MRI (Model L331)
- Z-0091-2026 · Aug 20, 2025 · Open, ClassifiedModel Number L231, PROPONENT DR EL MRI Pacemaker
- Z-0081-2026 · Aug 20, 2025 · Open, ClassifiedACCOLADE SR SL MRI (Model L310)
- Z-0088-2026 · Aug 20, 2025 · Open, ClassifiedModel Number L210 PROPONENT SR SL Pacemaker
- Z-0097-2026 · Aug 20, 2025 · Open, ClassifiedModel Number L131, ESSENTIO DR EL MRI Pacemaker
- Z-0094-2026 · Aug 20, 2025 · Open, ClassifiedModel Number L110, ESSENTIO SR SL MRI Pacemaker
- Z-0082-2026 · Aug 20, 2025 · Open, ClassifiedACCOLADE DR SL MRI (Model L311)
- Z-0089-2026 · Aug 20, 2025 · Open, ClassifiedModel Number L211 PROPONENT DR SL MRI Pacemaker
- Z-0823-2025 · Dec 12, 2024 · Open, ClassifiedPROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221), and EL MRI (Model Number L231).
- Z-0824-2025 · Dec 12, 2024 · Open, ClassifiedESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131)
- Z-2000-2021 · Jun 3, 2021 · Open, ClassifiedAccolade pacemaker, Models: a) L300 b) L301 c) L310 d) L311 e) L321 f) L331
- Z-1999-2021 · Jun 3, 2021 · Open, ClassifiedProponent pacemaker, Models: a) L200 b) L201 c) L209 d) L210 e) L211 f) L221 g) L231
- Z-2017-2021 · Jun 3, 2021 · Open, ClassifiedVITALIO DR EL (pacemaker), Model Numbers: a) J274 b) J277 c) K274 (added 7/15/2021) d) K277 (added 7/15/2021) e) K284 (added 7/15/2021) f) K287 (added 7/15/2021)
- Z-1998-2021 · Jun 3, 2021 · Open, ClassifiedEssentio pacemaker, Models: a) L100 b) L101 c) L110 d) L111 e) L121 f) L131
…and 11 more.
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall264 deaths16,996 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,513 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,906 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,975 reports
- ESSENTIO™ MRI DR L111Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls9 deaths1,790 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page