ZOOM® LATITUDE® 3120 (UDI 00802526562600)Recalls & reported problems
Boston Scientific Corporation · programmer/recorder/monitor · Class 3
All models of ZOOM LATITUDE (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 4
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
4 FDA recall records name this device's identifier, 4 still open.
8 other recalls cover devices cleared under the same FDA submission (P910077); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Nov 30, 2023Class IIOpen, Classified
The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.
Z-0673-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 7, 2022Class IIOpen, Classified
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Z-1669-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 7, 2022Class IIOpen, Classified
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Z-1671-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 7, 2022Class IIOpen, Classified
There is potential for unanticipated software anomalies to occur when using the Model 2868 Application loaded on the 3120 ZOOM LATITUTDE programmer when communicating with certain Boston Scientific transvenous defibrillators.
Z-1670-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P910077. One submission often covers many models, so check the product and lot details in each record.
- Z-1216-2023 · Feb 2, 2023 · Open, ClassifiedLATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
- Z-3260-2018 · Sep 10, 2018 · TerminatedModel 3300 LATITUDE(TM) Programming System Product Usage - The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communic
- Z-1445-2012 · Feb 24, 2012 · TerminatedBoston Scientific LATITUDE Patient Management System Communicator, Model 6476, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, Made in Malaysi
- Z-1096-2012 · Dec 19, 2011 · TerminatedLATITUDE Patient management System, Model 6488, software version 7.2. It is indicated for use to remotely communicate with a compatible Boston Scientific pulse generator and transf
- Z-0333-2012 · Oct 12, 2011 · TerminatedLATITUDE Patient Management System, Models 6441*, 6442*, 6465*, 6488 (USA). There are two software applications: the programmer software and the LATITUDE remote monitoring system
- Z-0037-2012 · Aug 12, 2011 · TerminatedLATITUDE Paceart Integration Software, version 1.00 Model 6472. LATITUDE Paceart Integration is software and therefore does not have any serial numbers. However, the software ver
- Z-1631-2010 · Sep 10, 2009 · TerminatedBoston Scientific, LATITUDE Communicator, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112. Models 6476 (USA) and 6468 (OUS)
- Z-0002-2007 · May 19, 2006 · TerminatedGuidant Zoom Latitude Programming System, Model 3120, Programmer/Recorder/Monitor (PRM). A portable cardiac rhythm management system designed to be used with Guidant implantable p
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page